      Quantitative evaluation of the progressive wear of powered interproximal reduction systems after repeated use : An in vitro study Translated title: Quantitative Untersuchung der fortschreitenden Abnutzung elektrisch angetriebener Interproximal-Reduktionssysteme bei wiederholter Anwendung : Eine In-vitro-Studie

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Purpose

          To evaluate the residual surface roughness of 5 common diamond-coated interproximal reduction (IPR) systems after consecutive in vitro applications in relation to system, diamond grain size, and instrument thickness.

          Methods

          IPR was performed on 80 extracted human incisors using motor-driven strips and discs under predefined conditions. The IPR auxiliaries were applied at 5 consecutive sessions of 20 s on intact interproximal surfaces, and the surface profile (R a, R z, R max) was analyzed at baseline and after each session with an optical profilometer.

          Results

          No overall significant difference in the roughness values was found between systems ( P = 0.07 for R a, P = 0.33 for R z, and P = 0.48 for R max). There was a significant average decrease of R a, R z, and R max for all systems for every unit increase in time by −0.171 μm ( P < 0.001), −3.297 ( P ≤ 0.001), and −2.788 μm ( P = 0.001), respectively. R a, R z, and R max values increased significantly, i.e., by 0.194 μm ( P = 0.003), 5.890 μm ( P = 0.001), and 5.319 μm ( P = 0.010) as instrument thickness increased by one unit. No significant reductions in R a, R z, and R max were observed across grain sizes (−0.008 μm [ P > 0.05], −0.244 μm [ P > 0.05], and −0.179 μm [ P > 0.05], respectively). There was no evidence of interaction between system and time as the P values for R a, R z, and R max were 0.88, 0.51, and 0.70, respectively.

          Conclusions

          All IPR materials presented significant gradual decrease of surface roughness after repeated applications. There were no significant roughness changes among auxiliaries of different grain sizes. Thinner auxiliaries showed significantly more roughness reduction, possibly requiring more frequent replacement than thick auxiliaries in clinical practice.

          Zusammenfassung

          Ziel

          Beurteilt werden sollte die verbleibende Oberflächenrauigkeit von 5 handelsüblichen diamantbeschichteten Interproximal-Reduktionssystemen nach wiederholter In-vitro-Anwendung, differenziert wurde dabei bezüglich System, Diamantkorngröße und Instrumentendicke.

          Methoden

          Interproximal-Reduktion (IPR) wurde mit motorgetriebenen Streifen oder Scheiben an 80 extrahierten menschlichen Inzisiven unter standardisierten Bedingungen durchgeführt. Die IPR-Systeme wurden für 5 aufeinander folgende Anwendungen von 20 s an intakten Interproximalflächen eingesetzt, die Oberflächenrauigkeit (R a, R z, R max) wurde sowohl initial als auch nach jeder Anwendung mit einem optischen Profilometer analysiert.

          Resultate

          Es wurden keine signifikanten Unterschiede der Oberflächenrauigkeit zwischen den unterschiedlichen Systemen festgestellt ( p = 0,07 für R a, p = 0,33 für R z, p = 0,48 für R max). Eine signifikante durchschnittliche Abnahme von R a, R z und R max von −0,171 μm ( p < 0,001), −3,297 μm ( p ≤ 0,001) bzw. −2,788 μm ( p = 0,001) nach jeder Anwendung wurde gefunden. R a, R z und R max Werte nahmen signifikant zu um 0,194 μm ( p = 0,003), 5,890 μm ( p = 0,001) bzw. 5,319 μm ( p = 0,010) bei zunehmender Instrumentendicke pro Einheit. Bezüglich der Diamantkorngrößen wurden keine signifikanten Unterschiede von R a, R z und R max (−0,008 μm [ p > 0,05], −0,244 μm [ p > 0,05] bzw. −0,179 μm [ p > 0,05]) gefunden. Es gab keine Anzeichen für eine Interaktion zwischen Systemtyp und Anwendungsanzahl für R a, R z und R max, mit p-Werten von 0,88, 0,51 bzw. 0,70.

