      Intrafraction displacement of prone versus supine prostate positioning monitored by real‐time electromagnetic tracking

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Implanted radiofrequency transponders were used for real‐time monitoring of the intrafraction prostate displacement between patients in the prone position and the same patients in the supine position. Thirteen patients had three transponders implanted transperineally and were treated prone with a custom‐fitted thermoplastic immobilization device. After collecting data from the last fraction, patients were realigned in the supine position and the displacements of the transponders were monitored for 5–7 minutes. Fourier transforms were applied to the data from each patient to determine periodicity and its amplitude. To remove auto correlation from the stream of displacement data, the distribution of short‐term and long‐term velocity components were calculated from Poincaré plots of paired sequential vector displacements. The mean absolute displacement was significantly greater prone than supine in the superior–inferior (SI) plane ( 1.2 ± 0.6 mm vs. 0.6 ± 0.4 mm , p = 0.015 ), but not for the lateral or anterior–posterior (AP) planes. Displacements were least in the lateral direction. Fourier analyses showed the amplitude of respiratory oscillations was much greater for the SI and AP planes in the prone versus the supine position. Analysis of Poincaré plots confirmed greater short‐term variance in the prone position, but no difference in the long‐term variance. The centroid of the implanted transponders was offset from the treatment isocenter by > 5 mm for 1.9% of the time versus 0.8% of the time for supine. These results confirmed significantly greater net intrafraction prostate displacement of patients in the prone position than in the supine position, but most of the difference was due to respiration‐induced motion that was most pronounced in the SI and AP directions. Because the respiratory motion remained within the action threshold and also within our 5 mm treatment planning margins, there is no compelling reason to choose one treatment position over the other.

          PACS number: 87.50.st

          Related collections

          Most cited references16

          • Record: found
          • Abstract: found
          • Article: not found

          Multi-institutional clinical experience with the Calypso System in localization and continuous, real-time monitoring of the prostate gland during external radiotherapy.

          To report the clinical experience with an electromagnetic treatment target positioning and continuous monitoring system in patients with localized prostate cancer receiving external beam radiotherapy. The Calypso System is a target positioning device that continuously monitors the location of three implanted electromagnetic transponders at a rate of 10 Hz. The system was used at five centers to position 41 patients over a full course of therapy. Electromagnetic positioning was compared to setup using skin marks and to stereoscopic X-ray localization of the transponders. Continuous monitoring was performed in 35 patients. The difference between skin mark vs. the Calypso System alignment was found to be >5 mm in vector length in more than 75% of fractions. Comparisons between the Calypso System and X-ray localization showed good agreement. Qualitatively, the continuous motion was unpredictable and varied from persistent drift to transient rapid movements. Displacements > or =3 and > or =5 mm for cumulative durations of at least 30 s were observed during 41% and 15% of sessions. In individual patients, the number of fractions with displacements > or =3 mm ranged from 3% to 87%; whereas the number of fractions with displacements > or =5 mm ranged from 0% to 56%. The Calypso System is a clinically efficient and objective localization method for positioning prostate patients undergoing radiotherapy. Initial treatment setup can be performed rapidly, accurately, and objectively before radiation delivery. The extent and frequency of prostate motion during radiotherapy delivery can be easily monitored and used for motion management.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Target localization and real-time tracking using the Calypso 4D localization system in patients with localized prostate cancer.

            The Calypso 4D Localization System is being developed to provide accurate, precise, objective, and continuous target localization during radiotherapy. This study involves the first human use of the system, to evaluate the localization accuracy of this technique compared with radiographic localization and to assess its ability to obtain real-time prostate-motion information. Three transponders were implanted in each of 20 patients. Eleven eligible patients of the 20 patients participated in a study arm that compared radiographic triangulated transponder locations to electromagnetically recorded transponder locations. Transponders were tracked for 8-min periods. The implantations were all successful, with no major complications. Intertransponder distances were largely stable. Comparison of the patient localization on the basis of transponder locations as per the Calypso system with the radiographic transponder localization showed an average (+/-SD) 3D difference of 1.5 +/- 0.9 mm. Upon tracking during 8 min, 2 of the 11 patients showed significant organ motion (>1 cm), with some motion lasting longer that 1 min. Calypso transponders can be used as magnetic intraprostatic fiducials. Clinical evaluation of this novel 4D nonionizing electromagnetic localization system with transponders indicates a comparable localization accuracy to isocenter, (within 2 mm) compared with X-ray localization.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Three-dimensional intrafractional movement of prostate measured during real-time tumor-tracking radiotherapy in supine and prone treatment positions.

