      Extraction of a CardioFix neurostimulator with concomitant laser-assisted lead and intracardiac cardiac defibrillator extraction due to Staphylococcus aureus pocket infection

      case-report

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Introduction KEY TEACHING POINTS Key Teaching Points • The CardioFit vagal nerve stimulator is an emerging implantable device in heart failure patients. • Extraction of this device is performed best with operators who have an in-depth knowledge of vital structures in the cervical area. • As lead extractions become more complex, a multidisciplinary approach involving cardiac surgeons will be paramount. This is a case report of the first reported laser lead extraction of a CardioFit vagal nerve stimulator owing to automatic implantable cardioverter-defibrillator (AICD) pocket infection. Case report A 61-year-old man with a history of ischemic cardiomyopathy status post dual-chamber Medtronic Evera XT AICD implanted in the left infraclavicular fossa area in 2007 was subsequently enrolled in the Innervate HF trial in 2013 and had a CardioFix (BioControl Medical LTD, Yehud, Isreal) neurotransmitter implanted. The patient developed severe Staphylococcus aureus infection of his AICD pocket shortly after a pulse generator change of his AICD. The decision was made to remove both implanted devices, including leads, owing to the infection. This is a description of the first reported extraction of the CardioFit neurostimulator owing to infection. Figure 1 shows the chest radiograph of the patient with both devices in place and the infected AICD pocket site. The CardioFit is a vagal nerve stimulator that is designed to modulate heart rate variability in patients with systolic heart failure. Its implantation requires dissection of the right cervical area to expose the right vagus nerve. Then a nerve stimulation lead is inserted with distal tip encircling the vagus nerve. This lead is an asymmetric bipolar multi-contact cuff that is passively fixated circumferentially on the vagus nerve. The lead is then tunneled subcutaneously down to the right infraclavicular area, where it is screwed into the pulse generator. A standard transvenous pacemaker lead with active fixation is placed in the RV apex and attached to the pulse generator. 1 In this particular case a Medtronic CapSureFix Novus 4076-52 cm active-fixation lead was used. The patient was taken to the hybrid operating room. As per our institution’s protocol for lead extraction, general anesthesia was used with endotracheal intubation. A transesophageal echo probe was inserted to evaluate for any signs of thrombus or vegetation around the leads. There was no evidence of baseline pericardial effusion. The patient was prepped and draped to include the neck, anterior chest, and abdomen out to the anterior axillary line and the bilateral groins. Prophylactic right femoral artery and venous sheaths were inserted in case emergent bypass was required. The location of the neurostimulator lead cuff was confirmed using fluoroscopy. An oblique 3 cm incision was made using the previous incision on the anterior border of the sternocleidomastoid muscle in the lower third of the neck. Dissection was performed with electrocautery through the platysma muscle to the level of the carotid sheath. The sternocleidomastoid muscle was retracted laterally. The vagus nerve and vagal nerve stimulator cuff were identified anterior to the internal jugular vein. The cuff was freed from the surrounding adhesions and the vagus nerve was exposed with sharp dissection. The locking mechanism on the stimulating lead cuff was undone and the lead was easily removed from the nerve (Figure 2). Careful inspection of the vagus nerve revealed no visible damage. The pulse generator of the vagal nerve stimulator was then exposed in the deltopectoral region and removed from the pocket. The stimulating lead was easily retracted into the pocket with gentle traction. A stiff stylet was inserted in the inner core of the right ventricular sensing lead and an attempt was made to unscrew the lead, which was unsuccessful. A locking stylet was inserted into the sensing lead and using gentle traction, the lead was extracted without any difficulty. The platysma muscle was reapproximated in the neck and both the right cervical and pocket wounds were closed with absorbable suture. Attention was then given to the infected AICD pocket. The infected device was removed in a similar fashion to the CardioFit device. The 2 leads of the AICD system were a Medtronic 6947-58 cm Sprint Quattro Secure active-fixation AICD lead in the right ventricular apex and a Medtronic CapSureFix Novus 4076-45 cm active-fixation lead placed in the right atrial appendage. These leads were easily extracted with the assistance of the Spectranetics Excimer CVX-300 Laser System (Colorado Springs, CO). Excisional debridement of the device capsule and infected skin edges were performed to the infected pocket and skin. A wound V.A.C. was placed on the ICD pocket wound. The patient left the operating room in stable condition and was discharged several days later following wound vac removal and delayed primary closure of his wound was performed 8 days later without incident. The patient was sent home with a LifeVest (Zoll Medical Corporation, Pittsburgh, PA) until reimplantation of a new AICD. The patient was given intravenous antibiotics appropriate for methicillin-sensitive Staphylococcus aureus for 10 days, then the patient was kept on oral antibiotics for an additional week. AICD reimplantation was subsequently scheduled with the patient’s primary electrophysiologist. Discussion The incidence of intracardiac device infections has seen steady growth over the past 2 decades. This is in part because of the growing number of pacemakers, defibrillators, and resynchronization devices. Extraction of these devices has traditionally been very complex and associated with a high mortality. With the emergence of new technology such as laser and mechanical rotational sheaths, lead extraction has become much more feasible2, 3, 4, 5. Devices such as the CardioFit, which has leads that are extracardiac, require additional skill and expertise. If careful dissection of the lead is not done, it can lead to damage of vital structures, including the carotid artery, the internal jugular vein, and the vagus nerve. As the heart failure patient population continues to grow, the use of devices for autonomic regulation and pressure sensing will also increase. These devices may have extracardiac as well as intracardiac components, making extraction a more in-depth and complex procedure. 5 The design of new implantable intracardiac devices needs to take into consideration the possibility of device infections and that extraction will be required. Therefore, ease of extraction is an important factor that needs to be incorporated in future device design. Operators that are explanting devices need to expand their surgical skills or adopt multidisciplinary teams to tackle such complex extractions. We present the first ever case of the CardioFit neurotransmitter device owing to infection. By incorporating a multidisciplinary effort involving cardiothoracic surgeons and cardiac electrophysiologists, we were able to safely and effectively remove all implanted devices without any morbidity to the patient. Conclusion Laser lead extraction procedures will increase in technical complexity as the heart failure population grows and there is a higher utilization of implantable intracardiac devices for pressure monitoring and autonomic regulation. Therefore, a multidisciplinary approach with electrophysiology and cardiac surgery is required for such complex cases.

