      Mechanical assessment of local bone quality to predict failure of locked plating in a proximal humerus fracture model.

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          The importance of osteoporosis in proximal humerus fractures is well recognized. However, the local distribution of bone quality in the humeral head may also have a significant effect because it remains unclear in what quality of bone screws of standard implants purchase. The goal of this study was to investigate whether the failure of proximal humerus locked plating can be predicted by the DensiProbe (ARI, Davos, Switzerland). A 2-part fracture with metaphyseal impaction was simulated in 12 fresh-frozen human cadaveric humeri. Using the DensiProbe, local bone quality was determined in the humeral head in the course of 6 proximal screws of a standard locking plate (Philos; Synthes GmbH, Solothurn, Switzerland). Cyclic mechanical testing with increasing axial loading until failure was performed. Bone mineral density (BMD) significantly correlated with cycles until failure. Head migration significantly increased between 1000 and 2000 loading cycles and significantly correlated with BMD after 3000 cycles. DensiProbe peak torque in all screw positions and their respective mean torque correlated significantly with the BMD values. In 3 positions, the peak torque significantly correlated with cycles to failure; here BMD significantly influenced mechanical stability. The validity of the DensiProbe was proven by the correlation between its peak torque measurements and BMD. The correlation between the peak torque and cycles to failure revealed the potential of the DensiProbe to predict the failure of locked plating in vitro. This method provides information about local bone quality, potentially making it suitable for intraoperative use by allowing the surgeon to take measures to improve stability.

          Related collections

          Author and article information

          Journal
          Orthopedics
          Orthopedics
          SLACK, Inc.
          1938-2367
          0147-7447
          Sep 2013
          : 36
          : 9
          Article
          10.3928/01477447-20130821-14
          24025003
          b3060a08-9c13-41d4-8e6a-356cdc8f7eae
          Copyright 2013, SLACK Incorporated.
          History

          Comments

          Comment on this article