      In vitro angiographic comparison of the flow-diversion performance of five neurovascular stents.

      Read this article at

      ScienceOpenPublisherPMC
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background and purpose Data differentiating flow diversion properties of commercially available low- and high-porosity stents are limited. This in vitro study applies angiographic analysis of intra-aneurysmal flow to compare the flow-diversion performance of five neurovascular devices in idealized sidewall and bifurcation aneurysm models. Methods Five commercial devices (Enterprise, Neuroform, LVIS, FRED, and Pipeline) were implanted in silicone sidewall and bifurcation aneurysm models under physiological average flow of blood analog fluid. High-speed angiographic images were acquired pre- and post-device implantation and contrast concentration-time curves within the aneurysm were recorded. The curves were quantified with five parameters to assess changes in contrast transport, and thus aneurysm hemodynamics, due to each device. Results Inter-device flow-diversion performance was more easily distinguished in the sidewall model than the bifurcation model. There were no obvious overall statistical trends in the bifurcation parameters but the Pipeline performed marginally better than the other devices. In the sidewall geometry, overall evidence suggests that the LVIS performed better than the Neuroform and Enterprise. The Pipeline and FRED devices were statistically superior to the three stents and Pipeline was superior to FRED in all sidewall parameters evaluated. Conclusions Based on this specific set of experiments, lower-porosity flow diverters perform significantly better in reducing intra-aneurysmal flow activity than higher-porosity stents in sidewall-type geometries. The LVIS device is potentially a better flow diverter than the Neuroform and Enterprise devices, while the Pipeline is potentially better than the FRED.

          Related collections

          Author and article information

          Journal
          Interv Neuroradiol
          Interventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciences
          SAGE Publications
          2385-2011
          1591-0199
          Apr 2018
          : 24
          : 2
          Affiliations
          [1 ] Department of Neurological Surgery, 12301 Stony Brook University , Stony Brook, NY, USA.
          Article
          10.1177/1591019917748317
          5847016
          29239685

          Comments

          Comment on this article