      Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology.

      Oral surgery, oral medicine, oral pathology, oral radiology, and endodontics

      Cone-Beam Computed Tomography, instrumentation, Humans, Male, Phantoms, Imaging, Practice Guidelines as Topic, Radiography, Dental, methods, Reference Values, Skull, radiography, Thermoluminescent Dosimetry, statistics & numerical data, Tomography, X-Ray Computed, United States

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          This study compares 2 measures of effective dose, E(1990) and E(2007), for 8 dentoalveolar and maxillofacial cone-beam computerized tomography (CBCT) units and a 64-slice multidetector CT (MDCT) unit. Average tissue-absorbed dose, equivalent dose, and effective dose were calculated using thermoluminescent dosimeter chips in a radiation analog dosimetry phantom. Effective doses were derived using 1990 and the superseding 2007 International Commission on Radiological Protection (ICRP) recommendations. Large-field of view (FOV) CBCT E(2007) ranged from 68 to 1,073 microSv. Medium-FOV CBCT E(2007) ranged from 69 to 560 microSv, whereas a similar-FOV MDCT produced 860 microSv. The E(2007) calculations were 23% to 224% greater than E(1990). The 2007 recommendations of the ICRP, which include salivary glands, extrathoracic region, and oral mucosa in the calculation of effective dose, result in an upward reassessment of fatal cancer risk from oral and maxillofacial radiographic examinations. Dental CBCT can be recommended as a dose-sparing technique in comparison with alternative medical CT scans for common oral and maxillofacial radiographic imaging tasks.

          Related collections

          Author and article information

          Journal
          18504152
          10.1016/j.tripleo.2008.03.018

          Comments

          Comment on this article