17
views
0
recommends
+1 Recommend
0 collections
    0
    shares
      • Record: found
      • Abstract: found
      • Article: not found

      Characterisation of wear particles produced by metal on metal and ceramic on metal hip prostheses under standard and microseparation simulation.

      Journal of Materials Science. Materials in Medicine
      Ceramics, Coated Materials, Biocompatible, Hip Prosthesis, adverse effects, Humans, In Vitro Techniques, Materials Testing, Metals, Microscopy, Electron, Scanning, Models, Biological, Particle Size, Prosthesis Failure

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          The failure of metal on polyethylene total hip replacements due to wear particle induced osteolysis and late aseptic loosening has focused interest upon alternative bearings, such as metal on metal implants. A recent advance in this field has been the development of a novel ceramic on metal implant. The characteristics of the wear particles generated in this low-wearing bearing have not been previously determined. The aims of this study were to characterise metal wear particles from metal on metal and ceramic on metal hips under standard and adverse (microseparation) wear conditions. Accurate characterisation of cobalt-chrome wear particles is difficult since the reactive nature of the particles prevents them from being isolated using acids and bases. A method was developed to isolate the metal wear particles using enzymes to digest serum containing lubricants from metal on metal and ceramic on metal hip simulations. High resolution scanning electron microscopy was then used to characterise the wear particles generated by both metal on metal and ceramic on metal implants under standard and microseparation wear conditions. The wear particles isolated from all simulations had a mean size of less than 50 nm with a rounded and irregular morphology. No significant difference was found between the size of wear particles generated under any conditions.

          Related collections

          Author and article information

          Comments

          Comment on this article