      Accuracy of Wrist-Worn Activity Monitors During Common Daily Physical Activities and Types of Structured Exercise: Evaluation Study

      research-article

      Read this article at

      ScienceOpenPublisherPMC
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          Wrist-worn activity monitors are often used to monitor heart rate (HR) and energy expenditure (EE) in a variety of settings including more recently in medical applications. The use of real-time physiological signals to inform medical systems including drug delivery systems and decision support systems will depend on the accuracy of the signals being measured, including accuracy of HR and EE. Prior studies assessed accuracy of wearables only during steady-state aerobic exercise.

          Objective

          The objective of this study was to validate the accuracy of both HR and EE for 2 common wrist-worn devices during a variety of dynamic activities that represent various physical activities associated with daily living including structured exercise.

          Methods

          We assessed the accuracy of both HR and EE for two common wrist-worn devices (Fitbit Charge 2 and Garmin vívosmart HR+) during dynamic activities. Over a 2-day period, 20 healthy adults (age: mean 27.5 [SD 6.0] years; body mass index: mean 22.5 [SD 2.3] kg/m 2; 11 females) performed a maximal oxygen uptake test, free-weight resistance circuit, interval training session, and activities of daily living. Validity was assessed using an HR chest strap (Polar) and portable indirect calorimetry (Cosmed). Accuracy of the commercial wearables versus research-grade standards was determined using Bland-Altman analysis, correlational analysis, and error bias.

          Results

          Fitbit and Garmin were reasonably accurate at measuring HR but with an overall negative bias. There was more error observed during high-intensity activities when there was a lack of repetitive wrist motion and when the exercise mode indicator was not used. The Garmin estimated HR with a mean relative error (RE, %) of −3.3% (SD 16.7), whereas Fitbit estimated HR with an RE of −4.7% (SD 19.6) across all activities. The highest error was observed during high-intensity intervals on bike (Fitbit: −11.4% [SD 35.7]; Garmin: −14.3% [SD 20.5]) and lowest error during high-intensity intervals on treadmill (Fitbit: −1.7% [SD 11.5]; Garmin: −0.5% [SD 9.4]). Fitbit and Garmin EE estimates differed significantly, with Garmin having less negative bias (Fitbit: −19.3% [SD 28.9], Garmin: −1.6% [SD 30.6], P<.001) across all activities, and with both correlating poorly with indirect calorimetry measures.

          Conclusions

          Two common wrist-worn devices (Fitbit Charge 2 and Garmin vívosmart HR+) show good HR accuracy, with a small negative bias, and reasonable EE estimates during low to moderate-intensity exercise and during a variety of common daily activities and exercise. Accuracy was compromised markedly when the activity indicator was not used on the watch or when activities involving less wrist motion such as cycle ergometry were done.

          Related collections

          Most cited references28

          • Record: found
          • Abstract: found
          • Article: found
          Is Open Access

          Accuracy in Wrist-Worn, Sensor-Based Measurements of Heart Rate and Energy Expenditure in a Diverse Cohort

          The ability to measure physical activity through wrist-worn devices provides an opportunity for cardiovascular medicine. However, the accuracy of commercial devices is largely unknown. The aim of this work is to assess the accuracy of seven commercially available wrist-worn devices in estimating heart rate (HR) and energy expenditure (EE) and to propose a wearable sensor evaluation framework. We evaluated the Apple Watch, Basis Peak, Fitbit Surge, Microsoft Band, Mio Alpha 2, PulseOn, and Samsung Gear S2. Participants wore devices while being simultaneously assessed with continuous telemetry and indirect calorimetry while sitting, walking, running, and cycling. Sixty volunteers (29 male, 31 female, age 38 ± 11 years) of diverse age, height, weight, skin tone, and fitness level were selected. Error in HR and EE was computed for each subject/device/activity combination. Devices reported the lowest error for cycling and the highest for walking. Device error was higher for males, greater body mass index, darker skin tone, and walking. Six of the devices achieved a median error for HR below 5% during cycling. No device achieved an error in EE below 20 percent. The Apple Watch achieved the lowest overall error in both HR and EE, while the Samsung Gear S2 reported the highest. In conclusion, most wrist-worn devices adequately measure HR in laboratory-based activities, but poorly estimate EE, suggesting caution in the use of EE measurements as part of health improvement programs. We propose reference standards for the validation of consumer health devices (http://precision.stanford.edu/).
            Bookmark
            • Record: found
            • Abstract: found
            • Article: found
            Is Open Access

            Using Fitness Trackers and Smartwatches to Measure Physical Activity in Research: Analysis of Consumer Wrist-Worn Wearables

