          Abstract

          Purpose

          To investigate the precision and accuracy of IOP measurements using a pneumatonometer and a tonometer tip cover (Tono-Pen ® tip cover) acting as a membrane between a cadaver eye model and pneumatonometer probe.

          Methods

          A total of 480 paired IOP measurements, with and without a Tono-Pen cover, were collected across 4 pressure levels of 7, 10, 20 and 30 mmHg. IOP measurements were obtained by three different pneumatonometer units paired with three different masked operators (three configurations). Four eyes were sampled for each eye pressure level. The sequence of eye pressure, configuration, and measurements with vs. without the Tono-Pen cover was randomized.

          Results

          With the Tono-Pen cover in place, there was a negative bias with a mean IOP difference of − 1.18 mmHg for all 480 paired samples compared with the measurements absent the cover. Compared with the test pressure settings (i.e., 7, 10, 20, 30 mmHg), the overall mean bias was + 0.35 mmHg with the Tono-Pen cover present. With the Tono-Pen cover present, the overall repeatability %CV (percent coefficient of variation) was 3.4% and the reproducibility %CV was 3.8% compared with a repeatability %CV of 3.2% and reproducibility %CV of 5.7% without the Tono-Pen cover.

          Conclusion

          Measurement of IOP via pneumatonometry with a Tono-Pen cover in place, also known as the excursion test method, yields precise, accurate and reproducible results. This developed method of pressure measurement is an acceptable and reliable form of IOP measurement.

          Related collections

          Most cited references6

          • Record: found
          • Abstract: found
          • Article: found
          Is Open Access

          Glaucoma is second leading cause of blindness globally

            Bookmark
            • Record: found
            • Abstract: found
            • Article: found
            Is Open Access

            Minimally-invasive glaucoma surgeries (MIGS) for open angle glaucoma: A systematic review and meta-analysis

            Background MIGS have been developed as a surgical alternative for glaucomatous patients. Purpose To analyze the change in intraocular pressure (IOP) and glaucoma medications using different MIGS devices (Trabectome, iStent, Excimer Laser Trabeculotomy (ELT), iStent Supra, CyPass, XEN, Hydrus, Fugo Blade, Ab interno canaloplasty, Goniscopy-assisted transluminal trabeculotomy) as a solo procedure or in association with phacoemulsification. Methods Randomized control trials (RCT) and non-RCT (non randomized comparative studies, NRS, and before-after studies) were included. Studies with at least one year of follow-up in patients affected by primary open angle glaucoma, pseudoexfoliative glaucoma or pigmentary glaucoma were considered. Risk of Bias assessment was performed using the Cochrane Risk of Bias and the ROBINS-I tools. The main outcome was the effect of MIGS devices compared to medical therapy, cataract surgery, other glaucoma surgeries and other MIGS on both IOP and use of glaucoma medications 12 months after surgery. Outcomes measures were the mean difference in the change of IOP and glaucoma medication compared to baseline at one and two years and all ocular adverse events. The current meta-analysis is registered on PROSPERO (reference n° CRD42016037280). Results Over a total of 3,069 studies, nine RCT and 21 case series with a total of 2.928 eyes were included. Main concerns about risk of bias in RCTs were lack of blinding, allocation concealment and attrition bias while in non-RCTs they were represented by patients’ selection, masking of participants and co-intervention management. Limited evidence was found based on both RCTs and non RCTs that compared MIGS surgery with medical therapy or other MIGS. In before-after series, MIGS surgery seemed effective in lowering both IOP and glaucoma drug use. MIGS showed a good safety profile: IOP spikes were the most frequent complications and no cases of infection or BCVA loss due to glaucoma were reported. Conclusions Although MIGS seem efficient in the reduction of the IOP and glaucoma medication and show good safety profile, this evidence is mainly derived from non-comparative studies and further, good quality RCTs are warranted.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Cerebrospinal fluid pressure is decreased in primary open-angle glaucoma.

              To compare cerebrospinal fluid (CSF) pressure in patients with primary open-angle glaucoma (POAG) with that in nonglaucomatous patients. Case-control study. Thirty-one thousand, seven hundred and eighty-six subjects underwent lumbar puncture (LP) between 1996 and 2007 at the Mayo Clinic, Rochester, Minnesota. Of these, 28 patients who had POAG and 49 patients who did not have POAG were analyzed. Retrospective review of medical records. Comparison of the 2 groups and factors associated with CSF pressure were analyzed by univariate and multivariate analyses. Demographics (age and gender), medical history, medication use, indication for LP, intraocular pressure (IOP), optic disc cup-to-disc ratio, visual field assessment, and CSF pressure. The mean CSF pressure +/- standard deviation was 13.0+/-4.2 mmHg in nonglaucoma patients and 9.2+/-2.9 mmHg in POAG patients (P<0.00005). The CSF pressure was lower in POAG patients regardless of indication for LP or age. Linear regression analysis showed that cup-to-disc ratio correlated independently with IOP (P<0.0001), CSF pressure (P<0.0001), and the translaminar pressure difference (P<0.0001). Multivariate analysis demonstrated that larger cup-to-disc ratio (P<0.0001) was associated with lower CSF pressure. Cerebrospinal fluid pressure is significantly lower in POAG patients compared with that in nonglaucomatous controls. These data support the notion that CSF pressure may play an important contributory role in the pathogenesis of POAG.
                Bookmark

                Author and article information

                Contributors
                tannerferg@gmail.com
                Journal
                Ophthalmol Ther
                Ophthalmol Ther
                Ophthalmology and Therapy
                Springer Healthcare (Cheshire )
                2193-8245
                2193-6528
                20 February 2020
                20 February 2020
                March 2020
                : 9
                : 1
                : 127-137
                Affiliations
                [1 ]GRID grid.239578.2, ISNI 0000 0001 0675 4725, Cole Eye Institute, , Cleveland Clinic, ; Cleveland, OH USA
                [2 ]GRID grid.66875.3a, ISNI 0000 0004 0459 167X, Department of Ophthalmology, , Mayo Clinic, ; Rochester, MN USA
                [3 ]GRID grid.478136.f, Vance Thompson Vision, ; Sioux Falls, SD USA
                [4 ]Value Creation Institute, Laguna Beach, CA USA
                Author information
                http://orcid.org/0000-0002-7754-0465
                Article
                235
                10.1007/s40123-020-00235-z
                7054586
                32078144
                78b6d565-2c56-4a37-b7a4-a43bcbb7eec7
                © The Author(s) 2020

                Open Access This article is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License, which permits any non-commercial use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc/4.0/

                History
                : 30 December 2019
                Categories
                Original Research
                Custom metadata
                © The Author(s) 2020

                diagnostics,intraocular pressure,open-angle glaucoma,ophthalmic techniques,tonometry