          Schlussfolgerungen

          Alle IPR-Systeme zeigten eine signifikante graduelle Abnahme der Oberflächenrauigkeit nach wiederholten Anwendungen. Es gab keine signifikanten Rauigkeitsdifferenzen zwischen Instrumenten mit unterschiedlichen Diamantkorngrößen. Bei dünneren Instrumenten gab es signifikant höhere Rauigkeitsverluste, was möglicherweise in der Praxis einen häufigeren Wechsel als bei dickeren Instrumenten nötig macht.

          Related collections

          Most cited references29

          • Record: found
          • Abstract: not found
          • Article: not found

          Asymmetry in tooth size: a factor in the etiology, diagnosis, and treatment of malocclusion

            Bookmark
            • Record: found
            • Abstract: not found
            • Article: not found

            2014 JCO study of orthodontic diagnosis and treatment procedures, Part 1: results and trends.

              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Anterior interocclusal relations. Part I.

              It is not uncommon for discrepancies in anterior interocclusal arch length to exist in the orthodontic patient. A frequent cause of this problem is a decrease in upper anterior dental arch length. This discrepancy, if left uncorrected, may result in an end-to-end incisal relation whereby the anterior teeth are unable to perform their function in the mutually protected dentition. This article will discuss the following procedures which may be useful in the correction of discrepancies in anterior interocclusal arch length: (1) repositioning of upright maxillary incisors to increase upper dental arch length, (2) mesiodistal enamel reduction to reduce mandibular dental arch length, promote stability, and improve gingival conditions, and (3) treatment of three-mandibular-incisor cases to decrease mandibular dental arch length, allow correction of crowded mandibular anterior teeth, and reduce protrusion of the lower anterior teeth.
                Bookmark

                Author and article information

                Contributors
                c.livas@acta.nl
                Journal
                J Orofac Orthop
                J Orofac Orthop
                Journal of Orofacial Orthopedics
                Springer Medizin (Heidelberg )
                1434-5293
                1615-6714
                13 November 2019
                13 November 2019
                2020
                : 81
                : 1
                : 22-29
                Affiliations
                [1 ]GRID grid.7177.6, ISNI 0000000084992262, Department of Orthodontics, Academic Centre for Dentistry Amsterdam (ACTA), , University of Amsterdam and VU University Amsterdam, ; Gustav Mahlerlaan 3004, 1081 LA Amsterdam, The Netherlands
                [2 ]GRID grid.5734.5, ISNI 0000 0001 0726 5157, Department of Orthodontics and Dentofacial Orthopedics, Dental School/Medical Faculty, , University of Bern, ; Freiburgstraße 7, 3010 Bern, Switzerland
                [3 ]GRID grid.5734.5, ISNI 0000 0001 0726 5157, Department of Restorative, Preventive and Pediatric Dentistry, Dental School/Medical Faculty, , University of Bern, ; Freiburgstraße 7, 3010 Bern, Switzerland
                Article
                200
                10.1007/s00056-019-00200-x
                6960219
                31720733
                642ce0cf-193b-4380-a028-05fc15a68108
                © The Author(s) 2019

                Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License ( http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

                History
                : 27 March 2019
                : 31 August 2019
                Funding
                Funded by: Academic Center for Dentistry Amsterdam (ACTA)
                Categories
                Original Article
                Custom metadata
                © Springer Medizin Verlag GmbH, ein Teil von Springer Nature 2020

                Orthopedics
                dental high-speed technique,surface roughness,optical profilometer,dental enamel,enamel stripping methods,zahnmedizinische hochgeschwindigkeitstechnik,oberflächenrauigkeit,optische profilometrie,zahnschmelz,methoden zur schmelzreduktion

                Comments

                Comment on this article