18
views
0
recommends
+1 Recommend
0 collections
    0
    shares
      • Record: found
      • Abstract: found
      • Article: found
      Is Open Access

      Estimating physical activity and sedentary behaviour in a free-living environment: A comparative study between Fitbit Charge 2 and Actigraph GT3X

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          Activity trackers such as the Fitbit Charge 2 enable users and researchers to monitor physical activity in daily life, which could be beneficial for changing behaviour. However, the accuracy of the Fitbit Charge 2 in a free-living environment is largely unknown.

          Objective

          To investigate the agreement between Fitbit Charge 2 and ActiGraph GT3X for the estimation of steps, energy expenditure, time in sedentary behaviour, and light and moderate-to-vigorous physical activity under free-living conditions, and further examine to what extent placing the ActiGraph on the wrist as opposed to the hip would affect the findings.

          Methods

          41 adults (n = 10 males, n = 31 females) were asked to wear a Fitbit Charge 2 device and two ActiGraph GT3X devices (one on the hip and one on the wrist) for seven consecutive days and fill out a log of wear times. Agreement was assessed through Bland-Altman plots combined with multilevel analysis.

          Results

          The Fitbit measured 1,492 steps/day more than the hip-worn ActiGraph (limits of agreement [LoA] = -2,250; 5,234), while for sedentary time, it measured 25 min/day less (LoA = -137; 87). Both Bland-Altman plots showed fixed bias. For time in light physical activity, the Fitbit measured 59 min/day more (LoA = -52;169). For time in moderate-to-vigorous physical activity, the Fitbit measured 31 min/day less (LoA = -132; 71) and for activity energy expenditure it measured 408 kcal/day more than the hip-worn ActiGraph (LoA = -385; 1,200). For the two latter outputs, the plots indicated proportional bias. Similar or more pronounced discrepancies, mostly in opposite direction, appeared when comparing to the wrist-worn ActiGraph.

          Conclusion

          Moderate to substantial differences between devices were found for most outputs, which could be due to differences in algorithms. Caution should be taken if replacing one device with another and when comparing results.

          Related collections

          Most cited references37

          • Record: found
          • Abstract: found
          • Article: found
          Is Open Access

          Accuracy of Fitbit Devices: Systematic Review and Narrative Syntheses of Quantitative Data

          Background Although designed as a consumer product to help motivate individuals to be physically active, Fitbit activity trackers are becoming increasingly popular as measurement tools in physical activity and health promotion research and are also commonly used to inform health care decisions. Objective The objective of this review was to systematically evaluate and report measurement accuracy for Fitbit activity trackers in controlled and free-living settings. Methods We conducted electronic searches using PubMed, EMBASE, CINAHL, and SPORTDiscus databases with a supplementary Google Scholar search. We considered original research published in English comparing Fitbit versus a reference- or research-standard criterion in healthy adults and those living with any health condition or disability. We assessed risk of bias using a modification of the Consensus-Based Standards for the Selection of Health Status Measurement Instruments. We explored measurement accuracy for steps, energy expenditure, sleep, time in activity, and distance using group percentage differences as the common rubric for error comparisons. We conducted descriptive analyses for frequency of accuracy comparisons within a ±3% error in controlled and ±10% error in free-living settings and assessed for potential bias of over- or underestimation. We secondarily explored how variations in body placement, ambulation speed, or type of activity influenced accuracy. Results We included 67 studies. Consistent evidence indicated that Fitbit devices were likely to meet acceptable accuracy for step count approximately half the time, with a tendency to underestimate steps in controlled testing and overestimate steps in free-living settings. Findings also suggested a greater tendency to provide accurate measures for steps during normal or self-paced walking with torso placement, during jogging with wrist placement, and during slow or very slow walking with ankle placement in adults with no mobility limitations. Consistent evidence indicated that Fitbit devices were unlikely to provide accurate measures for energy expenditure in any testing condition. Evidence from a few studies also suggested that, compared with research-grade accelerometers, Fitbit devices may provide similar measures for time in bed and time sleeping, while likely markedly overestimating time spent in higher-intensity activities and underestimating distance during faster-paced ambulation. However, further accuracy studies are warranted. Our point estimations for mean or median percentage error gave equal weighting to all accuracy comparisons, possibly misrepresenting the true point estimate for measurement bias for some of the testing conditions we examined. Conclusions Other than for measures of steps in adults with no limitations in mobility, discretion should be used when considering the use of Fitbit devices as an outcome measurement tool in research or to inform health care decisions, as there are seemingly a limited number of situations where the device is likely to provide accurate measurement.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: found
            Is Open Access

