      Instrumentation for the measurement of electric brain responses to transcranial magnetic stimulation.

      Medical & Biological Engineering & Computing
      Electrodes, Equipment Design, Humans, Magnetoencephalography, instrumentation

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          There is described a 60-channel EEG acquisition system designed for the recording of scalp-potential distributions starting just 2.5 ms after individual transcranial magnetic stimulation (TMS) pulses. The amplifier comprises gain-control and sample-and-hold circuits to prevent large artefacts from magnetically induced voltages in the leads. The maximum amplitude of the stimulus artefact during the 2.5 ms gating period is 1.7 microV, and 5 ms after the TMS pulse it is only 0.9 microV. It is also shown that mechanical forces to the electrodes under the stimulator coil are a potential source of artefacts, even though, with chlorided silver wire and Ag/AgCl-pellet electrodes, the artefact is smaller than 1 microV. The TMS-compatible multichannel EEG system makes it possible to locate TMS-evoked electric activity in the brain.

          Related collections

          Author and article information

          Journal
          10505382
          10.1007/BF02513307

          Chemistry
          Electrodes,Equipment Design,Humans,Magnetoencephalography,instrumentation
          Chemistry
          Electrodes, Equipment Design, Humans, Magnetoencephalography, instrumentation

          Comments

          Comment on this article