      Repeatability and Reproducibility of Foveal Avascular Zone Area Measurement on Normal Eyes by Different Optical Coherence Tomography Angiography Instruments

      research-article

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Purpose: To compare the foveal avascular zone (FAZ) area measurements produced by different optical coherence tomography angiography (OCTA). Methods: Healthy enrolled volunteers underwent OCTA using 2 different devices: Spectralis HRA+OCTA (Heidelberg Engineering, Heidelberg, Germany) and RS-3000 Advance (Nidek, Gamagori, Japan). Two graders measured FAZ in both superficial (SCP) and deep (DCP) retinal capillary plexuses. The SCP and DCP en face images were visualized automatically segmenting 2 separate slabs defined by the arbitrary segmentation lines created by the software of each OCT device. One grader repeated each measure twice. Results: Fifty-nine eyes were included. The mean FAZ was 0.33 ± 0.09 mm<sup>2</sup> at the SCP and 0.57 ± 0.17 mm<sup>2</sup> at the DCP measured with RS-3000 versus 0.30 ± 0.08 and 0.35 ± 0.08 mm<sup>2</sup>, respectively, measured with Spectralis. The measurements of the 2 devices were significantly different ( p < 0.0001). The intraoperator agreement was excellent at the SCP (intraclass correlation coefficient, ICC: 0.97 with Spectralis and 0.96 with RS-3000). At the DCP, it was good with Spectralis and fair with RS-3000 (ICC: 0.85 and 0.64, respectively). The interoperator agreement was excellent for Spectralis and good for RS-3000 at the SCP (ICC: 0.97 and 0.93, respectively). It was good at the DCP with both devices (ICC: 0.74 with RS-3000 and 0.81 with Spectralis). Conclusions: FAZ measurements obtained with different OCTA devices differ. These findings should be considered in follow-up studies of patients with retinal vascular diseases.

          Related collections

          Author and article information

          Journal
          ORE
          Ophthalmic Res
          10.1159/issn.0030-3747
          Ophthalmic Research
          S. Karger AG
          0030-3747
          1423-0259
          2018
          May 2018
          08 February 2018
          : 59
          : 4
          : 206-211
          Affiliations
          [_a] aDepartment of Ophthalmology, University of Padova, Padova, Italy
          [_b] bG.B. Bietti Foundation, IRCCS, Rome, Italy
          Author notes
          *Edoardo Midena, MD, Department of Ophthalmology, University of Padova, Via Giustiniani 2, IT–35128 Padova (Italy), E-Mail edoardo.midena@unipd.it
          Article
          485463 Ophthalmic Res 2018;59:206–211
          10.1159/000485463
          29421813
          831c4261-1c45-42d8-8503-011390e1d16a
          © 2018 S. Karger AG, Basel

          Copyright: All rights reserved. No part of this publication may be translated into other languages, reproduced or utilized in any form or by any means, electronic or mechanical, including photocopying, recording, microcopying, or by any information storage and retrieval system, without permission in writing from the publisher. Drug Dosage: The authors and the publisher have exerted every effort to ensure that drug selection and dosage set forth in this text are in accord with current recommendations and practice at the time of publication. However, in view of ongoing research, changes in government regulations, and the constant flow of information relating to drug therapy and drug reactions, the reader is urged to check the package insert for each drug for any changes in indications and dosage and for added warnings and precautions. This is particularly important when the recommended agent is a new and/or infrequently employed drug. Disclaimer: The statements, opinions and data contained in this publication are solely those of the individual authors and contributors and not of the publishers and the editor(s). The appearance of advertisements or/and product references in the publication is not a warranty, endorsement, or approval of the products or services advertised or of their effectiveness, quality or safety. The publisher and the editor(s) disclaim responsibility for any injury to persons or property resulting from any ideas, methods, instructions or products referred to in the content or advertisements.

          History
          : 15 August 2017
          : 20 November 2017
          Page count
          Figures: 3, Pages: 6
          Categories
          Original Paper

          Vision sciences,Ophthalmology & Optometry,Pathology
          Optical coherence tomography angiography,Deep retinal capillary plexus,Foveal avascular zone area,Superficial retinal capillary plexus

          Comments

          Comment on this article