      High-speed optical coherence tomography of corneal opacities.

      Ophthalmology
      Adult, Aged, Corneal Opacity, diagnosis, Female, Humans, Male, Middle Aged, Prospective Studies, Time Factors, Tomography, Optical Coherence, methods, standards, Ultrasonography

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          To evaluate corneal opacities with optical coherence tomography (OCT). Prospective observational case series. Twenty-three eyes of 19 patients with corneal opacities referred to a tertiary medical center. Twenty-three consecutive eyes with corneal opacities were imaged with a high-speed corneal OCT prototype (Carl Zeiss Meditec, Inc., Dublin, CA). The OCT system operates at a speed of 2000 axial scans per second and a wavelength of 1.3 microm. Slit-scanning tomography (Orbscan II, software version 3.12; Bausch & Lomb, Inc., Rochester, NY) and ultrasound pachymetry (Corneo-Gage Plus; Sonogage, Cleveland, OH) also were performed. Central cornea thickness was measured by the 3 instruments. In eyes with central opacities (n = 17), OCT central cornea thickness measurements were statistically equivalent to ultrasound pachymetry, whereas Orbscan II measurements were significantly less than ultrasound pachymetry (difference, -132.7+/-143 microm; P = 0.006). The OCT and ultrasound pachymetry results were obtained for all eyes, whereas Orbscan was unable to provide readings in 4 eyes. In eyes with off-center opacities (n = 6), OCT and Orbscan II central cornea thickness measurements were statistically equivalent to ultrasound pachymetry. The OCT measurements of epithelial thickness and scar depth also were demonstrated. Optical coherence tomography provides consistent pachymetry mapping, whereas Orbscan II significantly underestimates corneal thickness in the presence of central corneal scars. Optical coherence tomography could be valuable in the planning of surgical treatment of corneal scars.

          Related collections

          Author and article information

          Journal
          17307254
          10.1016/j.ophtha.2006.10.033

          Chemistry
          Adult,Aged,Corneal Opacity,diagnosis,Female,Humans,Male,Middle Aged,Prospective Studies,Time Factors,Tomography, Optical Coherence,methods,standards,Ultrasonography

          Comments

          Comment on this article