      validated microbiological and HPLC methods for the determination of moxifloxacin in pharmaceutical preparations and human plasma

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          The article presents a comparison between microbiological and high performance liquid chromatographic (HPLC) assays for quantification of moxifloxacin in tablets, ophthalmic solutions and human plasma. The microbiological method employed a cylinder-plate agar diffusion assay using a strain of Esherichia coli ATCC 25922 as the test organism and phosphate buffer (pH8) as the diluent. The calibration curves were linear (R 2 > 0.98) over a concentration range of 0.125 to 16 µgml -1. The within day and between days precisions were ≤ 4.47% and ≤ 6.39% respectively. Recovery values were between 89.4 and 110.2%. The HPLC assay used Hypersil ® BDS C 18 reversed phase column (250 ×4.6 mm, 5µm) with a mobile phase comprising 20 mM ammonium dihydrogen orthophosphate (pH3) and acetonitrile (75:25) and flowing at 1.5 ml/min. The detection was at 295nm. The calibration curves were linear (R 2 > 0.999) over the range of 0.125 to 16 µg ml -1. The within day and between days precisions were ≤ 4.07% and ≤ 5.09% respectively. Recovery values were between 97.7 and 107.6%. Similar potencies were obtained after the analysis of moxifloxacin tablets and ophthalmic solutions by both methods. Also pharmacokinetic parameters were calculated after the analysis of plasma samples of six male healthhy volunteers by both validated methods.

          Related collections

          Most cited references53

          • Record: found
          • Abstract: found
          • Article: not found

          Pharmacokinetics, safety, and tolerability of ascending single doses of moxifloxacin, a new 8-methoxy quinolone, administered to healthy subjects.

          The pharmacokinetics of moxifloxacin were investigated in six studies after oral administration of 50, 100, 200, 400, 600, and 800 mg. Eight healthy male volunteers were included in each study. With doses of up to 200 mg the study was performed as a double-blind, randomized group comparison (n = 6 verum and n = 2 matched placebo); with the higher doses the study was conducted with a double-blind, randomized, crossover design. Safety and tolerability were assessed by evaluation of vital signs, electrocardiograms, electroencephalograms, clinical chemistry parameters, results of urinalysis, and adverse events. The drug was well tolerated. The concentrations of moxifloxacin in plasma, urine, and saliva were determined by a validated high-pressure liquid chromatography assay with fluorescence detection. In addition, plasma and urine samples were analyzed by a bioassay. A good correlation between both methods was seen, indicating an absence of major active metabolites. The mean maximum concentrations of moxifloxacin in plasma (Cmax) ranged from 0.29 mg/liter (50-mg dose) to 4.73 mg/liter (800-mg dose) and were reached 0.5 to 4 h following drug administration. After reaching the Cmax, plasma moxifloxacin concentrations declined in a biphasic manner. Within 4 to 5 h they fell to about 30 to 55% of the Cmax, and thereafter a terminal half-life of 11 to 14 h accounted for the major part of the area under the concentration-time curve (AUC). During the absorption phase concentrations in saliva were even higher than those in plasma, whereas in the terminal phase a constant ratio of the concentration in saliva/concentration in plasma of between 0.5 and 1 was observed, indicating a correlation between unbound concentrations in plasma and levels in saliva (protein binding level, approximately 48%). AUC and Cmax increased proportionally to the dose over the whole range of doses investigated. Urinary excretion amounted to approximately 20% of the dose. Data on renal clearance (40 to 51 ml/min/1.73 m2) indicated partial tubular reabsorption of the drug. The pharmacokinetic parameters derived from compartmental and noncompartmental analyses were in good agreement. The kinetics could be described best by fitting the data to a two-compartment body model.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Validated spectrophotometric methods for the estimation of moxifloxacin in bulk and pharmaceutical formulations.

            New, simple, cost effective, accurate and reproducible UV-spectrophotometric methods were developed and validated for the estimation of moxifloxacin in bulk and pharmaceutical formulations. Moxifloxacin was estimated at 296 nm in 0.1N hydrochloric acid (pH 1.2) and at 289 nm in phosphate buffer (pH 7.4). Beer's law was obeyed in the concentration range of 1-12 microg ml(-1) (r2=0.9999) in hydrochloric acid and 1-14 microg ml(-1) (r2=0.9998) in the phosphate buffer medium. The apparent molar absorptivity and Sandell's sensitivity coefficient were found to be 4.63 x 10(4) l mol(-1) cm(-1) and 9.5 ng cm(-2)/0.001 A in hydrochloric acid; and 4.08 x 10(4) l mol(-1) cm(-1) and 10.8 ng cm(-2)/0.001 A in phosphate buffer media, respectively indicating the high sensitivity of the proposed methods. These methods were tested and validated for various parameters according to ICH guidelines. The detection and quantitation limits were found to be 0.0402, 0.1217 microg ml(-1) in hydrochloric acid and 0.0384, 0.1163 microg ml(-1) in phosphate buffer medium, respectively. The proposed methods were successfully applied for the determination of moxifloxacin in pharmaceutical formulations (tablets, i.v. infusions, eye drops and polymeric nanoparticles). The results demonstrated that the procedure is accurate, precise and reproducible (relative standard deviation <2%), while being simple, cheap and less time consuming and hence can be suitably applied for the estimation of moxifloxacin in different dosage forms and dissolution studies.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              An HPLC assay and a microbiological assay to determine levofloxacin in soft tissue, bone, bile and serum.

              A simple, specific and sensitive HPLC assay for levofloxacin in serum, bile, soft tissue and bone was evaluated and validated. The samples were prepared by protein precipitation with acids and methanol, which yielded high recoveries (for serum and bile>98% and for bone and soft tissue>90%). The compounds were separated on a reversed phase column with an acidic mobile phase containing triethylamine. The eluate was monitored by fluorescence detection. The HPLC assay is linear over the usable concentration range (0.1-40 microg/ml) and it provides good validation data for accuracy and precision. Although comparison of HPLC results to the results of a microbiological assay showed congruent results (regression coefficients>0.967). HPLC should be the method of choice for determination of levofloxacin in biological matrices.
                Bookmark

                Author and article information

                Journal
                Braz J Microbiol
                Braz. J. Microbiol
                bjm
                bjm
                Brazilian Journal of Microbiology
                Sociedade Brasileira de Microbiologia
                1517-8382
                1678-4405
                Oct-Dec 2012
                1 June 2012
                : 43
                : 4
                : 1291-1301
                Affiliations
                [1 ]Department of Pharmaceutical Microbiology, Faculty of Pharmacy , University of Tanta, Tanta, Egypt
                Author notes
                * Corresponding Author. Mailing address: Department of Pharmaceutical Microbiology, Faculty of Pharmacy, University of Tanta, Tanta, Egypt.; Tel.: +2 0106090750 +2 0106090810 Fax: +2 040 3335466.; E-mail: n.gamaleldin@ 123456gmail.com
                Article
                S1517-83822012000400008
                10.1590/S1517-83822012000400008
                3769051
                24031955
                98472318-b51a-4e3c-9098-4b39f9448738
                © Sociedade Brasileira de Microbiologia

                All the content of the journal, except where otherwise noted, is licensed under a Creative Commons License

                History
                : 04 April 2011
                : 17 November 2011
                : 07 June 2012
                Categories
                Medical Microbiology
                Research Paper

                moxifloxacin,microbiological assay,hplc assay,pharmaceutical analysis,pharmacokinetics

                Comments

                Comment on this article