            Reliability and validity of ten consumer activity trackers

            Background Activity trackers can potentially stimulate users to increase their physical activity behavior. The aim of this study was to examine the reliability and validity of ten consumer activity trackers for measuring step count in both laboratory and free-living conditions. Method Healthy adult volunteers (n = 33) walked twice on a treadmill (4.8 km/h) for 30 min while wearing ten different activity trackers (i.e. Lumoback, Fitbit Flex, Jawbone Up, Nike+ Fuelband SE, Misfit Shine, Withings Pulse, Fitbit Zip, Omron HJ-203, Yamax Digiwalker SW-200 and Moves mobile application). In free-living conditions, 56 volunteers wore the same activity trackers for one working day. Test-retest reliability was analyzed with the Intraclass Correlation Coefficient (ICC). Validity was evaluated by comparing each tracker with the gold standard (Optogait system for laboratory and ActivPAL for free-living conditions), using paired samples t-tests, mean absolute percentage errors, correlations and Bland-Altman plots. Results Test-retest analysis revealed high reliability for most trackers except for the Omron (ICC .14), Moves app (ICC .37) and Nike+ Fuelband (ICC .53). The mean absolute percentage errors of the trackers in laboratory and free-living conditions respectively, were: Lumoback (−0.2, −0.4), Fibit Flex (−5.7, 3.7), Jawbone Up (−1.0, 1.4), Nike+ Fuelband (−18, −24), Misfit Shine (0.2, 1.1), Withings Pulse (−0.5, −7.9), Fitbit Zip (−0.3, 1.2), Omron (2.5, −0.4), Digiwalker (−1.2, −5.9), and Moves app (9.6, −37.6). Bland-Altman plots demonstrated that the limits of agreement varied from 46 steps (Fitbit Zip) to 2422 steps (Nike+ Fuelband) in the laboratory condition, and 866 steps (Fitbit Zip) to 5150 steps (Moves app) in the free-living condition. Conclusion The reliability and validity of most trackers for measuring step count is good. The Fitbit Zip is the most valid whereas the reliability and validity of the Nike+ Fuelband is low.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: found
              Is Open Access

              Accuracy of Consumer Wearable Heart Rate Measurement During an Ecologically Valid 24-Hour Period: Intraindividual Validation Study

              Background Wrist-worn smart watches and fitness monitors (ie, wearables) have become widely adopted by consumers and are gaining increased attention from researchers for their potential contribution to naturalistic digital measurement of health in a scalable, mobile, and unobtrusive way. Various studies have examined the accuracy of these devices in controlled laboratory settings (eg, treadmill and stationary bike); however, no studies have investigated the heart rate accuracy of wearables during a continuous and ecologically valid 24-hour period of actual consumer device use conditions. Objective The aim of this study was to determine the heart rate accuracy of 2 popular wearable devices, the Apple Watch 3 and Fitbit Charge 2, as compared with the gold standard reference method, an ambulatory electrocardiogram (ECG), during consumer device use conditions in an individual. Data were collected across 5 daily conditions, including sitting, walking, running, activities of daily living (ADL; eg, chores, brushing teeth), and sleeping. Methods One participant, (first author; 29-year-old Caucasian male) completed a 24-hour ecologically valid protocol by wearing 2 popular wrist wearable devices (Apple Watch 3 and Fitbit Charge 2). In addition, an ambulatory ECG (Vrije Universiteit Ambulatory Monitoring System) was used as the gold standard reference method, which resulted in the collection of 102,740 individual heartbeats. A single-subject design was used to keep all variables constant except for wearable devices while providing a rapid response design to provide initial assessment of wearable accuracy for allowing the research cycle to keep pace with technological advancements. Accuracy of these devices compared with the gold standard ECG was assessed using mean error, mean absolute error, and mean absolute percent error. These data were supplemented with Bland-Altman analyses and concordance class correlation to assess agreement between devices. Results The Apple Watch 3 and Fitbit Charge 2 were generally highly accurate across the 24-hour condition. Specifically, the Apple Watch 3 had a mean difference of −1.80 beats per minute (bpm), a mean absolute error percent of 5.86%, and a mean agreement of 95% when compared with the ECG across 24 hours. The Fitbit Charge 2 had a mean difference of −3.47 bpm, a mean absolute error of 5.96%, and a mean agreement of 91% when compared with the ECG across 24 hours. These findings varied by condition. Conclusions The Apple Watch 3 and the Fitbit Charge 2 provided acceptable heart rate accuracy (<±10%) across the 24 hour and during each activity, except for the Apple Watch 3 during the daily activities condition. Overall, these findings provide preliminary support that these devices appear to be useful for implementing ambulatory measurement of cardiac activity in research studies, especially those where the specific advantages of these methods (eg, scalability, low participant burden) are particularly suited to the population or research question.
                Bookmark