          Related collections

          Most cited references2

          • Record: found
          • Abstract: found
          • Article: not found

          Increasing rates of cardiac device infections among Medicare beneficiaries: 1990-1999.

          Although cardiac devices have been found to reduce symptoms and mortality rates in appropriate patient populations, the implications of certain important risks, such as infection, are incompletely understood. The purpose of this study was to use a large population-based database to define the population that is at risk for cardiac device infections, determine the prevalence of device infections, and study changes in the rates of cardiac device implantation and infection in the past decade. Patients with cardiac device implantations and infections were identified with claims files from the Health Care Finance Administration for Medicare beneficiaries from January 1, 1990, through December 31, 1999. Rates of implantation of cardiac devices were determined. Time trend analyses were performed to determine the significance of the observed change in rates. Cardiac device implantation rates increased from 3.26 implantations per 1000 beneficiaries in 1990 to 4.64 implantations per 1000 beneficiaries in 1999, which represents an increase of 42% in 10 years (P for trend <.001). Cardiac device infections showed a larger increase, from 0.94 device infections per 1000 beneficiaries in 1990 to 2.11 device infections per 1000 beneficiaries in 1999, which represents an increase of 124% during the study period (P for trend <.001). During the previous decade, there was a significant increase in both cardiac device implantations and infections in elderly patients, although the increase in the rates of device infections was substantially higher. Additional studies are needed to better understand the relationship and timing between cardiac device implantation and infection.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: found
            Is Open Access

            An Implantable Left Atrial Pressure Sensor Lead Designed for Percutaneous Extraction Using Standard Techniques

            Background An implantable left atrial pressure (LAP) monitoring system for guiding the management of patients with advanced heart failure has the potential to require extraction, particularly in the setting of infection. The LAP sensor lead was designed to be suitable for ease of percutaneous extraction using standard techniques for extracting pacemaker and defibrillator leads. The clinical experience, to date, with percutaneous extraction of the LAP sensor lead is presented. Methods A total of 82 patients underwent successful implantation of the LAP sensor lead using transseptal catheterization. Five patients of the 82 patients during a cumulative follow-up period of 267 patient-years (median of 2.9 years/patient) underwent percutaneous extraction using manual traction with a locking stylet and/or an excimer laser sheath to bore through adhesions. The distal fixation anchors of the LAP sensor lead are designed to fold forward during extraction so that the sensor module can easily separate from the interatrial septum. Results Percutaneous extraction of the LAP sensor lead was accomplished successfully in all five patients with no embolic events, vascular tears, perforations, or other complications requiring surgical intervention. Manual traction alone was sufficient to detach the LAP sensor lead from the interatrial septum in all cases. Use of the excimer laser sheath was needed in selected cases to bore through scar tissue within the venous insertion site, but not within the heart. Conclusions The extraction of the LAP sensor lead was accomplished safely using standard techniques and equipment for percutaneously extracting pacemaker and defibrillator leads.
              Bookmark

              Author and article information

              Contributors
              Journal
              HeartRhythm Case Rep
              HeartRhythm Case Rep
              HeartRhythm Case Reports
              Elsevier
              2214-0271
              16 March 2016
              May 2016
              16 March 2016
              : 2
              : 3
              : 208-210
              Affiliations
              [* ]Division of Cardiac Electrophysiology, University of California–San Diego, La Jolla, California
              []Division of Cardiac Surgery, University of California–San Diego, La Jolla, California
              Author notes
              [* ] Address reprints and correspondence: Dr Shrinivas Hebsur, Division of Cardiac Electrophysiology, University of California–San Diego, 9444 Medical Center Dr MC 7411, 3rd Floor, Rm 3-089D, La Jolla, CA 92037Division of Cardiac Electrophysiology, University of California–San Diego shebsur@ 123456gmail.com
              Article
              S2214-0271(15)00263-8
              10.1016/j.hrcr.2015.11.007
              5419734
              71c48c33-e3d2-4be5-989f-a0e1ea013879
              © 2016 Heart Rhythm Society. Published by Elsevier Inc.

              This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).

              History
              Categories
              Case Report

              laser,extraction,cardiofit,defibrillator,extracardiac,surgical