      The Effect of Scan Length on the Assessment of BOLD Delay in Ischemic Stroke

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Objectives: To evaluate the impact of resting-state functional MRI scan length on the diagnostic accuracy, image quality and lesion volume estimation of BOLD delay maps used for brain perfusion assessment in acute ischemic stroke.

          Methods: Sixty-three acute ischemic stroke patients received a 340 s resting-state functional MRI within 24 h of stroke symptom onset. BOLD delay maps were calculated from the full scan and four shortened versions (68 s, 136 s, 204 s, 272 s). The BOLD delay lesions on these maps were compared in terms of spatial overlap and volumetric agreement with the lesions derived from the full scans and with time-to-maximum (Tmax) lesions derived from DSC-MRI in a subset of patients (n = 10). In addition, the interpretability and quality of these maps were compared across different scan lengths using mixed models.

          Results: Shortened BOLD delay scans showed a small volumetric bias (ranging from 0.05 to 5.3 mL; between a 0.13% volumetric underestimation and a 7.7% overestimation relative to the mean of the volumes, depending on scan length) compared to the full scan. Decreased scan length was associated with decreased spatial overlap with both the BOLD delay lesions derived from the full scans and with Tmax lesions. Only the two shortest scan lengths (68 and 136 s) were associated with substantially decreased interpretability, decreased structure clarity, and increased noisiness of BOLD delay maps.

          Conclusions: BOLD delay maps derived from resting-state fMRI scans lasting 272 and 204 s provide sufficient diagnostic quality and adequate assessment of perfusion lesion volumes. Such shortened scans may be helpful in situations where quick clinical decisions need to be made.

          Related collections

          Most cited references46

          • Record: found
          • Abstract: found
          • Article: not found

          Tracer arrival timing-insensitive technique for estimating flow in MR perfusion-weighted imaging using singular value decomposition with a block-circulant deconvolution matrix.

          Relative cerebral blood flow (CBF) and tissue mean transit time (MTT) estimates from bolus-tracking MR perfusion-weighted imaging (PWI) have been shown to be sensitive to delay and dispersion when using singular value decomposition (SVD) with a single measured arterial input function. This study proposes a technique that is made time-shift insensitive by the use of a block-circulant matrix for deconvolution with (oSVD) and without (cSVD) minimization of oscillation of the derived residue function. The performances of these methods are compared with standard SVD (sSVD) in both numerical simulations and in clinically acquired data. An additional index of disturbed hemodynamics (oDelay) is proposed that represents the tracer arrival time difference between the AIF and tissue signal. Results show that PWI estimates from sSVD are weighted by tracer arrival time differences, while those from oSVD and cSVD are not. oSVD also provides estimates that are less sensitive to blood volume compared to cSVD. Using PWI data that can be routinely collected clinically, oSVD shows promise in providing tracer arrival timing-insensitive flow estimates and hence a more specific indicator of ischemic injury. Shift maps can continue to provide a sensitive reflection of disturbed hemodynamics. Copyright 2003 Wiley-Liss, Inc.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Optimal Tmax threshold for predicting penumbral tissue in acute stroke.

            We sought to assess whether the volume of the ischemic penumbra can be estimated more accurately by altering the threshold selected for defining perfusion-weighting imaging (PWI) lesions. DEFUSE is a multicenter study in which consecutive acute stroke patients were treated with intravenous tissue-type plasminogen activator 3 to 6 hours after stroke onset. Magnetic resonance imaging scans were obtained before, 3 to 6 hours after, and 30 days after treatment. Baseline and posttreatment PWI volumes were defined according to increasing Tmax delay thresholds (>2, >4, >6, and >8 seconds). Penumbra salvage was defined as the difference between the baseline PWI lesion and the final infarct volume (30-day fluid-attenuated inversion recovery sequence). We hypothesized that the optimal PWI threshold would provide the strongest correlations between penumbra salvage volumes and various clinical and imaging-based outcomes. Thirty-three patients met the inclusion criteria. The correlation between infarct growth and penumbra salvage volume was significantly better for PWI lesions defined by Tmax >6 seconds versus Tmax >2 seconds, as was the difference in median penumbra salvage volume in patients with a favorable versus an unfavorable clinical response. Among patients who did not experience early reperfusion, the Tmax >4 seconds threshold provided a more accurate prediction of final infarct volume than the >2 seconds threshold. Defining PWI lesions based on a stricter Tmax threshold than the standard >2 seconds delay appears to provide more a reliable estimate of the volume of the ischemic penumbra in stroke patients imaged between 3 and 6 hours after symptom onset. A threshold between 4 and 6 seconds appears optimal for early identification of critically hypoperfused tissue.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: found
              Is Open Access