                Author and article information

                Contributors
                Role: ConceptualizationRole: Data curationRole: Formal analysisRole: MethodologyRole: Writing – original draft
                Role: Formal analysisRole: MethodologyRole: Writing – review & editing
                Role: Formal analysisRole: Writing – review & editing
                Role: Formal analysisRole: Writing – review & editing
                Role: Writing – review & editing
                Role: Writing – review & editing
                Role: Writing – review & editing
                Role: Writing – review & editing
                Role: ConceptualizationRole: Writing – review & editing
                Role: MethodologyRole: Writing – review & editing
                Role: Editor
                Journal
                PLoS One
                PLoS ONE
                plos
                plosone
                PLoS ONE
                Public Library of Science (San Francisco, CA USA )
                1932-6203
                11 June 2020
                2020
                : 15
                : 6
                : e0234426
                Affiliations
                [1 ] Research Unit for Dietary Studies, The Parker Institute, Bispebjerg and Frederiksberg Hospital, The Capital Region, København, Denmark
                [2 ] Department of Public Health, University of Southern Denmark, Esbjerg, Denmark
                [3 ] Biomathematics & Statistics Scotland (James Hutton Institute), Aberdeen, Scotland, United Kingdom
                [4 ] School of Psychology, Faculty of Medicine and Health, University of Leeds, Leeds, England, United Kingdom
                [5 ] Centro Interdisciplinar para o Estudo da Performance Humana, Faculdade de Motricidade Humana, Universidade de Lisboa, Lisbon, Portugal
                [6 ] The National Institute of Public Health, University of Southern Denmark, Copenhagen, Denmark
                [7 ] The Boden Institute of Obesity, Nutrition, Exercise & Eating Disorders, The University of Sydney, Sydney, Australia
                [8 ] Department of Public Health, Section for General Practice, University of Copenhagen, Copenhagen, Denmark
                Linneaus University, SWEDEN
                Author notes

                Competing Interests: The authors have declared that no competing interests exist.

                Author information
                http://orcid.org/0000-0003-4921-6593
                Article
                PONE-D-19-24901
                10.1371/journal.pone.0234426
                7289355
                32525912
                7e0fce74-6125-492b-96b7-f0955e42c625
                © 2020 Mikkelsen et al

                This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

                History
                : 4 September 2019
                : 25 May 2020
                Page count
                Figures: 1, Tables: 2, Pages: 13
                Funding
                Funded by: funder-id http://dx.doi.org/10.13039/501100007601, Horizon 2020;
                Award ID: 643309
                Award Recipient :
                Funded by: funder-id http://dx.doi.org/10.13039/100001275, Oak Foundation;
                Award ID: OCAY-18-774-OFIL
                The NoHoW study has received funding from the European Union’s Horizon 2020 research and innovation programme under grant agreement No. 643309. The Parker Institute is supported by a core grant from the Oak Foundation (grant agreement number: OCAY-18-774-OFIL). The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
                Categories
                Research Article
                Biology and Life Sciences
                Anatomy
                Musculoskeletal System
                Body Limbs
                Arms
                Wrist
                Medicine and Health Sciences
                Anatomy
                Musculoskeletal System
                Body Limbs
                Arms
                Wrist
                Biology and Life Sciences
                Anatomy
                Musculoskeletal System
                Pelvis
                Hip
                Medicine and Health Sciences
                Anatomy
                Musculoskeletal System
                Pelvis
                Hip
                Medicine and Health Sciences
                Public and Occupational Health
                Physical Activity
                Engineering and Technology
                Electronics
                Accelerometers
                Biology and Life Sciences
                Biochemistry
                Bioenergetics
                Medicine and Health Sciences
                Public and Occupational Health
                Physical Activity
                Physical Fitness
                Exercise
                Medicine and Health Sciences
                Sports and Exercise Medicine
                Exercise
                Biology and Life Sciences
                Sports Science
                Sports and Exercise Medicine
                Exercise
                Medicine and Health Sciences
                Cardiology
                Heart Rate
                Research and Analysis Methods
                Research Facilities
                Research Laboratories
                Custom metadata
                There are legal restrictions on sharing data that contain potentially identifying or sensitive person information. The restrictions are imposed by The Danish Data Protection Agency ( https://www.datatilsynet.dk/english/). Data used in the current study will be made available upon request after application to the NoHoW data controller. The application procedure can be obtained from the Research Unit for Dietary Studies at The Parker Institute ( bfh-eek@ 123456regionh.dk ) or from the NoHoW data controller Graham Horgan ( Graham.Horgan@ 123456bioss.ac.uk ).

                Uncategorized
                Uncategorized

                Comments

                Comment on this article