          Abstract

          Background and Objective

          Conflicting results were found between the I-gel™ and the LMA-Supreme™ during anesthesia, so we conducted a meta-analysis of randomized controlled trials (RCTs) to compare the effectiveness and safety of the I-gel™ vs. the LMA-Supreme™during anesthesia.

          Methods

          A comprehensive search was conducted using Pubmed, EMbase, ISI Web of Knowledge, the Cochrane Library, China Journal Full-text Database, Chinese Biomedical Database, Chinese Scientific Journals Full-text Database, CMA Digital Periodicals, and Google scholar to find RCTs that compare the LMA-S™ with the i-gel™during anesthesia. Two reviewers independently selected trials, extracted data, and assessed the methodological qualities and evidence levels. Data were analyzed by RevMan 5.0 and comprehensive meta-analysis software.

          Results

          Ten RCTs were included. There were no significant differences in oropharyngeal leak pressures (mean difference [MD] 0.72, 95% confidence interval [CI] –1.10 2.53), device placement time (MD –1.3, 95%CI –4.07 1.44), first attempt insertion success (risk ratio [RR] 1.01, 95% CI 0.9 1.14), grade 3 and 4 fiberoptic view (RR 0.89, 95%CI 0.65 1.21), and blood on removal (RR 0.62, 95%CI 0.32 1.22) between the i-gel™ and the LMA-Supreme™, respectively. However, the LMA-Supreme™was associated with easier gastric tube insertion (RR 1.17, 95%CI 1.07 1.29), and more sore throat (RR 2.56, 95%CI 1.60 4.12) than the i-gel™ group.

          Conclusions

          The LMA-Supreme™ and i-gel™ were similarly successful and rapidly inserted. However, the LMA-Supreme™ was shown to be easier for gastric tube insertion and associated with more sore throat compared with the i-gel™.

          Related collections

          Most cited references16

          • Record: found
          • Abstract: found
          • Article: not found

          Initial anatomic investigations of the I-gel airway: a novel supraglottic airway without inflatable cuff.

          The I-gel airway is a novel supraglottic airway that uses an anatomically designed mask made of a gel-like thermoplastic elastomer. We studied the positioning and mechanics of this new device in 65 non-embalmed cadavers with 73 endoscopies (eight had repeat insertion), 16 neck dissections, and six neck radiographs. A full view of the glottis (percentage of glottic opening score 100%) occurred in 44/73 insertions, whereas only 3/73 insertions had epiglottis-only views. Including the eight repeat insertions with a different size, a glottic opening score of > 50% was obtained in all 65 cadavers. The mean percentage of glottic opening score for the 73 insertions was 82% (95% confidence interval 75-89%). In each of the neck dissections and radiographs the bowl of the device covered the laryngeal inlet. We found that the I-gel effectively conformed to the perilaryngeal anatomy despite the lack of an inflatable cuff and it consistently achieved proper positioning for supraglottic ventilation.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Crossover comparison of the laryngeal mask supreme and the i-gel in simulated difficult airway scenario in anesthetized patients.

            The single-use supraglottic airway devices LMA-Supreme (LMA-S; Laryngeal Mask Company, Henley-on-Thames, United Kingdom) and i-gel (Intersurgical Ltd, Wokingham, Berkshire, United Kingdom) have a second tube for gastric tube insertion. Only the LMA-S has an inflatable cuff. They have the same clinical indications and might be useful for difficult airway management. This prospective, crossover, randomized controlled trial was performed in a simulated difficult airway scenario using an extrication collar limiting mouth opening and neck movement. Sixty patients were included. Both devices were placed in random order in each patient. Primary outcome was overall success rate. Other measurements were time to successful ventilation, airway leak pressure, fiberoptic glottic view, and adverse events. Success rate for the LMA-S was 95% versus 93% for the i-gel (P = 1.000). LMA-S needed shorter insertion time (34 +/- 12 s vs. 42 +/- 23 s, P = 0.024). Tidal volumes and airway leak pressure were similar (LMA-S 26 +/- 8 cm H20; i-gel 27 +/- 9 cm H20; P = 0.441). Fiberoptic view through the i-gel showed less epiglottic downfolding. Overall agreement in insertion outcome was 54 (successes) and 1 (failure) or 55 (92%) of 60 patients. The difference in success rate was 1.7% (95% CI -11.3% to 7.6%). Both airway devices had similar insertion success and clinical performance in the simulated difficult airway situation. The authors found less epiglottic downfolding and better fiberoptic view but longer insertion time with the i-gel. Our study shows that both devices are feasible for emergency airway management in patients with reduced neck movement and limited mouth opening.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Comparison of the LMA Supreme vs the i-gel in paralysed patients undergoing gynaecological laparoscopic surgery with controlled ventilation.