              To quantify three-dimensional (3D) movement of the prostate gland with the patient in the supine and prone positions and to analyze the movement frequency for each treatment position. The real-time tumor-tracking radiotherapy (RTRT) system was developed to identify the 3D position of a 2-mm gold marker implanted in the prostate 30 times/s using two sets of fluoroscopic images. The linear accelerator was triggered to irradiate the tumor only when the gold marker was located within the region of the planned coordinates relative to the isocenter. Ten patients with prostate cancer treated with RTRT were the subjects of this study. The coordinates of the gold marker were recorded every 0.033 s during RTRT in the supine treatment position for 2 min. The patient was then moved to the prone position, and the marker was tracked for 2 min to acquire data regarding movement in this position. Measurements were taken 5 times for each patient (once a week); a total of 50 sets for the 10 patients was analyzed. The raw data from the RTRT system were filtered to reduce system noise, and the amplitude of movement was then calculated. The discrete Fourier transform of the unfiltered data was performed for the frequency analysis of prostate movement. No apparent difference in movement was found among individuals. The amplitude of 3D movement was 0.1-2.7 mm in the supine and 0.4-24 mm in the prone positions. The amplitude in the supine position was statistically smaller in all directions than that in the prone position (p < 0.0001). The amplitude in the craniocaudal and AP directions was larger than in the left-right direction in the prone position (p < 0.0001). No characteristic movement frequency was detected in the supine position. The respiratory frequency was detected for all patients regarding movement in the craniocaudal and AP directions in the prone position. The results of the frequency analysis suggest that prostate movement is affected by the respiratory cycle and is influenced by bowel movement in the prone position. The results of this study have confirmed that internal organ motion is less frequent in the supine position than in the prone position in the treatment of prostate cancer. RTRT would be useful in reducing uncertainty due to the effects of the respiratory cycle, especially in the prone position.
                Bookmark

                Author and article information

                Contributors
                wbutler@wheelinghospital.org
                Journal
                J Appl Clin Med Phys
                J Appl Clin Med Phys
                10.1002/(ISSN)1526-9914
                ACM2
                Journal of Applied Clinical Medical Physics
                John Wiley and Sons Inc. (Hoboken )
                1526-9914
                04 March 2013
                March 2013
                : 14
                : 2 ( doiID: 10.1002/acm2.2013.14.issue-2 )
                : 198-208
                Affiliations
                [ 1 ] Schiffler Cancer Center Wheeling WV USA
                Author notes
                [*] [* ]Corresponding author: Wayne M. Butler, Schiffler Cancer Center, Wheeling Hospital, 1 Medical Park, Wheeling, WV 26003, USA; phone: 304‐243‐3983; fax: 304‐243‐5047; email: wbutler@ 123456wheelinghospital.org
                Article
                ACM20198
                10.1120/jacmp.v14i2.4141
                5714377
                23470943
                813dfa17-f91a-4892-94ad-1c2a5e2020dc
                © 2013 The Authors.

                This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.

                History
                : 27 July 2012
                : 15 November 2012
                Page count
                Figures: 6, Tables: 3, References: 17, Pages: 11, Words: 5282
                Categories
                Radiation Oncology Physics
                Radiation Oncology Physics
                Custom metadata
                2.0
                acm20198
                March 2013
                Converter:WILEY_ML3GV2_TO_NLMPMC version:5.2.5 mode:remove_FC converted:16.11.2017

                prostate,external beam radiation,calypso,fiducials,intrafraction motion

                Comments

                Comment on this article