      Evaluating the forced oscillation technique in the detection of early smoking-induced respiratory changes

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          Early detection of the effects of smoking is of the utmost importance in the prevention of chronic obstructive pulmonary disease (COPD). The forced oscillation technique (FOT) is easy to perform since it requires only tidal breathing and offers a detailed approach to investigate the mechanical properties of the respiratory system. The FOT was recently suggested as an attractive alternative for diagnosing initial obstruction in COPD, which may be helpful in detecting COPD in its initial phases. Thus, the purpose of this study was twofold: (1) to evaluate the ability of FOT to detect early smoking-induced respiratory alterations; and (2) to compare the sensitivity of FOT with spirometry in a sample of low tobacco-dose subjects.

          Methods

          Results from a group of 28 smokers with a tobacco consumption of 11.2 ± 7.3 pack-years were compared with a control group formed by 28 healthy subjects using receiver operating characteristic (ROC) curves and a questionnaire as a gold standard. The early adverse effects of smoking were adequately detected by the absolute value of the respiratory impedance ( Z4Hz), the intercept resistance ( R0), and the respiratory system dynamic compliance ( Crs, dyn). Z4Hz was the most accurate parameter (Se = 75%, Sp = 75%), followed by R0 and Crs, dyn. The performances of the FOT parameters in the detection of the early effects of smoking were higher than that of spirometry (p < 0.05).

          Conclusion

          This study shows that FOT can be used to detect early smoking-induced respiratory changes while these pathologic changes are still potentially reversible. These findings support the use of FOT as a versatile clinical diagnostic tool in aiding COPD prevention and treatment.

          Related collections

          Most cited references47

          • Record: found
          • Abstract: found
          • Article: not found

          The meaning and use of the area under a receiver operating characteristic (ROC) curve.

          A representation and interpretation of the area under a receiver operating characteristic (ROC) curve obtained by the "rating" method, or by mathematical predictions based on patient characteristics, is presented. It is shown that in such a setting the area represents the probability that a randomly chosen diseased subject is (correctly) rated or ranked with greater suspicion than a randomly chosen non-diseased subject. Moreover, this probability of a correct ranking is the same quantity that is estimated by the already well-studied nonparametric Wilcoxon statistic. These two relationships are exploited to (a) provide rapid closed-form expressions for the approximate magnitude of the sampling variability, i.e., standard error that one uses to accompany the area under a smoothed ROC curve, (b) guide in determining the size of the sample required to provide a sufficiently reliable estimate of this area, and (c) determine how large sample sizes should be to ensure that one can statistically detect differences in the accuracy of diagnostic techniques.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Receiver-operating characteristic (ROC) plots: a fundamental evaluation tool in clinical medicine.

            The clinical performance of a laboratory test can be described in terms of diagnostic accuracy, or the ability to correctly classify subjects into clinically relevant subgroups. Diagnostic accuracy refers to the quality of the information provided by the classification device and should be distinguished from the usefulness, or actual practical value, of the information. Receiver-operating characteristic (ROC) plots provide a pure index of accuracy by demonstrating the limits of a test's ability to discriminate between alternative states of health over the complete spectrum of operating conditions. Furthermore, ROC plots occupy a central or unifying position in the process of assessing and using diagnostic tools. Once the plot is generated, a user can readily go on to many other activities such as performing quantitative ROC analysis and comparisons of tests, using likelihood ratio to revise the probability of disease in individual subjects, selecting decision thresholds, using logistic-regression analysis, using discriminant-function analysis, or incorporating the tool into a clinical strategy by using decision analysis.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Measuring the accuracy of diagnostic systems.

              J Swets (1988)
              Diagnostic systems of several kinds are used to distinguish between two classes of events, essentially "signals" and "noise". For them, analysis in terms of the "relative operating characteristic" of signal detection theory provides a precise and valid measure of diagnostic accuracy. It is the only measure available that is uninfluenced by decision biases and prior probabilities, and it places the performances of diverse systems on a common, easily interpreted scale. Representative values of this measure are reported here for systems in medical imaging, materials testing, weather forecasting, information retrieval, polygraph lie detection, and aptitude testing. Though the measure itself is sound, the values obtained from tests of diagnostic systems often require qualification because the test data on which they are based are of unsure quality. A common set of problems in testing is faced in all fields. How well these problems are handled, or can be handled in a given field, determines the degree of confidence that can be placed in a measured value of accuracy. Some fields fare much better than others.
                Bookmark

                Author and article information

                Journal
                Biomed Eng Online
                BioMedical Engineering OnLine
                BioMed Central
                1475-925X
                2009
                25 September 2009
                : 8
                : 22
                Affiliations
                [1 ]Department of Physiology, Biomedical Instrumentation Laboratory, State University of Rio de Janeiro, Rio de Janeiro, Brazil
                [2 ]Faculty of Medical Sciences, Pulmonary Function Laboratory, State University of Rio de Janeiro, Rio de Janeiro, Brazil
                [3 ]Department of Physiology, BioVasc Research Laboratory, State University of Rio de Janeiro, Rio de Janeiro, Brazil
                Article
                1475-925X-8-22
                10.1186/1475-925X-8-22
                2760559
                19781078
                7788ff51-f199-4a01-9e49-48b5e90c0499
                Copyright © 2009 Faria et al; licensee BioMed Central Ltd.

                This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

                History
                : 9 June 2009
                : 25 September 2009
                Categories
                Research

                Biomedical engineering
                Biomedical engineering

                Comments

                Comment on this article