      Syncope after successful implantation of atrioventricular synchronous leadless pacemaker caused by polymorphic ventricular tachycardia

      case-report

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Introduction Leadless pacing is an alternative to transvenous single-chamber ventricular pacing (VVI/R) with a favorable safety and efficacy profile in selected patients. 1 Until recently, the Micra (Model MC1VR01; Medtronic Inc, Minneapolis, MN) was the only leadless pacing system approved for clinical use in the United States. The leadless pacemaker is an attractive option for patients with permanent atrial fibrillation who require infrequent ventricular back-up pacing and in those in whom atrioventricular (AV) synchrony is not pursued, especially when vascular access is limited (ie, hemodialysis patients) or the risk of intravascular infection is deemed high. The recently approved Micra AV (Model MC1AVR1; Medtronic Inc) represents a major development in leadless pacing technology. Micra AV uses its accelerometer to sense the atrial mechanical events and deliver leadless AV-synchronous ventricular pacing (VDD). 2 The Micra AV therefore has the potential to provide VDD pacing with a lower rate of pocket/lead-related complications vs transvenous systems. Pacing-facilitated ventricular arrhythmias have previously been reported after Micra implantation.3, 4, 5 Here, we report the first case of bradycardia-dependent polymorphic ventricular tachycardia (PMVT) caused by atrial undersensing in a patient implanted with the Micra AV leadless pacemaker. Case report A 92-year-old woman with history of severe aortic stenosis, hypothyroidism, type 2 diabetes mellitus, and recurrent urinary tract infections presented to the hospital with progressive fatigue and dyspnea over a 2-week period and found to be in sinus rhythm with complete heart block and stable junctional escape rhythm (Supplemental Figure 1a). Blood biochemistry was within normal limits, as were high-sensitivity cardiac biomarkers and thyroid-stimulating hormone levels. Transthoracic echocardiogram showed mild left ventricular hypertrophy with preserved left ventricular systolic function. Considering her infectious risk and comorbidities but intact sinus node function, she was referred for Micra AV insertion. The patient was brought to our electrophysiology laboratory in the fasting state. Under conscious sedation, ultrasound-guided right femoral venous access was obtained and upsized to accept the 27 French delivery system. The Micra AV was successfully deployed on the right ventricular mid-septum (Figure 1) using the standard implant technique. After sheath removal, hemostasis was secured with a hemostatic silk suture and manual compression; there were no acute complications. The pacemaker was programmed to VDDR 60–105 beats per minute (bpm) and found to have satisfactory acute implant parameters (impedance 710 W; R wave 8.3 mV; ventricular capture threshold 0.38 V at 0.24 ms). Furthermore, the leadless pacemaker appropriately sensed the atrial mechanical events and delivered AV synchronous ventricular pacing. She was returned to her inpatient bed in stable condition. Figure 1 Thoracic fluoroscopy view depicting the Micra delivery system after deployment of the Micra device (Medtronic, Minneapolis, MN). View is confirming proper placement at the septal aspect of the right ventricle on a: standard left anterior oblique view and b: right anterior oblique view. Overnight, the patient suffered a syncopal event while supine with seizure-like movements for about 12 seconds correlating with an episode of PMVT (Figure 2); the tachycardia terminated spontaneously. Investigations including blood biochemistry, cardiac enzymes, a transthoracic echocardiogram, and noninvasive ischemia assessment were unremarkable. Figure 2 Telemetry strip showing initiation of polymorphic ventricular tachycardia following a paced beat (blue arrows) after premature ventricular contraction (orange arrows). Dissociated atrial activity is marked with green arrows. Accelerometer sensor failed to detect the mechanical atrial activity and led to premature pacing at the lower programmed rate. Figure shows the short-long-short setup for polymorphic