      Modifications to mobile chest radiography technique during the COVID-19 pandemic – implications of X-raying through side room windows

      research-article

      Read this article at

      ScienceOpenPublisherPMC
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Introduction

          Modifications to common radiographic techniques have resulted from the challenges presented by the COVID-19 pandemic. Reports exist regarding the potential benefits of undertaking mobile radiography through side room windows. The aim of this study was to evaluate the impact on image quality and exposure factors when undertaking such examinations.

          Methods

          A phantom based study was undertaken using a digital X-ray room. Control acquisitions, using a commercially available image quality test tool, were performed using standard mobile chest radiography acquisition factors. Image quality (physical and visual), incidence surface air kerma (ISAK), Exposure Index (EI) and Deviation Index (DI) were recorded. Image quality and radiation dose were further assessed for two additional (experimental) scenarios, where a side room window was located immediately adjacent to the exit port of the light beam diaphragm. The goal of experimental scenario one was to modify exposure factors to maintain the control ISAK. The goal of experimental scenario two was to modify exposure factors to maintain the control EI and DI. Dose and image quality data were compared between the three scenarios.

          Results

          To maintain the pre-window (control) ISAK (76 μGy), tube output needed a three-fold increase (90 kV/4 mAs versus 90 kV/11.25 mAs). To maintain EI/DI a more modest increase in tube output was required (90 kV/8 mAs/ISAK 54 μGy). Physical and visual assessments of spatial resolution and signal-to-noise ratio were indifferent between the three scenarios. There was a slight statistically significant reduction in contrast-to-noise ratio when imaging through the glass window (2.3 versus 1.4 and 1.2; P=0.005).

          Conclusions

          Undertaking mobile X-ray examinations through side room windows is potentially feasible but does require an increase in tube output and is likely to be limited by minor reductions in image quality.

          Implications for practice

          Mobile examinations performed through side room windows should only be used in limited circumstances and future clinical evaluation of this technique is warranted.

          Related collections

          Author and article information

          Contributors
          Journal
          Radiography (Lond)
          Radiography (Lond)
          Radiography (London, England : 1995)
          Published by Elsevier Ltd on behalf of The College of Radiographers.
          1078-8174
          1532-2831
          3 August 2020
          3 August 2020
          Affiliations
          [a ]School of Allied Health Professions, Keele University, Staffordshire, UK
          [b ]Department of Radiology, Warrington and Halton Teaching Hospitals NHS Foundation Trust, Warrington, UK
          [c ]LPD Lab Services, Blackburn, UK
          [d ]Medical Imaging Department, Teesside University, Middlesbrough, UK
          Author notes
          []Corresponding author. Mackay Building, School of Allied Health Professions, Keele University, Staffordshire, UK a.england@ 123456keele.ac.uk
          Article
          S1078-8174(20)30139-5
          10.1016/j.radi.2020.07.015
          7396953
          32855021
          8d412199-9bb0-4337-b681-e5375ef25605
          © 2020 Published by Elsevier Ltd on behalf of The College of Radiographers.

          Since January 2020 Elsevier has created a COVID-19 resource centre with free information in English and Mandarin on the novel coronavirus COVID-19. The COVID-19 resource centre is hosted on Elsevier Connect, the company's public news and information website. Elsevier hereby grants permission to make all its COVID-19-related research that is available on the COVID-19 resource centre - including this research content - immediately available in PubMed Central and other publicly funded repositories, such as the WHO COVID database with rights for unrestricted research re-use and analyses in any form or by any means with acknowledgement of the original source. These permissions are granted for free by Elsevier for as long as the COVID-19 resource centre remains active.

          History
          : 29 June 2020
          : 21 July 2020
          : 23 July 2020
          Categories
          Article

          covid-19,mobile radiography,modified technique,image quality,radiation dose

          Comments

          Comment on this article