      Reproducibility of the external surface position in left‐breast DIBH radiotherapy with spirometer‐based monitoring

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Deep inspiration breath hold (DIBH) in left‐sided breast cancer radiotherapy treatments allows for a reduction in cardiac and pulmonary doses without compromising target coverage. The selection of the most appropriate technology for DIBH monitoring is a crucial issue. We evaluated the stability and reproducibility of DIBHs controlled by a spirometric device, by assessing the variability of the external surface position within a single DIBH (intra‐DIBH) and between DIBHs performed in the same treatment session (intrafraction) or in different sessions (interfraction). The study included seven left‐breast cancer patients treated with spirometer‐based DIBH radiotherapy. Infrared optical tracking was used to record the 3D coordinates of seven to eleven passive markers placed on the patient's thoraco‐abdominal surface during 29‐43 DIBHs performed in six to eight treatment sessions. The obtained results showed displacements of the external surface between different sessions up to 6.3 mm along a single direction, even at constant inspired volumes. The median value of the interfraction variability in the position of breast passive markers was 2.9 mm (range 1.9‐4.8 mm) in the latero‐lateral direction, 3.6 mm (range 2.2‐4.6 mm) in the antero‐posterior direction, and 4.3 mm (range 2.8‐6.2 mm) in the cranio‐caudal direction. There were no significant dose distribution variations for target and organs at risk with respect to the treatment plan, confirming the adequacy of the applied clinical margins (15 mm) to compensate for the measured setup uncertainties. This study demonstrates that spirometer‐based control does not guarantee a stable and reproducible position of the external surface in left‐breast DIBH radiotherapy, suggesting the need for more robust DIBH monitoring techniques when reduced margins and setup uncertainties are required for improving normal tissue sparing and decreasing cardiac and pulmonary toxicity.

          PACS number: 87.55.Km

          Related collections

          Most cited references24

          • Record: found
          • Abstract: found
          • Article: not found

          Errors and margins in radiotherapy.

          Clinical radiotherapy procedures aim at high accuracy. However, there are many error sources that act during treatment preparation and execution that limit the accuracy. As a consequence, a safety margin is required to ensure that the planned dose is actually delivered to the target for (almost) all patients. Before treatment planning, a planning computed tomography scan is made. In particular, motion of skin with respect to the internal anatomy limits the reproducibility of this step, introducing a systematic setup error. The second important error source is organ motion. The tumor is imaged in an arbitrary position, leading to a systematic organ motion error. The image may also be distorted because of the interference of the scanning process and organ motion. A further systematic error introduced during treatment planning is caused by the delineation process. During treatment, the most important errors are setup error and organ motion leading to day-to-day variations. There are many ways to define the margins required for these errors. In this article, an overview is given of errors in radiotherapy and margin recipes, based on physical and biological considerations. Respiration motion is treated separately.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Deep inspiration breath-hold technique for lung tumors: the potential value of target immobilization and reduced lung density in dose escalation.

            This study evaluates the dosimetric benefits and feasibility of a deep inspiration breath-hold (DIBH) technique in the treatment of lung tumors. The technique has two distinct features--deep inspiration, which reduces lung density, and breath-hold, which immobilizes lung tumors, thereby allowing for reduced margins. Both of these properties can potentially reduce the amount of normal lung tissue in the high-dose region, thus reducing morbidity and improving the possibility of dose escalation. Five patients treated for non-small cell lung carcinoma (Stage IIA-IIIB) received computed tomography (CT) scans under 4 respiration conditions: free-breathing, DIBH, shallow inspiration breath-hold, and shallow expiration breath-hold. The free-breathing and DIBH scans were used to generate 3-dimensional conformal treatment plans for comparison, while the shallow inspiration and expiration scans determined the extent of tumor motion under free-breathing conditions. To acquire the breath-hold scans, the patients are brought to reproducible respiration levels using spirometry, and for DIBH, modified slow vital capacity maneuvers. Planning target volumes (PTVs) for free-breathing plans included a margin for setup error (0.75 cm) plus a margin equal to the extent of tumor motion due to respiration (1-2 cm). Planning target volumes for DIBH plans included the same margin for setup error, with a reduced margin for residual uncertainty in tumor position (0.2-0.5 cm) as determined from repeat fluoroscopic movies. To simulate the effects of respiration-gated treatments and estimate the role of target immobilization alone (i.e., without the benefit of reduced lung density), a third plan is generated from the free-breathing scan using a PTV with the same margins as for DIBH plans. The treatment plan comparison suggests that, on average, the DIBH technique can reduce the volume of lung receiving more than 25 Gy by 30% compared to free-breathing plans, while respiration gating can reduce the volume by 18%. The DIBH maneuver was found to be highly reproducible, with intra breath-hold reproducibility of 1.0 (+/- 0.9) mm and inter breath-hold reproducibility of 2.5 (+/- 1.6) mm, as determined from diaphragm position. Patients were able to perform 10-13 breath-holds in one session, with a comfortable breath-hold duration of 12-16 s. Patients tolerate DIBH maneuvers well and can perform them in a highly reproducible fashion. Compared to conventional free-breathing treatment, the DIBH technique benefits from reduced margins, as a result of the suppressed target motion, as well as a decreased lung density; both contribute to moving normal lung tissue out of the high-dose region. Because less normal lung tissue is irradiated to high dose, the possibility for dose escalation is significantly improved.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Using surface imaging and visual coaching to improve the reproducibility and stability of deep-inspiration breath hold for left-breast-cancer radiotherapy.