arrhythmia in this patient. We felt that undersensing of the A wave led to paced beat at the lower VVI mode rate followed by premature ventricular contractions (PVCs), which triggered a short-long-short (S-L-S) status leading to the polymorphic arrhythmia. Pacing mode was kept as VDD but the A4 (atrium active contraction) phase sensing property was adjusted, and we added beta blockers to her medications. Intervention resulted in a lower sinus node rate, no further PVCs, and better atrial activity sensing (Figure 3). Figure 3 A: The correlation between recorded surface electrocardiogram and accelerometer sensing of ventricular mechanical changes during different phases: A3: passive ventricular filling, A4: active atrium contraction. Shown is a successful electromechanical correlation after adjusting A4 sensing and treatment with beta blockers. PVAB = postventricular atrial blanking period. B: Changes in accelerometer sensing threshold over time in this patient. Improper sensing after implantation is noted (threshold > 1.6 m/s2). It also shows improvement in the A4 acceleration sensing after adjustments made to sensing parameters and titrating up beta blocker dose over time. Repeat ECG showed well-synchronized paced ventricular rhythm (Supplemental Figure 1b) and no further arrhythmia was noted on telemetry monitoring. She recovered and was discharged home to the care of her family. Discussion Novel technologies offer new opportunities and challenges. The recently approved Micra AV pacemaker represents a major development in leadless pacing, having the potential of delivering leadless AV-synchronous ventricular pacing (VDD). Here, we report the first case of bradycardia-dependent PMVT caused by atrial undersensing in a patient implanted with the Micra AV leadless pacemaker. Single-chamber VVI leadless pacing is commonly used in clinical practice and associated with lower pocket- and lead-related complications when compared to transvenous pacing systems. 1 , 6 , 7 Until very recently, leadless pacing with the Micra device (Model MC1VR01) was limited by single-chamber ventricular sensing, making it unsuitable for patients in which AV synchrony is desired. The Micra AV (Model MC1AVR1) uses its accelerometer to sense the atrial mechanical events in sinus rhythm to deliver atrial-triggered ventricular pacing (VVD), thereby preserving AV synchrony. The implant technique and delivery system for Micra AV are the same as for Micra such that Micra implanters can seamlessly opt for the Micra AV system on the basis of the clinical pacing indication. The Micra Atrial tracking using a Ventricular accelerometer 2 (MARVEL 2) study compared leadless VDD vs VVI pacing with the Micra AV in 75 patients with sinus rhythm and complete heart block. Atrioventricular synchrony increased from 26.8% to 89.2% with VVI vs VDD pacing, respectively, and no serious complications were reported. 2 The benefits of atrial-based vs ventricular-based pacing beyond a mild reduction in atrial fibrillation have not been consistently demonstrated, whereas the acute and chronic complication rates are definitely higher with dual-chamber transvenous systems. 8 Nevertheless, the prospect of reliable leadless VDD pacing capabilities will likely significantly expand the indications for leadless pacemakers. The potential proarrhythmic effect of cardiac pacing is a rare but well-described phenomenon with transvenous pacemakers. 9 Conventional pacing modes such as VVI/R and DDD/R have been observed to facilitate the induction of ventricular tachycardia (VT) and fibrillation (VF). In a combined post hoc analysis of the PainFree Rx II and EnTrust trials, Sweeney and colleagues 10 found that pacing-associated S-L-S sequences were observed in 29.8% of VT/VF events and pacing was adjudicated to be causal in 2.6% to 5.2% of VT/VF episodes. Algorithms to minimize ventricular pacing, such as managed ventricular pacing (Medtronic, Inc, Minneapolis, MN), have also been observed to facilitate and even trigger VT/VF, albeit in an exceedingly small minority of patients. 