      GPU-based fast cone beam CT reconstruction from undersampled and noisy projection data via total variation.

      Medical physics
      Algorithms, Artifacts, Child, Computer Graphics, Computer Simulation, Cone-Beam Computed Tomography, methods, Humans, Image Processing, Computer-Assisted, Models, Statistical, Phantoms, Imaging, Radiographic Image Interpretation, Computer-Assisted, Radiotherapy, Radiotherapy Planning, Computer-Assisted, X-Rays

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Cone-beam CT (CBCT) plays an important role in image guided radiation therapy (IGRT). However, the large radiation dose from serial CBCT scans in most IGRT procedures raises a clinical concern, especially for pediatric patients who are essentially excluded from receiving IGRT for this reason. The goal of this work is to develop a fast GPU-based algorithm to reconstruct CBCT from undersampled and noisy projection data so as to lower the imaging dose. The CBCT is reconstructed by minimizing an energy functional consisting of a data fidelity term and a total variation regularization term. The authors developed a GPU-friendly version of the forward-backward splitting algorithm to solve this model. A multigrid technique is also employed. It is found that 20-40 x-ray projections are sufficient to reconstruct images with satisfactory quality for IGRT. The reconstruction time ranges from 77 to 130 s on an NVIDIA Tesla C1060 (NVIDIA, Santa Clara, CA) GPU card, depending on the number of projections used, which is estimated about 100 times faster than similar iterative reconstruction approaches. Moreover, phantom studies indicate that the algorithm enables the CBCT to be reconstructed under a scanning protocol with as low as 0.1 mA s/projection. Comparing with currently widely used full-fan head and neck scanning protocol of approximately 360 projections with 0.4 mA s/projection, it is estimated that an overall 36-72 times dose reduction has been achieved in our fast CBCT reconstruction algorithm. This work indicates that the developed GPU-based CBCT reconstruction algorithm is capable of lowering imaging dose considerably. The high computation efficiency in this algorithm makes the iterative CBCT reconstruction approach applicable in real clinical environments.

          Related collections

          Most cited references15

          • Record: found
          • Abstract: not found
          • Article: not found

          Compressed sensing

            Bookmark
            • Record: found
            • Abstract: not found
            • Article: not found

            Near-Optimal Signal Recovery From Random Projections: Universal Encoding Strategies?

              Bookmark
              • Record: found
              • Abstract: not found
              • Article: not found

              Signal Recovery by Proximal Forward-Backward Splitting

                Bookmark

                Author and article information

                Comments

                Comment on this article