      Does objective measurement of tracheal tube cuff pressures minimise adverse effects and maintain accurate cuff pressures? A systematic review and meta-analysis.

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Correct inflation pressures of the tracheal cuff are recommended to ensure adequate ventilation and prevent aspiration and adverse events. However there are conflicting views on which measurement to employ. The aim of this review was to examine whether adjustment of cuff pressure guided by objective measurement, compared with subjective measurement or observation of the pressure value alone, was able to prevent patient-related adverse effects and maintain accurate cuff pressures. A search of PubMed, Web of Science, Embase, CINAHL and ScienceDirect was conducted using keywords 'cuff pressure' and 'measure*' and related synonyms. Included studies were randomised or pseudo-randomised controlled trials investigating mechanically ventilated patients both in the intensive care unit and during surgery. Outcomes included adverse effects and the comparison of pressure measurements. Pooled analyses were performed to calculate risk ratios, effect sizes and 95% confidence intervals. Meta-analysis found preliminary evidence that adjustment of cuff pressure guided by objective measurement as compared with subjective measurement or observation of the pressure value alone, has benefit in preventing adverse effects. These included cough at two hours (odds ratio [OR] 0.42, confidence interval [CI] 0.23 to 0.79, P=0.007), hoarseness at 24 hours (OR 0.49, CI 0.31 to 0.76, P <0.002), sore throat (OR 0.73, CI 0.54 to 0.97, P <0.03), lesions of the trachea and incidences of silent aspiration (P=0.001), as well as maintaining accurate cuff pressures (Hedges' g 1.61, CI 2.69 to 0.53, P=0.003). Subjective measurement to guide adjustment or observation of the pressure value alone may lead to patient-related adverse effects and inaccuracies. It is recommended that an objective form of measurement be used.

          Related collections

          Author and article information

          Journal
          Anaesth Intensive Care
          Anaesthesia and intensive care
          0310-057X
          0310-057X
          September 2016
          : 44
          : 5
          Affiliations
          [1 ] Physiotherapist, Physiotherapy Department, Royal Brisbane and Women's Hospital, Brisbane, Queensland.
          [2 ] Professor of Anaesthesiology, Department of Anaesthesia and Perioperative Medicine, Royal Brisbane and Women's Hospital, Chairman of Anaesthesiology, Senior Staff Specialist School of Medicine, University of Queensland, Brisbane, Queensland.
          [3 ] Principal Research Fellow, Physiotherapy Department, Royal Brisbane and Women's Hospital, Principal Research Fellow, School of Medicine, University of Queensland, Principal Research Fellow, School of Allied Health Sciences, Griffith University, Brisbane, Queensland.
          Article
          20150412
          10.1177/0310057X1604400503
          27608338
          8cb3bb2c-2c75-4544-9f8c-074e85710c64
          History

          airway, intubation, tracheal, tracheostomy, intensive care