      Experimental validation of intact and implanted distal femur finite element models.

      Journal of Biomechanics
      Biomechanical Phenomena, Cementation, Femur, Finite Element Analysis, Humans, Knee Prosthesis, Materials Testing, Stress, Mechanical

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Four finite element (FE) models of intact and distal femur of knee replacements were validated relative to measured bone strains. FE models of linear tetrahedrons were used. Femoral replacements with cemented stemless, cemented and noncemented femoral stems of the PFC Sigma Modular Knee System were analyzed. Bone strains were recorded at ten locations on the cortex. The magnitude of the FE bone strains corresponded to the mean measured strains, with an overall agreement of 10%. Linear regression between the FE and mean experimental strains produced slopes between 0.94 and 1.06 and R(2) values between 0.92 and 0.99. RSME values were less than 12%. The FE models were able to adequately replicate the mechanical behavior of distal femur reconstructions.