              Real-time motion analytics during brain MRI improve data quality and reduce costs

              Head motion systematically distorts clinical and research MRI data. Motion artifacts have biased findings from many structural and functional brain MRI studies. An effective way to remove motion artifacts is to exclude MRI data frames affected by head motion. However, such post-hoc frame censoring can lead to data loss rates of 50% or more in our pediatric patient cohorts. Hence, many scanner operators collect additional ‘buffer data’, an expensive practice that, by itself, does not guarantee sufficient high-quality MRI data for a given participant. Therefore, we developed an easy-to-setup, easy-to-use Framewise Integrated Real-time MRI Monitoring (FIRMM) software suite that provides scanner operators with head motion analytics in real-time, allowing them to scan each subject until the desired amount of low-movement data has been collected. Our analyses show that using FIRMM to identify the ideal scan time for each person can reduce total brain MRI scan times and associated costs by 50% or more.
                Bookmark

                Author and article information

                Contributors
                Journal
                Front Neurol
                Front Neurol
                Front. Neurol.
                Frontiers in Neurology
                Frontiers Media S.A.
                1664-2295
                05 May 2020
                2020
                : 11
                : 381
                Affiliations
                [1] 1Center for Stroke Research, Charité - Universitätsmedizin Berlin , Berlin, Germany
                [2] 2Institute of Biometry and Clinical Epidemiology, Charité – Universitätsmedizin Berlin , Berlin, Germany
                [3] 3Berlin Institute of Health (BIH) , Berlin, Germany
                [4] 4Department of Neurophysics, Max Planck Institute for Human Cognitive and Brain Sciences , Leipzig, Germany
                [5] 5Center for Cognitive Neuroscience Berlin, Free University , Berlin, Germany
                [6] 6Berlin School of Mind and Brain, Humboldt Universität zu Berlin , Berlin, Germany
                [7] 7Department of Neurology, Max Planck Institute for Human Cognitive and Brain Sciences , Leipzig, Germany
                Author notes

                Edited by: Peter Sörös, University of Oldenburg, Germany

                Reviewed by: Bradley J. MacIntosh, Sunnybrook Research Institute (SRI), Canada; Seena Dehkharghani, New York University, United States

                *Correspondence: Ahmed A. Khalil ahmed-abdelrahim.khalil@ 123456charite.de

                This article was submitted to Applied Neuroimaging, a section of the journal Frontiers in Neurology

                Article
                10.3389/fneur.2020.00381
                7214917
                32431665
                80c4bf20-722a-42ba-952f-4ec18d0f2aeb
                Copyright © 2020 Tanrıtanır, Villringer, Galinovic, Grittner, Kirilina, Fiebach, Villringer and Khalil.

                This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

                History
                : 05 December 2019
                : 15 April 2020
                Page count
                Figures: 6, Tables: 3, Equations: 0, References: 66, Pages: 12, Words: 8716
                Funding
                Funded by: Bundesministerium für Bildung und Forschung 10.13039/501100002347
                Categories
                Neurology
                Original Research

                Neurology
                perfusion,acute stroke,bold delay,scan length,mri
                Neurology
                perfusion, acute stroke, bold delay, scan length, mri

                Comments

                Comment on this article