      Aging stability of complete blood count and white blood cell differential parameters analyzed by Abbott CELL-DYN Sapphire hematology analyzer.

      International Journal of Laboratory Hematology

      Time Factors, Specimen Handling, Reproducibility of Results, Protein Stability, methods, instrumentation, Leukocyte Count, Humans, Blood Cell Count, Autoanalysis

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          This study presents the results of an aging stability study of complete blood count (CBC) and leukocyte differential parameters using the Abbott CELL-DYN Sapphire hematology analyzer. Stability studies showed no substantial change in CBC parameters up to 24-48 h at +23 +/- 2 degrees C (room temperature), except for optical platelet count (PLTo). For specimens aged over 24, the value of impedance platelet count yielded more reliable results than the routine PLTo. White blood cell (WBC) differential parameters, except eosinophils, were stable for up to 48 h at +23 +/- 2 degrees C. CBC parameters were stable for 72 h, except mean platelet volume, which slightly increased between 48 and 72 h, at +4 degrees C. WBC differentials were stable 48-72 h, with a slight decrease observed in absolute neutrophils and lymphocytes at +4 degrees C.

          Related collections

          Author and article information

          Journal
          18190587
          10.1111/j.1751-553X.2007.01009.x

          Comments

          Comment on this article