11 , 12 Several hypotheses have been proposed to account for “R-on-T” or S-L-S-mediated PMVT. At a tissue level, alternations in activation cycle length between paced complexes and PVCs increase the heterogeneity of repolarization, as repolarization is a rate-dependent process, setting the stage for ventricular arrhythmias. 13 An appropriately timed PVC in the vulnerable period of ventricular repolarization can find parts of the myocardium available for conduction while others may be refractory, leading to functional reentry and PMVT/VF. Fortunately, most patients do not experience such events, which raises the possibility of subclinical defects in repolarization being unmasked by additional stressors (eg, acute bradycardia, electrolyte abnormalities, ischemia) in those patients that do suffer from S-L-S-mediated PMVT. Leadless pacemaker–facilitated ventricular arrhythmias have previously been reported with the Micra pacing system.3, 4, 5 Da Costa and colleagues 3 reported a case of VF temporally associated to Micra insertion but without documenting the mechanism of initiation of tachycardia such that a causal relationship cannot be ascertained. More recently, Amin and colleagues 5 and Olsen and colleagues 4 each reported a case of Micra-facilitated sustained ventricular arrhythmias, the former managed with repositioning the Micra and the latter requiring deactivation of the Micra and implantation of a transvenous pacemaker. The number of cases of leadless pacemaker–facilitated or leadless pacemaker–triggered ventricular arrhythmias remains small. However, this may underestimate the true incidence of such events, as the Micra devices are not designed to store arrhythmia events; hence the events could therefore go undetected. Our case is different and highlights a complication associated with leadless VDD pacing, a novel technology. In our patient, atrial undersensing during VDD pacing led to S-L-S sequences or “R-on-T,” ultimately triggering PMVT. The phenomenon of atrial undersensing leading to PMVT with VDD pacing has previously been described with transvenous pacing systems. 14 The induction of PMVT from an S-L-S sequence is likely a low-probability stochastic event. However, the rate of inappropriate atrial sensing is significantly higher with the Micra AV than with transvenous VDD systems, potentially exposing patients to a higher burden of S-L-S sequences. In fact, in the MARVEL 2 study, AV synchrony was present in 89.2% of patients after a 20-minute rest period but was as low as 69.8% while standing. 2 Relying on the mechanical forces of the right atrium (RA) leaves this technology vulnerable to RA hemodynamic changes. Enlarged RA, hemodynamic acute shifts, and atrial tachycardia/arrhythmia may compromise the A4 wave sensed by the accelerometer and the device will switch to VVI mode. Therefore, patient selection should consider all the contributing factors. Further work is needed to ascertain the burden of S-L-S and risk of VT/VF caused by atrial undersensing with the Micra AV pacemaker. Arrhythmia electrogram storage should be within the scope of the available technology and would be very useful in better defining the incidence of atrial undersensing–induced ventricular arrhythmias with the Micra AV. Elsokkari and colleagues 15 reported a case of PMVT in a patient with complete heart block switched from VDD 50-120 bpm to VVI 50 bpm transvenous pacing. They noted that this change in pacing mode was the functional equivalent of the patient having had de novo AV junction ablation. This empiric observation forms the basis for a higher backup pacing rate after AV junction ablation. A similar sequence of events was observed in our patient in that VDD pacing at the sinus rate was replaced by functional VVI pacing at the lower rate interval because of atrial undersensing. This abrupt change in rate may have acted as an additional stressor to the patient’s ventricular repolarization homeostatic mechanisms. This also highlights the limitations of the accelerometer-based sensing of atrial mechanical events. It is highly likely that this technology will be further improved with subsequent iterations of the device. Conclusion Novel technologies offer new opportunities and challenges. Here, we report the first case of bradycardia-dependent PMVT caused by atrial undersensing in a patient implanted with the Micra AV leadless pacemaker.

