          Abstract

          Previous studies reported interest in intraoperative shear-wave elastography (SWE) guidance for brain-tumor and epilepsy surgeries. Focal cortical dysplasia (FCD) surgery is one of the most appropriate indications for using SWE guidance. The aim of this study was to evaluate the efficacy of ultrasound SWE techniques for the intraoperative detection of FCDs. We retrospectively analyzed data from 18 adult patients with drug-resistant epilepsy associated with FCD who had undergone SWE-guided surgery. Conventional B-mode images detected FCD in 2 patients (11.1%), while SWE detected FCD in 14 patients (77.8%). The stiffness ratios between MRI-positive and -negative cases were significantly different (3.6 ± 0.4 vs. 2.2 ± 0.6, respectively; p < 0.001). FCDs were significantly more frequently detected by interoperative SWE in women (OR 4.7, 95% CI (1.7–12.7); p = 0.004) and in patients in whom FCD was visible on magnetic resonance imaging (MRI; OR 2.3, 95% CI (1.3–4.3); p = 0.04). At 1 year after surgery and at last follow-up (mean = 21 months), seizure outcome was good (International League Against Epilepsy (ILAE) Class 1 or 2) in 72.2% and 55.6% of patients, respectively. Despite some limitations, our study highlighted the potential of SWE as an intraoperative tool to detect FCD. Future technical developments should allow for optimizing intraoperative surgical-cavity evaluation from the perspective of complete FCD resection. Interobserver reliability of SWE measurements should also be assessed by further studies.

          Related collections

          Most cited references43

          • Record: found
          • Abstract: found
          • Article: found
          Is Open Access

          Ultrasound Elastography: Review of Techniques and Clinical Applications

          Elastography-based imaging techniques have received substantial attention in recent years for non-invasive assessment of tissue mechanical properties. These techniques take advantage of changed soft tissue elasticity in various pathologies to yield qualitative and quantitative information that can be used for diagnostic purposes. Measurements are acquired in specialized imaging modes that can detect tissue stiffness in response to an applied mechanical force (compression or shear wave). Ultrasound-based methods are of particular interest due to its many inherent advantages, such as wide availability including at the bedside and relatively low cost. Several ultrasound elastography techniques using different excitation methods have been developed. In general, these can be classified into strain imaging methods that use internal or external compression stimuli, and shear wave imaging that use ultrasound-generated traveling shear wave stimuli. While ultrasound elastography has shown promising results for non-invasive assessment of liver fibrosis, new applications in breast, thyroid, prostate, kidney and lymph node imaging are emerging. Here, we review the basic principles, foundation physics, and limitations of ultrasound elastography and summarize its current clinical use and ongoing developments in various clinical applications.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Ultrasound elastography: principles and techniques.

            Ultrasonography has been widely used for diagnosis since it was first introduced in clinical practice in the 1970's. Since then, new ultrasound modalities have been developed, such as Doppler imaging, which provides new information for diagnosis. Elastography was developed in the 1990's to map tissue stiffness, and reproduces/replaces the palpation performed by clinicians. In this paper, we introduce the principles of elastography and give a technical summary for the main elastography techniques: from quasi-static methods that require a static compression of the tissue to dynamic methods that uses the propagation of mechanical waves in the body. Several dynamic methods are discussed: vibro-acoustography, Acoustic Radiation Force Impulsion (ARFI), transient elastography, shear wave imaging, etc. This paper aims to help the reader at understanding the differences between the different methods of this promising imaging modality that may become a significant tool in medical imaging.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Shear-Wave Elastography: Basic Physics and Musculoskeletal Applications.

              In the past 2 decades, sonoelastography has been progressively used as a tool to help evaluate soft-tissue elasticity and add to information obtained with conventional gray-scale and Doppler ultrasonographic techniques. Recently introduced on clinical scanners, shear-wave elastography (SWE) is considered to be more objective, quantitative, and reproducible than compression sonoelastography with increasing applications to the musculoskeletal system. SWE uses an acoustic radiation force pulse sequence to generate shear waves, which propagate perpendicular to the ultrasound beam, causing transient displacements. The distribution of shear-wave velocities at each pixel is directly related to the shear modulus, an absolute measure of the tissue's elastic properties. Shear-wave images are automatically coregistered with standard B-mode images to provide quantitative color elastograms with anatomic specificity. Shear waves propagate faster through stiffer contracted tissue, as well as along the long axis of tendon and muscle. SWE has a promising role in determining the severity of disease and treatment follow-up of various musculoskeletal tissues including tendons, muscles, nerves, and ligaments. This article describes the basic ultrasound physics of SWE and its applications in the evaluation of various traumatic and pathologic conditions of the musculoskeletal system. (©)RSNA, 2017.
                Bookmark

                Author and article information

                Contributors
                Role: Academic Editor
                Journal
                J Clin Med
                J Clin Med
                jcm
                Journal of Clinical Medicine
                MDPI
                2077-0383
                03 March 2021
                March 2021
                : 10
                : 5
                : 1049
                Affiliations
                [1 ]Department of Neurosurgery, La Pitié-Salpêtrière University Hospital, Assistance Publique-Hôpitaux de Paris, 75013 Paris, France; stephane.clemenceau@ 123456aphp.fr (S.C.); alexandre.carpentier@ 123456aphp.fr (A.C.)
                [2 ]Faculty of Medicine, Sorbonne University, 75005 Paris, France
                [3 ]Paris Brain Institute (ICM, INSERM, UMRS 1127, CNRS, UMR 7225), 75013 Paris, France
                Author notes
                [* ]Correspondence: bertrand.mathon@ 123456aphp.fr ; Tel.: +33-1-4216-3408
                Author information
                https://orcid.org/0000-0002-9182-5846
                Article
                jcm-10-01049
                10.3390/jcm10051049
                7961510
                33802551
                6f2963d0-377c-4eb6-be42-7ba5e84f9d35
                © 2021 by the authors.

                Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license ( http://creativecommons.org/licenses/by/4.0/).

                History
                : 10 January 2021
                : 28 February 2021
                Categories
                Article

                epilepsy surgery,drug-resistant epilepsy,neurosurgery,shear-wave elastography,real-time guidance,epileptogenic zone,brain surgery,focal epilepsy,ultrasound,seizure outcome

                Comments

                Comment on this article