      Adaptação do Bird Mark 7 para oferta de pressão positiva contínua nas vias aéreas em ventilação não-invasiva: estudo em modelo mecânico Translated title: Adapting the Bird Mark 7 to deliver noninvasive continuous positive airway pressure: a bench study

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          OBJETIVO: Testar a eficiência da adaptação do ventilador Bird Mark 7 para oferecer pressão positiva contínua nas vias aéreas, conhecida como continuous positive airway pressure (CPAP) em inglês, em ventilação não-invasiva. MÉTODOS: Estudo experimental utilizando um modelo mecânico do sistema respiratório. O Bird Mark 7 foi alimentado com 400 e 500 kPa e foi testado em CPAP de 5, 10 e 15 cmH2O. Para avaliar a eficiência da adaptação foram analisados os seguintes variáveis: diferença entre a CPAP pré-determinada e a CPAP realmente atingida (CPAPreal); área da pressão da via aérea sob o nível de CPAP ajustado (ÁREA CPAP); e volume corrente gerado. RESULTADOS: A adaptação do Bird Mark 7 para oferecer CPAP em ventilação não-invasiva conseguiu atingir o volume corrente esperado em todas as situações de esforço inspiratório (normal ou elevado), pressão de alimentação (400 ou 500 kPa) e valor de CPAP (5, 10 ou 15 cmH2O). Para os CPAPs de 5 e 10 cmH2O, o CPAPreal foi muito próximo do pré-determinado, e a ÁREA CPAP teve valor próximo de zero. Para o CPAP de 15 cmH2O, o CPAPreal ficou abaixo do pré-determinado, e a ÁREA CPAP teve valor elevado. CONCLUSÃO: A eficiência da adaptação do Bird Mark 7 para oferecer CPAP em ventilação não-invasiva foi boa para os valores de CPAP de 5 e 10 cmH2O e insuficiente para CPAP de 15 cmH2O. Se adaptado como em nosso estudo, o Bird Mark 7 pode ser uma opção para oferta de CPAP até 10 cmH2O em locais onde equipamentos de ventilação não-invasiva são escassos ou inexistentes.

          Translated abstract

          OBJECTIVE: To test the efficiency of the Bird Mark 7 ventilator adapted to deliver continuous positive airway pressure (CPAP) in noninvasive positive pressure ventilation. METHODS: This was an experimental study using a mechanical model of the respiratory system. A Bird Mark 7 ventilator was supplied with 400 and 500 kPa and tested at CPAP of 5, 10 and 15 cmH2O. The following variables were analyzed: difference between the preset CPAP and the CPAP actually attained CPAP (trueCPAP); area of airway pressure at the CPAP level employed (AREA CPAP); and tidal volume generated. RESULTS: Adapting the Bird Mark 7 to offer CPAP achieved the expected tidal volume in all situations of inspiratory effort (normal or high), ventilator pressure supply (400 or 500 kPa) and CPAP value (5, 10 or 15 cmH2O). At a CPAP of 5 or 10 cmH2O, the trueCPAP was near the preset level, and the AREA CPAP was near zero. However, at a CPAP of 15 cmH2O, the value remained below the preset, and the AREA CPAP was high. CONCLUSION: The efficiency of Bird Mark 7 adaptation in offering CPAP was satisfactory at 5 and 10 cmH2O but insufficient at 15 cmH2O. If adapted as described in our study, the Bird Mark 7 might be an option for offering CPAP up to 10 cmH2O in areas where little or no equipment is available.

          Related collections

          Most cited references12

          • Record: found
          • Abstract: not found
          • Article: not found

          International Consensus Conferences in Intensive Care Medicine: non-invasive positive pressure ventilation in acute respiratory failure. Organised jointly by the American Thoracic Society, the European Respiratory Society, the European Society of Intensive Care Medicine, and the Société de Réanimation de Langue Française, and approved by the ATS Board of Directors, December 2000.

          T. Evans (2000)
            Bookmark
            • Record: found
            • Abstract: not found
            • Article: not found

            A eficácia da aplicação de pressão positiva continua nas vias aéreas (CPAP), com utilização do Bird Mark 7, em pacientes em pós-operatório de cirurgia

              Bookmark
              • Record: found
              • Abstract: not found
              • Article: not found

              Effects of different continuous positive airway pressure devices and periodic hyperinflations on respiratory function

                Bookmark

                Author and article information

                Contributors
                Role: ND
                Role: ND
                Role: ND
                Role: ND
                Role: ND
                Journal
                jbpneu
                Jornal Brasileiro de Pneumologia
                J. bras. pneumol.
                Sociedade Brasileira de Pneumologia e Tisiologia (São Paulo )
                1806-3756
                March 2008
                : 34
                : 3
                : 167-172
                Affiliations
                [1 ] Universidade de São Paulo Brazil
                [2 ] Universidade de São Paulo Brazil
                Article
                S1806-37132008000300007
                10.1590/S1806-37132008000300007
                997b0a53-c766-4e32-a641-2e9ec3be45b0

                http://creativecommons.org/licenses/by/4.0/

                History
                Product

                SciELO Brazil

                Self URI (journal page): http://www.scielo.br/scielo.php?script=sci_serial&pid=1806-3713&lng=en
                Categories
                RESPIRATORY SYSTEM

                Respiratory medicine
                Ventilators, mechanical,Positive-pressure respiration,Continuous positive airway pressure,Respiradores mecânicos,Respiração com pressão positiva,Pressão positiva contínua nas vias aéreas

                Comments

                Comment on this article