Introduction
Leadless pacing is an alternative to transvenous single-chamber ventricular pacing
(VVI/R) with a favorable safety and efficacy profile in selected patients.
1
Until recently, the Micra (Model MC1VR01; Medtronic Inc, Minneapolis, MN) was the
only leadless pacing system approved for clinical use in the United States. The leadless
pacemaker is an attractive option for patients with permanent atrial fibrillation
who require infrequent ventricular back-up pacing and in those in whom atrioventricular
(AV) synchrony is not pursued, especially when vascular access is limited (ie, hemodialysis
patients) or the risk of intravascular infection is deemed high. The recently approved
Micra AV (Model MC1AVR1; Medtronic Inc) represents a major development in leadless
pacing technology. Micra AV uses its accelerometer to sense the atrial mechanical
events and deliver leadless AV-synchronous ventricular pacing (VDD).
2
The Micra AV therefore has the potential to provide VDD pacing with a lower rate of
pocket/lead-related complications vs transvenous systems. Pacing-facilitated ventricular
arrhythmias have previously been reported after Micra implantation.3, 4, 5 Here, we
report the first case of bradycardia-dependent polymorphic ventricular tachycardia
(PMVT) caused by atrial undersensing in a patient implanted with the Micra AV leadless
pacemaker.
Case report
A 92-year-old woman with history of severe aortic stenosis, hypothyroidism, type 2
diabetes mellitus, and recurrent urinary tract infections presented to the hospital
with progressive fatigue and dyspnea over a 2-week period and found to be in sinus
rhythm with complete heart block and stable junctional escape rhythm (Supplemental
Figure 1a). Blood biochemistry was within normal limits, as were high-sensitivity
cardiac biomarkers and thyroid-stimulating hormone levels. Transthoracic echocardiogram
showed mild left ventricular hypertrophy with preserved left ventricular systolic
function. Considering her infectious risk and comorbidities but intact sinus node
function, she was referred for Micra AV insertion.
The patient was brought to our electrophysiology laboratory in the fasting state.
Under conscious sedation, ultrasound-guided right femoral venous access was obtained
and upsized to accept the 27 French delivery system. The Micra AV was successfully
deployed on the right ventricular mid-septum (Figure 1) using the standard implant
technique. After sheath removal, hemostasis was secured with a hemostatic silk suture
and manual compression; there were no acute complications. The pacemaker was programmed
to VDDR 60–105 beats per minute (bpm) and found to have satisfactory acute implant
parameters (impedance 710 W; R wave 8.3 mV; ventricular capture threshold 0.38 V at
0.24 ms). Furthermore, the leadless pacemaker appropriately sensed the atrial mechanical
events and delivered AV synchronous ventricular pacing. She was returned to her inpatient
bed in stable condition.
Figure 1
Thoracic fluoroscopy view depicting the Micra delivery system after deployment of
the Micra device (Medtronic, Minneapolis, MN). View is confirming proper placement
at the septal aspect of the right ventricle on a: standard left anterior oblique view
and b: right anterior oblique view.
Overnight, the patient suffered a syncopal event while supine with seizure-like movements
for about 12 seconds correlating with an episode of PMVT (Figure 2); the tachycardia
terminated spontaneously. Investigations including blood biochemistry, cardiac enzymes,
a transthoracic echocardiogram, and noninvasive ischemia assessment were unremarkable.
Figure 2
Telemetry strip showing initiation of polymorphic ventricular tachycardia following
a paced beat (blue arrows) after premature ventricular contraction (orange arrows).
Dissociated atrial activity is marked with green arrows. Accelerometer sensor failed
to detect the mechanical atrial activity and led to premature pacing at the lower
programmed rate. Figure shows the short-long-short setup for polymorphic arrhythmia
in this patient.
We felt that undersensing of the A wave led to paced beat at the lower VVI mode rate
followed by premature ventricular contractions (PVCs), which triggered a short-long-short
(S-L-S) status leading to the polymorphic arrhythmia. Pacing mode was kept as VDD
but the A4 (atrium active contraction) phase sensing property was adjusted, and we
added beta blockers to her medications. Intervention resulted in a lower sinus node
rate, no further PVCs, and better atrial activity sensing (Figure 3).
Figure 3
A: The correlation between recorded surface electrocardiogram and accelerometer sensing
of ventricular mechanical changes during different phases: A3: passive ventricular
filling, A4: active atrium contraction. Shown is a successful electromechanical correlation
after adjusting A4 sensing and treatment with beta blockers. PVAB = postventricular
atrial blanking period. B: Changes in accelerometer sensing threshold over time in
this patient. Improper sensing after implantation is noted (threshold > 1.6 m/s2).
It also shows improvement in the A4 acceleration sensing after adjustments made to
sensing parameters and titrating up beta blocker dose over time.
Repeat ECG showed well-synchronized paced ventricular rhythm (Supplemental Figure 1b)
and no further arrhythmia was noted on telemetry monitoring. She recovered and was
discharged home to the care of her family.
Discussion
Novel technologies offer new opportunities and challenges. The recently approved Micra
AV pacemaker represents a major development in leadless pacing, having the potential
of delivering leadless AV-synchronous ventricular pacing (VDD). Here, we report the
first case of bradycardia-dependent PMVT caused by atrial undersensing in a patient
implanted with the Micra AV leadless pacemaker.
Single-chamber VVI leadless pacing is commonly used in clinical practice and associated
with lower pocket- and lead-related complications when compared to transvenous pacing
systems.
