      Simethicone use during gastrointestinal endoscopy: Position statement of the Gastroenterological Society of Australia.

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Concern has been raised regarding the use of simethicone, a de-foaming agent, during endoscopic procedures. Following reports of simethicone residue in endoscope channels despite high level disinfection, an endoscope manufacturer recommended that it not be used due to concerns of biofilm formation and a possible increased risk of microorganism transmission. However, a detailed mucosal assessment is essential in performing high-standard endoscopic procedures. This is impaired by bubbles within the gastrointestinal lumen. The Gastroenterological Society of Australia's Infection Control in Endoscopy Guidelines (ICEG) Committee conducted a literature search utilizing the MEDLINE database. Further references were sourced from published paper bibliographies. Following a review of the available evidence, and drawing on extensive clinical experience, the multidisciplinary ICEG committee considered the risks and benefits of simethicone use in formulating four recommendations. Published reports have documented residual liquid or crystalline simethicone in endoscope channels after high level disinfection. There are no data confirming that simethicone can be cleared from channels by brushing. Multiple series report benefits of simethicone use during gastroscopy and colonoscopy in improving mucosal assessment, adenoma detection rate, and reducing procedure time. There are no published reports of adverse events related specifically to the use of simethicone, delivered either orally or via any endoscope channel. An assessment of the risks and benefits supports the continued use of simethicone during endoscopic procedures. Strict adherence to instrument reprocessing protocols is essential.

          Related collections

          Author and article information

          Journal
          J Gastroenterol Hepatol
          Journal of gastroenterology and hepatology
          Wiley
          1440-1746
          0815-9319
          Dec 2019
          : 34
          : 12
          Affiliations
          [1 ] Medical School, University of Queensland, Brisbane, Queensland, Australia.
          [2 ] Gastroenterological Society of Australia, Melbourne, Victoria, Australia.
          [3 ] Barwon Health, Deakin University, Geelong, Victoria, Australia.
          [4 ] Australasian Society of Infectious Diseases, Hobart, Tasmania, Australia.
          [5 ] Gastroenterological Nurses College of Australia, Beaumaris, Victoria, Australia.
          [6 ] Australasian College for Infection Prevention and Control, Hobart, Tasmania, Australia.
          [7 ] Australian School of Advanced Medicine, Macquarie University, Sydney, New South Wales, Australia.
          Article
          10.1111/jgh.14757
          31242327
          99093f2b-5878-4e54-9a85-b5d10930b790
          History

          gastrointestinal,gastroscopy,simethicone,endoscopy,colonoscopy,endoscope reprocessing

          Comments

          Comment on this article