            Background New fitness trackers and smartwatches are released to the consumer market every year. These devices are equipped with different sensors, algorithms, and accompanying mobile apps. With recent advances in mobile sensor technology, privately collected physical activity data can be used as an addition to existing methods for health data collection in research. Furthermore, data collected from these devices have possible applications in patient diagnostics and treatment. With an increasing number of diverse brands, there is a need for an overview of device sensor support, as well as device applicability in research projects. Objective The objective of this study was to examine the availability of wrist-worn fitness wearables and analyze availability of relevant fitness sensors from 2011 to 2017. Furthermore, the study was designed to assess brand usage in research projects, compare common brands in terms of developer access to collected health data, and features to consider when deciding which brand to use in future research. Methods We searched for devices and brand names in six wearable device databases. For each brand, we identified additional devices on official brand websites. The search was limited to wrist-worn fitness wearables with accelerometers, for which we mapped brand, release year, and supported sensors relevant for fitness tracking. In addition, we conducted a Medical Literature Analysis and Retrieval System Online (MEDLINE) and ClinicalTrials search to determine brand usage in research projects. Finally, we investigated developer accessibility to the health data collected by identified brands. Results We identified 423 unique devices from 132 different brands. Forty-seven percent of brands released only one device. Introduction of new brands peaked in 2014, and the highest number of new devices was introduced in 2015. Sensor support increased every year, and in addition to the accelerometer, a photoplethysmograph, for estimating heart rate, was the most common sensor. Out of the brands currently available, the five most often used in research projects are Fitbit, Garmin, Misfit, Apple, and Polar. Fitbit is used in twice as many validation studies as any other brands and is registered in ClinicalTrials studies 10 times as often as other brands. Conclusions The wearable landscape is in constant change. New devices and brands are released every year, promising improved measurements and user experience. At the same time, other brands disappear from the consumer market for various reasons. Advances in device quality offer new opportunities for research. However, only a few well-established brands are frequently used in research projects, and even less are thoroughly validated.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Validity of consumer-based physical activity monitors.

              Many consumer-based monitors are marketed to provide personal information on the levels of physical activity and daily energy expenditure (EE), but little or no information is available to substantiate their validity. This study aimed to examine the validity of EE estimates from a variety of consumer-based, physical activity monitors under free-living conditions. Sixty (26.4 ± 5.7 yr) healthy males (n = 30) and females (n = 30) wore eight different types of activity monitors simultaneously while completing a 69-min protocol. The monitors included the BodyMedia FIT armband worn on the left arm, the DirectLife monitor around the neck, the Fitbit One, the Fitbit Zip, and the ActiGraph worn on the belt, as well as the Jawbone Up and Basis B1 Band monitor on the wrist. The validity of the EE estimates from each monitor was evaluated relative to criterion values concurrently obtained from a portable metabolic system (i.e., Oxycon Mobile). Differences from criterion measures were expressed as a mean absolute percent error and were evaluated using 95% equivalence testing. For overall group comparisons, the mean absolute percent error values (computed as the average absolute value of the group-level errors) were 9.3%, 10.1%, 10.4%, 12.2%, 12.6%, 12.8%, 13.0%, and 23.5% for the BodyMedia FIT, Fitbit Zip, Fitbit One, Jawbone Up, ActiGraph, DirectLife, NikeFuel Band, and Basis B1 Band, respectively. The results from the equivalence testing showed that the estimates from the BodyMedia FIT, Fitbit Zip, and NikeFuel Band (90% confidence interval = 341.1-359.4) were each within the 10% equivalence zone around the indirect calorimetry estimate. The indicators of the agreement clearly favored the BodyMedia FIT armband, but promising preliminary findings were also observed with the Fitbit Zip.
                Bookmark

                Author and article information

                Contributors
                Journal
                JMIR Mhealth Uhealth
                JMIR Mhealth Uhealth
                JMU
                JMIR mHealth and uHealth
                JMIR Publications (Toronto, Canada )
                2291-5222
                December 2018
                10 December 2018
                : 6
                : 12
                : e10338
                Affiliations
                [1 ] Department of Biomedical Engineering Oregon Health & Science University Portland, OR United States
                [2 ] School of Kinesiology and Health Science York University Toronto, ON Canada
                [3 ] Harold Schnitzer Diabetes Health Center Oregon Health & Science University Portland, OR United States
                [4 ] Harvard John A Paulson School of Engineering and Applied Sciences Harvard University Cambridge, MA United States
                [5 ] Children's Mercy Kansas City Kansas City, MO United States
                [6 ] Institute for Diabetes, Obesity & Metabolism University of Pennsylvania Perelman School of Medicine Philadelphia, PA United States
                [7 ] Department of Pediatrics University of Kansas Medical Center Kansas City, KS United States
                Author notes
                Corresponding Author: Peter G Jacobs jacobsp@ 123456ohsu.edu
                Author information
                http://orcid.org/0000-0001-9612-1314
                http://orcid.org/0000-0002-8514-1405
                http://orcid.org/0000-0002-9374-8469
                http://orcid.org/0000-0002-0537-8243
                http://orcid.org/0000-0003-1618-8290
                http://orcid.org/0000-0001-5333-6892
                http://orcid.org/0000-0002-3293-9114
                http://orcid.org/0000-0002-2368-0331
                http://orcid.org/0000-0002-9253-838X
                http://orcid.org/0000-0002-8902-6965
                http://orcid.org/0000-0003-1179-5374
                http://orcid.org/0000-0001-6556-7559
                http://orcid.org/0000-0001-9897-4783
                Article
                v6i12e10338
                10.2196/10338
                6305876
                30530451
                8dac90bf-7aa1-4474-85ef-26c599a6bc20
                ©Ravi Kondama Reddy, Rubin Pooni, Dessi P Zaharieva, Brian Senf, Joseph El Youssef, Eyal Dassau, Francis J Doyle III, Mark A Clements, Michael R Rickels, Susana R Patton, Jessica R Castle, Michael C Riddell, Peter G Jacobs. Originally published in JMIR Mhealth and Uhealth (http://mhealth.jmir.org), 10.12.2018.

                This is an open-access article distributed under the terms of the Creative Commons Attribution License ( https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR mhealth and uhealth, is properly cited. The complete bibliographic information, a link to the original publication on http://mhealth.jmir.org/.as well as this copyright and license information must be included.

                History
                : 19 March 2018
                : 23 April 2018
                : 17 June 2018
                : 5 September 2018
                Categories
                Original Paper
                Original Paper

                heart rate,energy metabolism,fitness trackers,high-intensity interval training,artificial pancreas