              Late cardiac complications may arise after left-breast radiation therapy. Deep-inspiration breath hold (DIBH) allows reduction of the irradiated heart volume at the same time as it reduces tumor bed motion and increases lung sparing. In the present study, we have evaluated the improvement in reproducibility and stability of the DIBH for left-breast-cancer treatment when visual coaching is provided with the aid of 3D video surface imaging and video goggles. Five left-breast-cancer patients and fifteen healthy volunteers were asked to perform a series of DIBHs without and with visual coaching. Reproducibility and stability of DIBH were measured for each individual with and without visual coaching. The average reproducibility and stability changed from 2.1 mm and 1.5 mm, respectively, without visual feedback to 0.5 mm and 0.7 mm with visual feedback, showing a significant statistical difference (p 2 mm) in reproducibility and stability were observed in 35% and 15% of the subjects, respectively. The average chest wall excursion of the DIBH with respect to the free breathing preceding the DIBH was found to be 11.3 mm. The reproducibility and stability of the DIBH improve significantly from the visual coaching provided to the patient, especially in those patients with poor reproducibility and stability.
                Bookmark

                Author and article information

                Contributors
                aurora.fassi@polimi.it
                Journal
                J Appl Clin Med Phys
                J Appl Clin Med Phys
                10.1002/(ISSN)1526-9914
                ACM2
                Journal of Applied Clinical Medical Physics
                John Wiley and Sons Inc. (Hoboken )
                1526-9914
                04 January 2014
                January 2014
                : 15
                : 1 ( doiID: 10.1002/acm2.2014.15.issue-1 )
                : 130-140
                Affiliations
                [ 1 ] Dipartimento di Elettronica Informazione e Bioingegneria, Politecnico di Milano Milano Italy
                [ 2 ] Department of Radiation Oncology Fondazione Salvatore Maugeri Pavia Italy
                [ 3 ] Medical Physics Fondazione Salvatore Maugeri Pavia Italy
                [ 4 ] Bioengineering Unit CNAO Foundation Pavia Italy
                Author notes
                [*] [* ] a Corresponding author: Aurora Fassi, Dipartimento di Elettronica, Informazione e Bioingegneria, Politecnico di Milano, Piazza Leonardo da Vinci 32, 20133 Milano, Italy; phone: (+39) 02 2399 9022; fax: (+39) 02 2399 9000; email: aurora.fassi@ 123456polimi.it

                Article
                ACM20130
                10.1120/jacmp.v15i1.4494
                5711250
                24423845
                82813c70-61d6-4e3e-b346-16e5049d75de
                © 2014 The Authors.

                This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.

                History
                : 13 April 2013
                : 12 August 2013
                Page count
                Figures: 5, Tables: 2, References: 29, Pages: 11, Words: 5495
                Categories
                Radiation Oncology Physics
                Radiation Oncology Physics
                Custom metadata
                2.0
                acm20130
                January 2014
                Converter:WILEY_ML3GV2_TO_NLMPMC version:5.2.5 mode:remove_FC converted:17.11.2017

                left‐breast cancer,dibh radiotherapy,spirometric monitoring,optical tracking,surface reproducibility

                Comments

                Comment on this article