              We compared the efficacy of the inflatable cuff of the LMA Supreme against the non-inflatable i-gel cuff in providing an adequate seal for laparoscopic surgery in the Trendelenburg position in 100 female patients. There was no difference in our primary outcome, oropharyngeal leak pressure, between the LMA Supreme and the i-gel (mean (SD) 26.4 (5.1) vs 25.0 (5.7) cmH(2) O, respectively; p = 0.18). Forty-seven (94%) LMA Supremes and 48 (96%) i-gels were successfully inserted on the first attempt, with similar ease, and comparable times to the first capnograph trace (mean (SD) 14.3 (4.7) s for the LMA Supreme vs 15.4 (8.2) s for the i-gel; p = 0.4). Gastric tube insertion was easier and achieved more quickly with the LMA Supreme vs the i-gel (9.0 (2.5) s vs 15.1 (7.3) s, respectively; p < 0.001). After creation of the pneumoperitoneum, there was a smaller difference between expired and inspired tidal volumes with the LMA Supreme (21.5 (15.2) ml) than with the i-gel (31.2 (23.5) ml; p = 0.009). There was blood on removal of two LMA Supremes and one i-gel. Four patients in the LMA Supreme group and one patient in the i-gel group experienced mild postoperative sore throat.
                Bookmark

                Author and article information

                Contributors
                Role: Editor
                Journal
                PLoS One
                PLoS ONE
                plos
                plosone
                PLoS ONE
                Public Library of Science (San Francisco, USA )
                1932-6203
                2013
                12 August 2013
                : 8
                : 8
                : e71910
                Affiliations
                [1 ]Department of Anesthesiology, The First Hospital of China Medical University, Shenyang, China
                [2 ]Department of Anesthesiology, Fengtian Hospital of Shengyang Medical College, Shenyang, China
                [3 ]Department of Neurosurgery, The First Hospital of Shenyang, Shenyang, China
                [4 ]Department of Ophthalmology, The First Hospital of China Medical University, Shenyang, China
                [5 ]Department of Clinical Epidemiology and Evidence Medicine, The First Hospital of China Medical University, Shenyang, China
                University of Colorado, United States of America
                Author notes

                Competing Interests: The authors have declared that no competing interests exist.

                Conceived and designed the experiments: JHJ XGC. Performed the experiments: JHJ LL XGC XFC XMW. Analyzed the data: XMW LL JHJ. Contributed reagents/materials/analysis tools: LL XGC XFC. Wrote the paper: JHJ LL XFC XGC.

                Article
                PONE-D-12-39575
                10.1371/journal.pone.0071910
                3741209
                23951266
                5ebb162a-ee72-4a82-a3a2-6ed827d8b2e9
                Copyright @ 2013

                This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

                History
                : 15 December 2012
                : 6 July 2013
                Page count
                Pages: 8
                Funding
                No current external funding sources for this study.
                Categories
                Research Article
                Medicine
                Anatomy and Physiology
                Respiratory System
                Respiratory Physiology
                Anesthesiology
                Anesthesiology Technology
                Anesthetic Mechanisms
                General Anesthesia
                Pain Management
                Regional Anesthesia
                Drugs and Devices
                Medical Devices
                Surgery
                Anesthesia

                Uncategorized
                Uncategorized

                Comments

                Comment on this article