      In vivo biocompatibility evaluation of a new resilient, hard-carbon, thin-film coating for ventricular assist devices.

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          The purpose of this study was to evaluate in vivo the biocompatibility of BioMedFlex (BMF), a new resilient, hard-carbon, thin-film coating, as a blood journal bearing material in Cleveland Heart's (Charlotte, NC, USA) continuous-flow right and left ventricular assist devices (RVADs and LVADs). BMF was applied to RVAD rotating assemblies or both rotating and stator assemblies in three chronic bovine studies. In one case, an LVAD with a BMF-coated stator was also implanted. Cases 1 and 3 were electively terminated at 18 and 29 days, respectively, with average measured pump flows of 4.9 L/min (RVAD) in Case 1 and 5.7 L/min (RVAD) plus 5.7 L/min (LVAD) in Case 3. Case 2 was terminated prematurely after 9 days because of sepsis. The sepsis, combined with running the pump at minimum speed (2000 rpm), presented a worst-case biocompatibility challenge. Postexplant evaluation of the blood-contacting journal bearing surfaces showed no biologic deposition in any of the four pumps. Thrombus inside the RVAD inlet cannula in Case 3 is believed to be the origin of a nonadherent thrombus wrapped around one of the primary impeller blades. In conclusion, we demonstrated that BMF coatings can provide good biocompatibility in the journal bearing for ventricular assist devices.

          Related collections

          Author and article information

          Journal
          Artif Organs
          Artificial organs
          Wiley
          1525-1594
          0160-564X
          Dec 2010
          : 34
          : 12
          Affiliations
          [1 ] Department of Biomedical Engineering, Lerner Research Institute, Cleveland Clinic, OH 44195, USA.
          Article
          AOR985
          10.1111/j.1525-1594.2009.00985.x
          20545668
          64417f73-7c9e-41eb-95c8-317caff25959
          History

          Comments

          Comment on this article