      Prosthesis-patient mismatch after mitral valve replacement: a single-centered retrospective analysis in East China

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          Prosthesis–patient mismatch (PPM) may affect the clinical outcomes of patients undergoing mitral valve replacement (MVR) surgery. We aimed to investigate the incidence of PPM of the mitral position in our center and analyze the possible predictors of PPM as well as its effect on short-term outcomes.

          Methods

          We retrospectively examined all consecutive patients with isolated or concomitant MVR at our center from 2013 to 2015. PPM was defined as an indexed effective orifice area (iEOA) of ≤1.2 cm2/m2. After inclusion and exclusion, a total of 1067 patients were analyzed. The baseline information were collected and compared between the two groups. Multivariate logistic regression analysis was conducted to determine the preoperative predictors of PPM as well as the effect of PPM on early mortality.

          Results

          A total of 1067 patients were included in the study. PPM was detected in 15.9% of the patients while 12 patients (1.12%) met the criteria for severe PPM. Patients with PPM compared to the non-PPM patients had higher age, larger body surface area and were more likely to be male and obese. Logistic regression analysis showed that higher age, larger BSA, bioprosthesis and smaller left ventricle end-diastolic diameter were predictors of PPM. There were no significant differences between the PPM and non-PPM groups regarding post-operative complications. Logistic regression analysis showed that PPM was not a risk factor of short-term mortality ( P = 0.654). Also, there were no significant differences regarding short−/mid-term heart function between the PPM and non PPM groups ( P = 0.902).

          Conclusions

          Our results demonstrated that higher age, bioprosthesis, larger BSA and smaller left ventricle size were associated with mitral PPM. However, PPM was not associated with poorer early outcomes after MVR surgery. In eastern of China, the prevalence of mitral valve stenosis is high; therefore, whether the standard PPM criteria are suitable for patients of this district needs to be further verified.

          Related collections

          Most cited references22

          • Record: found
          • Abstract: found
          • Article: not found

          The problem of valve prosthesis-patient mismatch.

          Valve prostheses have played an important part in the past two decades in the management of patients with valvular heart disease. However, many of the devices used in valve replacement have introduced new clinical problems. This paper deals with some of the problems associated with valve replacement, including one not previously emphasized--valve prosthesis-patient mismatch, which may cause obstruction to ventricular outflow and/or inflow.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Prosthesis-patient mismatch: definition, clinical impact, and prevention.

            Prosthesis-patient mismatch (PPM) is present when the effective orifice area of the inserted prosthetic valve is too small in relation to body size. Its main haemodynamic consequence is to generate higher than expected gradients through normally functioning prosthetic valves. This review updates the present knowledge about the impact of PPM on clinical outcomes. PPM is common (20-70% of aortic valve replacements) and has been shown to be associated with worse haemodynamic function, less regression of left ventricular hypertrophy, more cardiac events, and lower survival. Moreover, as opposed to most other risk factors, PPM can largely be prevented by using a prospective strategy at the time of operation.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Impact of prosthesis-patient mismatch on survival after mitral valve replacement.

              We recently reported that valve prosthesis-patient mismatch (PPM) is associated with persisting pulmonary hypertension after mitral valve replacement. Thus, the objective of this study was to evaluate the impact of PPM on mortality in patients undergoing mitral valve replacement. The indexed valve effective orifice area was estimated for each type and size of prosthesis being implanted in 929 consecutive patients and used to define PPM as not clinically significant if > 1.2 cm2/m2, as moderate if > 0.9 and < or = 1.2 cm2/m2, and as severe if < or = 0.9 cm2/m2. Moderate PPM was present in 69% of patients; severe PPM was seen in 9%. For patients with severe PPM, 6-year survival (74+/-5%) and 12-year survival (63+/-7%) were significantly less than for patients with moderate PPM (84+/-1% and 76+/-2%; P=0.027) or nonsignificant PPM (90+/-2% and 82+/-4%; P=0.002). On multivariate analysis, severe PPM was associated with higher mortality (hazard ratio, 3.2; 95% confidence interval, 1.5 to 6.8; P=0.003). Severe PPM is an independent predictor of mortality after mitral valve replacement. As opposed to other independent risk factors, PPM may be avoided or its severity may be reduced with the use of a prospective strategy at the time of operation. For patients identified as being at risk for severe PPM, every effort should be made to implant a prosthesis with a larger effective orifice area.
                Bookmark

                Author and article information

                Contributors
                maike1524@126.com
                doctorzjn@126.com
                0086-571-87236843 , zhengjunnan@zju.edu.cn
                188551058@qq.com
                Journal
                J Cardiothorac Surg
                J Cardiothorac Surg
                Journal of Cardiothoracic Surgery
                BioMed Central (London )
                1749-8090
                3 October 2018
                3 October 2018
                2018
                : 13
                : 100
                Affiliations
                ISNI 0000 0004 1803 6319, GRID grid.452661.2, Department of Cardiothoracic Surgery, , the First Affiliated Hospital of Zhejiang University, ; No.79 Qingchun Road, Hangzhou, 310003 China
                Article
                788
                10.1186/s13019-018-0788-4
                6169081
                30285794
                ad1b3735-01c2-41b3-b9d5-b0d963ced381
                © The Author(s). 2018

                Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License ( http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver ( http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

                History
                : 14 May 2018
                : 25 September 2018
                Categories
                Research Article
                Custom metadata
                © The Author(s) 2018

                Surgery
                prosthesis-patient mismatch,mitral valve replacement,effective orifice area,short-term mortality

                Comments

                Comment on this article