      Oxyhemoglobin saturation measurements by green spectral shift.

      Optics letters
      Algorithms, Blood Chemical Analysis, methods, Computer Simulation, Humans, Models, Chemical, Oximetry, Oxyhemoglobins, analysis, Spectrum Analysis

      Read this article at

      ScienceOpenPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          From an analysis of new hemoglobin solution transmission spectra at various oxygen saturations (SO2), path lengths, and pH, we find the determination of SO2 by using the classical oximetry technique to be poorly calibrated. We used this data set to develop a proposed method for SO2 determination based on the spectral shift of the hemoglobin transmission minimum between 475 and 510 nm. The method does not require accurate knowledge of hemoglobin extinction coefficients and is linear in relation to SO2 despite changes in path length, pH, or hemoglobin concentration.