      Clinical evaluation of the accuracy of the Evident RCM Mark II Apex Locator.

      1 , , ,
      Journal of endodontics
      Elsevier BV

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          The accuracy of the Evident RCM Mark II in locating the apical foramen was evaluated in 37 human teeth scheduled for extraction. After extraction, the difference between the position of the apical foramen determined electronically and its real anatomical location was measured under a binocular microscope, using a micrometer mobile in x, y, and z coordinates. The results show that in 86% (n = 37) of the cases, an accurate location +/- 0.5 mm of the apical foramen was obtained.