      Field mapping without reference scan using asymmetric echo-planar techniques.

      Magnetic Resonance in Medicine
      Echo-Planar Imaging, instrumentation, methods, statistics & numerical data, Fourier Analysis, Humans, Phantoms, Imaging, Software, Time Factors

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Improvements in Bo mapping and shimming were achieved by measuring the static field information in multiple subsequent echoes generated by an asymmetric echo-planar readout gradient train. With careful compensation, eddy current effects were shown to affect the adjustment of the shim coils minimally. In addition to reducing the time required for field mapping by two-fold, the sensitivity was simultaneously optimized irrespective of the prevalent T2* present, thereby minimizing the error of the static field measurement to below 0.1 Hz. With adiabatic low flip-angle excitation, the time required for field mapping was below 1 second.