      In vivo human crystalline lens topography

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Custom high-resolution high-speed anterior segment spectral domain optical coherence tomography (OCT) was used to characterize three-dimensionally (3-D) the human crystalline lens in vivo. The system was provided with custom algorithms for denoising and segmentation of the images, as well as for fan (scanning) and optical (refraction) distortion correction, to provide fully quantitative images of the anterior and posterior crystalline lens surfaces. The method was tested on an artificial eye with known surfaces geometry and on a human lens in vitro, and demonstrated on three human lenses in vivo . Not correcting for distortion overestimated the anterior lens radius by 25% and the posterior lens radius by more than 65%. In vivo lens surfaces were fitted by biconicoids and Zernike polynomials after distortion correction. The anterior lens radii of curvature ranged from 10.27 to 14.14 mm, and the posterior lens radii of curvature ranged from 6.12 to 7.54 mm. Surface asphericities ranged from −0.04 to −1.96. The lens surfaces were well fitted by quadrics (with variation smaller than 2%, for 5-mm pupils), with low amounts of high order terms. Surface lens astigmatism was significant, with the anterior lens typically showing horizontal astigmatism ( Z 2 2 ranging from −11 to −1 µm) and the posterior lens showing vertical astigmatism ( Z 2 2 ranging from 6 to 10 µm).

          Related collections

          Most cited references62

          • Record: found
          • Abstract: found
          • Article: not found

          Optical coherence tomography.

          A technique called optical coherence tomography (OCT) has been developed for noninvasive cross-sectional imaging in biological systems. OCT uses low-coherence interferometry to produce a two-dimensional image of optical scattering from internal tissue microstructures in a way that is analogous to ultrasonic pulse-echo imaging. OCT has longitudinal and lateral spatial resolutions of a few micrometers and can detect reflected signals as small as approximately 10(-10) of the incident optical power. Tomographic imaging is demonstrated in vitro in the peripapillary area of the retina and in the coronary artery, two clinically relevant examples that are representative of transparent and turbid media, respectively.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Biometric, optical and physical changes in the isolated human crystalline lens with age in relation to presbyopia.

            The biometric, optical and physical properties of 19 pairs of isolated human eye-bank lenses ranging in age from 5 to 96 years were compared. Lens focal length and spherical aberration were measured using a scanning laser apparatus, lens thickness and the lens surface curvatures were measured by digitizing the lens profiles and equivalent refractive indices were calculated for each lens using this data. The second lens from each donor was used to measure resistance to physical deformation by providing a compressive force to the lens. The lens capsule was then removed from each lens and each measurement was repeated to ascertain what role the capsule plays in determining these optical and physical characteristics. Age dependent changes in lens focal length, lens surface curvatures and lens resistance to physical deformation are described. Isolated lens focal length was found to be significantly linearly correlated with both the anterior and posterior surface curvatures. No age dependent change in equivalent refractive index of the isolated lens was found. Although decapsulating human lenses causes similar changes in focal length to that which we have shown to occur when human lenses are mechanically stretched into an unaccommodated state, the effects are due to nonsystematic changes in lens curvatures. These studies reinforce the conclusion that lens hardening must be considered as an important factor in the development of presbyopia, that age changes in the human lens are not limited to the loss of accommodation that characterizes presbyopia but that the lens optical and physical properties change substantially with age in a complex manner.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              The shape of the aging human lens: curvature, equivalent refractive index and the lens paradox.

              Scheimpflug slit images of the crystalline lens are distorted due to the refracting properties of the cornea and because they are obliquely viewed. We measured the aspheric curvature of the lens of 102 subjects ranging in age between 16 and 65 years and applied correction for these distortions. The procedure was validated by measuring an artificial eye and pseudophakic patients with intraocular lenses of known dimensions. Compared to previous studies using Scheimpflug photography, the decrease of the radius of the anterior lens surface with age was smaller, and the absolute value for the radius of the anterior and posterior lens surface was significantly smaller. A slight decrease of the posterior lens radius with age could be demonstrated. Generally, front and back surfaces were hyperbolic. Axial length was measured of 42 subjects enabling calculation of the equivalent refractive index of the lens, which showed a small, but highly significant decrease with age. These new findings explain the lens paradox and may serve as a basis for modelling the refractive properties of the lens.
                Bookmark

                Author and article information

                Journal
                Biomed Opt Express
                Biomed Opt Express
                BOE
                Biomedical Optics Express
                Optical Society of America
                2156-7085
                12 September 2012
                01 October 2012
                12 September 2012
                : 3
                : 10
                : 2471-2488
                Affiliations
                [1 ]Instituto de Óptica “Daza de Valdés”, Consejo Superior de Investigaciones Científicas, C/Serrano 121, 28006 Madrid, Spain
                Author notes
                Article
                170171
                10.1364/BOE.3.002471
                3470006
                23082289
                7f745e2d-e4e5-41b4-8a43-c16e7e51abdf
                ©2012 Optical Society of America

                This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-No Derivative Works 3.0 Unported License, which permits download and redistribution, provided that the original work is properly cited. This license restricts the article from being modified or used commercially.

                History
                : 7 June 2012
                : 26 July 2012
                : 24 August 2012
                Funding
                Funded by: MICINN
                Award ID: FIS2008-02065
                Award ID: FIS2011-25637
                Funded by: EURHORCs-ESF
                Award ID: EURYI-05-102-ES
                Award ID: ERC-2011-AdG-294099
                Award ID: CEN-2091021
                Categories
                Ophthalmology Applications
                Custom metadata
                True
                0

                Vision sciences
                (110.4500) optical coherence tomography,(110.6880) three-dimensional image acquisition,(120.4640) optical instruments,(330.7327) visual optics, ophthalmic instrumentation,(120.4800) optical standards and testing,(120.6650) surface measurements, figure

                Comments

                Comment on this article