      Establishing New Cut-Off Limits for Galactose 1-Phosphate-Uridyltransferase Deficiency for the Dutch Newborn Screening Programme

      research-article

      Read this article at

      ScienceOpenPublisherPMC
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Newborn screening for classical galactosemia in the Netherlands is performed by five laboratories and is based on the measurement of galactose 1-phosphate-uridyltransferase (GALT) activity and total galactose (TGAL) in heel prick blood spots. Unexpected problems with the GALT assay posed a challenge to switch to a new assay. The aim of this study was to make an analytical and clinical evaluation of GALT assays to replace the current assay and to establish new cut-off values (COVs).

          First, the manual assay from PerkinElmer (NG-1100) and the GSP assay were compared by analyzing 626 anonymous heel prick samples in parallel. Secondly, a manual GSP method was evaluated and 2,052 samples were compared with the automated GSP assay. Finally, a clinical evaluation was performed by collecting data from 93 referred newborns.

          No satisfactory correlation was observed between GALT activity measured with the manual NG-1100 assay and the automated GSP assay. An acceptable correlation was found between the manual and automated GSP assay. Intra- and inter-assay variation of the automated GSP were 1.8–10.0% and 3.1–13.9%, respectively. Evaluation of clinical data demonstrated that adjusting the COVs for GALT to 2.0 U/dl and TGAL to 1,100 μmol/l improved specificity of screening for classical galactosemia.

          An assay designed for automated processing to measure GALT activity in heel prick samples works equally well when processed manually. We therefore adopted both methods in the Dutch screening laboratories. As a result of this evaluation new COVs for GALT and TGAL have been introduced and are valid from July 2015.

          Related collections

          Author and article information

          Contributors
          ekemper@ysl.nl
          Journal
          JIMD Rep
          JIMD Rep
          JIMD Reports
          Springer Berlin Heidelberg (Berlin, Heidelberg )
          2192-8304
          2192-8312
          21 May 2016
          2017
          : 32
          : 1-6
          Affiliations
          [1 ] ISNI 0000 0004 0501 4532, GRID grid.414559.8, Department of Clinical Chemistry, , IJsselland Hospital, ; Prins Constantijnweg 2, 2906ZC Capelle aan den IJssel, The Netherlands
          [2 ] ISNI 0000000404654431, GRID grid.5650.6, Newborn Screening Laboratory Amsterdam, , Academic Medical Center, ; Amsterdam, The Netherlands
          [3 ] ISNI 0000000404654431, GRID grid.5650.6, Department of Pediatrics, , Emma Children’s Hospital, Academic Medical Center, ; Amsterdam, The Netherlands
          [4 ] ISNI 0000 0001 2208 0118, GRID grid.31147.30, , Centre for Infectious Diseases Research, Diagnostics and Screening, National Institute for Public Health and the Environment, ; Bilthoven, The Netherlands
          [5 ]Swiss Newborn Screening Laboratory, Children’s Research Center, Zurich, Switzerland
          [11 ] ISNI 0000 0001 0726 4330, GRID grid.412341.1, Division of Metabolism, , University Children’s Hospital, ; Zurich, Switzerland
          [12 ] ISNI 0000 0004 0459 167X, GRID grid.66875.3a, Div of Child and Adolescent Neurology, , Mayo Clinic, ; Rochester, Minnesota USA
          [13 ] ISNI 0000000121901201, GRID grid.83440.3b, Health, Genetics & Genomic Med Clinical and Molecular Genetics Unit, , Univ Coll London Inst of Child, ; London, UK
          [14 ] ISNI 0000 0001 0328 4908, GRID grid.5253.1, Managing Editor JIMD Univ Children’s Hospital Heidelberg, , Univ Children’s Hospital Heidelberg, ; Heidelberg, Germany
          [15 ] ISNI 0000 0001 2217 8588, GRID grid.265219.b, Hayward Genetics Ctr,SL#31, , Tulane Univ Medical Health Ctr, ; New Orleans, Louisiana USA
          [16 ] ISNI 0000 0000 8853 2677, GRID grid.5361.1, Division of Human Genetics, , Medical University Innsbruck, ; Innsbruck, Austria
          Author notes

          Communicated by: Piero Rinaldo, MD, PhD

          Article
          PMC5362554 PMC5362554 5362554 560
          10.1007/8904_2016_560
          5362554
          27207469
          81045712-66c8-41a0-ba03-9ab1ffb31f71
          © SSIEM and Springer-Verlag Berlin Heidelberg 2016
          History
          : 04 February 2016
          : 10 February 2016
          : 19 March 2016
          Categories
          Article
          Custom metadata
          © SSIEM and Springer-Verlag Berlin Heidelberg 2017

          Galactose 1-phosphate-uridyltransferase,Galactosemia,Heel prick,Newborn screening,Blood spots

          Comments

          Comment on this article