          Related collections

          Most cited references12

          • Record: found
          • Abstract: found
          • Article: not found

          Incidence and predictors of short- and long-term complications in pacemaker therapy: the FOLLOWPACE study.

          Today quantitative information about the type of complications and their incidence during long-term pacemaker (PM) follow-up is scarce. To assess the incidence and determinants of short- and long-term complications after first pacemaker implantation for bradycardia. A prospective multicenter cohort study (the FOLLOWPACE study) was conducted among 1517 patients receiving a PM between January 2003 and November 2007. The independent association of patient and implantation-procedure characteristics with the incidence of PM complications was analyzed using multivariable Cox regression analysis. A total of 1517 patients in 23 Dutch PM centers were followed for a mean of 5.8 years (SD 1.1), resulting in 8797 patient-years. Within 2 months, 188 (12.4%) patients developed PM complications. Male gender, age at implantation, body mass index, a history of cerebrovascular accident, congestive heart failure, use of anticoagulant drugs, and passive atrial lead fixation were independent predictors for complications within 2 months, yielding a C-index of 0.62 (95% confidence interval 0.57-0.66). Annual hospital implanting volume did not additionally contribute to the prediction of short-term complications. Thereafter, 140 (9.2%) patients experienced complications, mostly lead-related complications (n = 84). Independent predictors for long-term complications were age, body mass index, hypertension, and a dual-chamber device, yielding a C-index of 0.62 (95% confidence interval 0.57-0.67). The occurrence of a short-term PM complication was not predictive of future PM complications. Complication incidence in modern pacing therapy is still substantial. Most complications occur early after PM implantation. Although various patient- and procedure-related characteristics are independent predictors for early and late complications, their ability to identify the patient at high risk is rather poor. This relatively high incidence of PM complications and their poor prediction underscores the usefulness of current guidelines for regular follow-up of patients with PM. Copyright © 2012 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found
            Is Open Access

            Atrioventricular synchronous pacing using a leadless ventricular pacemaker: Results from the MARVEL 2 study

            This study reports on the performance of a leadless ventricular pacemaker with automated, enhanced accelerometer-based algorithms that provide atrioventricular (AV) synchronous pacing.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Cardiovascular outcomes with atrial-based pacing compared with ventricular pacing: meta-analysis of randomized trials, using individual patient data.

              Several randomized trials have compared atrial-based (dual-chamber or atrial) pacing with ventricular pacing in patients with bradycardia. No trial has shown a mortality reduction, and only 1 small trial suggested a reduction in stroke. The goal of this review was to determine whether atrial-based pacing prevents major cardiovascular events. A systematic review was performed of publications since 1980. For inclusion, trials had to compare an atrial-based with a ventricular-based pacing mode; use a randomized, controlled, parallel design; and have data on mortality, stroke, heart failure, or atrial fibrillation. Individual patient data were obtained from 5 of the 8 identified studies, representing 95% of patients in the 8 trials, and a total of 35 000 patient-years of follow-up. There was no significant heterogeneity among the results of the individual trials. There was no significant reduction in mortality (hazard ratio [HR], 0.95; 95% confidence interval [CI], 0.87 to 1.03; P=0.19) or heart failure (HR, 0.89; 95% CI, 0.77 to 1.03; P=0.15) with atrial-based pacing. There was a significant reduction in atrial fibrillation (HR, 0.80; 95% CI, 0.72 to 0.89; P=0.00003) and a reduction in stroke that was of borderline significance (HR, 0.81; 95% CI, 0.67 to 0.99; P=0.035). There was no convincing evidence that any patient subgroup received special benefit from atrial-based pacing. Compared with ventricular pacing, the use of atrial-based pacing does not improve survival or reduce heart failure or cardiovascular death. However, atrial-based pacing reduces the incidence of atrial fibrillation and may modestly reduce stroke.
                Bookmark

                Author and article information

                Contributors
                Journal
                HeartRhythm Case Rep
                HeartRhythm Case Rep
                HeartRhythm Case Reports
                Elsevier
                2214-0271
                18 May 2020
                August 2020
                18 May 2020
                : 6
                : 8
                : 503-506
                Affiliations
                [1]Cardiac Electrophysiology Department, Brigham and Women’s Hospital, Harvard Medical School, Boston, Massachusetts
                Author notes
                [] Address reprint requests and correspondence: Dr Ahmad Halawa, Clinical Cardiac Electrophysiology Department, Brigham and Women’s Hospital, 3rd Floor, Watkins Cardiovascular Center, Suite C, Carl J. and Ruth Shapiro Cardiovascular Center, 70 Francis Street, Boston, MA 02115. ahalawa@ 123456bwh.harvard.edu
                Article
                S2214-0271(20)30089-0
                10.1016/j.hrcr.2020.05.003
                7424308
                32817828
                9aed81c7-0f9f-4537-b971-5ac4feac3116
                © 2020 Heart Rhythm Society. Published by Elsevier Inc.

                This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).

                History
                Categories
                Case Report

                atrioventricular synchrony,leadless pacemaker,polymorphic ventricular tachycardia,short-long-short,syncope

                Comments

                Comment on this article