1
,
6
,
7
Until very recently, leadless pacing with the Micra device (Model MC1VR01) was limited
by single-chamber ventricular sensing, making it unsuitable for patients in which
AV synchrony is desired. The Micra AV (Model MC1AVR1) uses its accelerometer to sense
the atrial mechanical events in sinus rhythm to deliver atrial-triggered ventricular
pacing (VVD), thereby preserving AV synchrony. The implant technique and delivery
system for Micra AV are the same as for Micra such that Micra implanters can seamlessly
opt for the Micra AV system on the basis of the clinical pacing indication. The Micra
Atrial tracking using a Ventricular accelerometer 2 (MARVEL 2) study compared leadless
VDD vs VVI pacing with the Micra AV in 75 patients with sinus rhythm and complete
heart block. Atrioventricular synchrony increased from 26.8% to 89.2% with VVI vs
VDD pacing, respectively, and no serious complications were reported.
2
The benefits of atrial-based vs ventricular-based pacing beyond a mild reduction in
atrial fibrillation have not been consistently demonstrated, whereas the acute and
chronic complication rates are definitely higher with dual-chamber transvenous systems.
8
Nevertheless, the prospect of reliable leadless VDD pacing capabilities will likely
significantly expand the indications for leadless pacemakers.
The potential proarrhythmic effect of cardiac pacing is a rare but well-described
phenomenon with transvenous pacemakers.
9
Conventional pacing modes such as VVI/R and DDD/R have been observed to facilitate
the induction of ventricular tachycardia (VT) and fibrillation (VF). In a combined
post hoc analysis of the PainFree Rx II and EnTrust trials, Sweeney and colleagues
10
found that pacing-associated S-L-S sequences were observed in 29.8% of VT/VF events
and pacing was adjudicated to be causal in 2.6% to 5.2% of VT/VF episodes. Algorithms
to minimize ventricular pacing, such as managed ventricular pacing (Medtronic, Inc,
Minneapolis, MN), have also been observed to facilitate and even trigger VT/VF, albeit
in an exceedingly small minority of patients.
11
,
12
Several hypotheses have been proposed to account for “R-on-T” or S-L-S-mediated PMVT.
At a tissue level, alternations in activation cycle length between paced complexes
and PVCs increase the heterogeneity of repolarization, as repolarization is a rate-dependent
process, setting the stage for ventricular arrhythmias.
13
An appropriately timed PVC in the vulnerable period of ventricular repolarization
can find parts of the myocardium available for conduction while others may be refractory,
leading to functional reentry and PMVT/VF. Fortunately, most patients do not experience
such events, which raises the possibility of subclinical defects in repolarization
being unmasked by additional stressors (eg, acute bradycardia, electrolyte abnormalities,
ischemia) in those patients that do suffer from S-L-S-mediated PMVT.
Leadless pacemaker–facilitated ventricular arrhythmias have previously been reported
with the Micra pacing system.3, 4, 5 Da Costa and colleagues
3
reported a case of VF temporally associated to Micra insertion but without documenting
the mechanism of initiation of tachycardia such that a causal relationship cannot
be ascertained. More recently, Amin and colleagues
5
and Olsen and colleagues
4
each reported a case of Micra-facilitated sustained ventricular arrhythmias, the former
managed with repositioning the Micra and the latter requiring deactivation of the
Micra and implantation of a transvenous pacemaker. The number of cases of leadless
pacemaker–facilitated or leadless pacemaker–triggered ventricular arrhythmias remains
small. However, this may underestimate the true incidence of such events, as the Micra
devices are not designed to store arrhythmia events; hence the events could therefore
go undetected.
Our case is different and highlights a complication associated with leadless VDD pacing,
a novel technology. In our patient, atrial undersensing during VDD pacing led to S-L-S
sequences or “R-on-T,” ultimately triggering PMVT. The phenomenon of atrial undersensing
leading to PMVT with VDD pacing has previously been described with transvenous pacing
systems.
14
The induction of PMVT from an S-L-S sequence is likely a low-probability stochastic
event. However, the rate of inappropriate atrial sensing is significantly higher with
the Micra AV than with transvenous VDD systems, potentially exposing patients to a
higher burden of S-L-S sequences. In fact, in the MARVEL 2 study, AV synchrony was
present in 89.2% of patients after a 20-minute rest period but was as low as 69.8%
while standing.
2
Relying on the mechanical forces of the right atrium (RA) leaves this technology vulnerable
to RA hemodynamic changes. Enlarged RA, hemodynamic acute shifts, and atrial tachycardia/arrhythmia
may compromise the A4 wave sensed by the accelerometer and the device will switch
to VVI mode. Therefore, patient selection should consider all the contributing factors.
Further work is needed to ascertain the burden of S-L-S and risk of VT/VF caused by
atrial undersensing with the Micra AV pacemaker. Arrhythmia electrogram storage should
be within the scope of the available technology and would be very useful in better
defining the incidence of atrial undersensing–induced ventricular arrhythmias with
the Micra AV.
Elsokkari and colleagues
15
reported a case of PMVT in a patient with complete heart block switched from VDD 50-120
bpm to VVI 50 bpm transvenous pacing. They noted that this change in pacing mode was
the functional equivalent of the patient having had de novo AV junction ablation.
This empiric observation forms the basis for a higher backup pacing rate after AV
junction ablation. A similar sequence of events was observed in our patient in that
VDD pacing at the sinus rate was replaced by functional VVI pacing at the lower rate
interval because of atrial undersensing. This abrupt change in rate may have acted
as an additional stressor to the patient’s ventricular repolarization homeostatic
mechanisms. This also highlights the limitations of the accelerometer-based sensing
of atrial mechanical events. It is highly likely that this technology will be further
improved with subsequent iterations of the device.
Conclusion
Novel technologies offer new opportunities and challenges. Here, we report the first
case of bradycardia-dependent PMVT caused by atrial undersensing in a patient implanted
with the Micra AV leadless pacemaker.