      Resolution of T-wave oversensing with implantable cardioverter defibrillator generator replacement in an adolescent.

      1
      Pacing and clinical electrophysiology : PACE
      Wiley

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Inappropriate delivery of therapy due to T-wave oversensing is a frequent complication of intracardiac cardioverter defibrillators (ICD). A 16-year-old adolescent with congenital long QT syndrome type 3 underwent single-chamber implantable cardioverter defibrillator (ICD) (EnTrust VR, Medtronic Inc., Minneapolis, MN, USA) implantation.He received two appropriate and four inappropriate shocks. The inappropriate shocks were due to T-wave oversensing. Following the replacement of ICD with a different brand generator (Vitality 2 VR, Boston Scientific Inc., Natick, MA, USA), no more T-wave oversensing is noted. Proper filtering is imperative in ICDs in order to prevent T-wave oversensing-related inappropriate delivery of therapy.

          Related collections

          Author and article information

          Journal
          Pacing Clin Electrophysiol
          Pacing and clinical electrophysiology : PACE
          Wiley
          0147-8389
          0147-8389
          Jul 2007
          : 30
          : 7
          Affiliations
          [1 ] Arkansas Children's Hospital, University of Arkansas for Medical Sciences, Little Rock, Arkansas 72202, USA. tuzcuvolkan@uams.edu
          Article
          PACE785
          10.1111/j.1540-8159.2007.00785.x
          17584279
          a5e99b22-21d6-4c85-a5fe-69afc00769f9
          History

          Comments

          Comment on this article