      Novel dry electrodes for ECG monitoring.

      Physiological measurement
      Artifacts, Cardiac Electrophysiology, instrumentation, Elasticity, Electric Capacitance, diagnostic use, Electric Impedance, Electrocardiography, Ambulatory, Electrodes, Equipment Design, Equipment Failure Analysis, Evoked Potentials, physiology, Humans

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          The development, fabrication and characterization of two novel dry bioelectrodes--conductive and capacitive ones--for biopotential monitoring are presented. The new electrodes have the potential to improve the applicability of dry electrodes in ambulant recording of ECG by reducing motion artifacts as well as the contact impedance to the skin. Furthermore, a passive filter network is integrated into the new electrodes to suppress slow offset fluctuation of the ECG signal caused e.g. by motions like breathing or changes in the electrode-skin interface properties. Compared to standard gel electrodes these new electrodes exhibit equivalent and superior contact impedances and biosignals. The integrated filter network effectively suppresses fluctuating offset potentials.