Whenzheng Li , MD a , Fang Li , BS a , He Wang , PHD b , Xueying Long , MD a , Obin Ghimire , MD a , Yigang Pei , MD a , ∗ , Xiangcheng Xiao , MD c , Jianping Ning , MD c
30 June 2017
chronic dialysis catheter, HR-MRCP, HR-T2WI, thrombotic complications, tip position
Catheter-related thrombotic complications(TCs) can occur during the long term use of a chronic dialysis catheter (CDC), including fibrin sheath (FS), mural thrombosis (MT), venous thrombosis (VT), and intraluminal clots (IC), which has not been reported with MRI. The aim of our study was to evaluate the determination of catheter tip position (TP) and resolution of TCs in patients with transjugular CDC scheduled into the superior vena cava using high resolution magnetic resonance cholangiopancreatography (HR-MRCP) and T 2-weighted imaging (HR-T 2WI).
The study protocol was approved by the local Research Ethics Committee. Informed consent was obtained from all patients. In total, 41 consecutively enrolled transjugular CDC patients with suspected catheter dysfunction were scanned with HRMRCP and HR-T 2WI. The distance from the top to the tip of the catheter and the presence and nature of catheter TCs were assessed by 2 experienced radiologists. Chest x-ray was taken within 1 to 2 days and CDC was withdrawn within 3 to 10 days from those patients with TCs identified by HR-MRI.
A total of 38 subjects successfully underwent HR-MRI, including 13 normal and 25 with TCs (fibrin sheath [FS]: n = 21, mural thrombosis [MT]: n = 7, venous thrombosis [VT]: n = 3, intraluminal clots [IC]: n = 4). There was no significant difference between HR-MRCP and chest x-ray in catheter TP determination ( P = .124). Normal catheter appeared as “double eyes” on HR-T 2WI and “double tracks” on HR-MRCP. TCs appeared as follows: FS displayed as a “thin ring” (<1mm) around the catheter, MT as patchy hyperintensity and VT as a “thick ring” (>5mm) on HR-T 2WI. Unilateral IC appeared as a “single eye” on HR-T 2WI and a “single track” on HR-MRCP (n = 3). Bilateral IC appeared as neither “eye” nor “track” (n = 1). Catheter withdrawal confirmed FS (n = 16), MT (n = 6), VT (n = 1), and IC (n = 4).
This is an open access article distributed under the Creative Commons Attribution-Share Alike License 4.0, which allows others to remix, tweak, and build upon the work, even for commercial purposes, as long as the author is credited and the new creations are licensed under the identical terms. http://creativecommons.org/licenses/by-sa/4.0