      Evaluation of an electrochemical sensor for measuring blood ketones.

      Diabetes Care
      3-Hydroxybutyric Acid, blood, Blood Specimen Collection, methods, Diabetic Ketoacidosis, diagnosis, Electrochemistry, instrumentation, Humans, Ketone Bodies, Reference Values, Reproducibility of Results

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          To evaluate the performance of a hand-held ketone sensor that is able to measure blood beta-hydroxybutyrate (beta-HBA) concentrations within 30 s in patients with diabetic ketoacidosis (DKA) and patients who attend a weight management clinic. Two groups of patients were studied: 19 patients admitted with DKA and 156 patients attending a weight management clinic. Paired capillary and venous whole blood samples were measured using the ketone sensor and also using an enzymatic laboratory reference method. The ketone sensor accurately measured beta-HBA concentrations in patients with DKA (limits of agreement -0.9 to + 1.0 mmol/l) or starvation-induced ketonemia (limits of agreement -0.5 to +0.5 mmol/l). This ketone sensor accurately measures whole blood beta-HBA concentrations within 30 s.