      Biomechanical Comparison of Integrated Fixation Cage Versus Anterior Cervical Plate and Cage in Anterior Cervical Corpectomy and Fusion (ACCF): A Finite Element Analysis

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          Anterior cervical plate and cage fixation system (ACPC) used in anterior cervical corpectomy and fusion (ACCF) is reported to incur excess complications. This study aimed to introduce integrated fixation cage (IFC) into ACCF to eliminate the anterior cervical plate (ACP)-related complications.

          Material/Methods

          One validated intact and 3 ACCF-simulated C3–C7 cervical spine models were developed. In ACCF models, C5 was corpectomied and fixed by IFC or ACPC. For each model, 1.0 Nm moments of flexion, extension, lateral bending, and torsion were imposed on the C3 vertebra. The range of motion (ROM) of each segment and the stress distribution on screw-vertebra interface, bone graft, and cage-endplate were recorded and analyzed.

          Results

          ROMs of C3–C7 were not different in any motion condition between IFC and ACPC models. The maximal von Mises stress on screw-vertebra interface of the IFC model was lower than that of the ACPC models in flexion, extension, and lateral bending, but higher in rotation. The maximal von Mises stress on bone graft of the IFC model was higher compared with the ACPC models, except in flexion. The IFC model showed a higher maximal von Mises stress on cage-endplate interface in all motion planes.

          Conclusions

          Based on finite element analysis, IFC provided identical C3–C7 construct stability as ACPC. Compared with ACPC, IFC showed better biomechanical performance on screw-vertebra interface and bone graft, but worse biomechanical performance on cage-endplate interface.

          Related collections

          Most cited references31

          • Record: found
          • Abstract: found
          • Article: not found

          Wolff's Law and bone's structural adaptations to mechanical usage: an overview for clinicians.

          H. Frost (1994)
          Basic Multicellular Unit-based bone remodeling can lead to the removal or conservation of bone, but cannot add to it. Decreased mechanical usage (MU) and acute disuse result in loss of bone next to marrow; normal and hypervigorous MU result in bone conservation. Bone modeling by resorption and formation drifts can add bone and reshape the trabeculae and cortex to strengthen them but collectively they do not remove bone. Hypervigorous MU turns this modeling on, and its architectural effects then lower typical peak bone strains caused by future loads of the same kind to a threshold range. Decreased and normal MU leave this modeling off. Where typical peak bone strains stay below a 50 microstrain region (the MESr) the largest disuse effects on remodeling occur. Larger strains depress it and make it conserve existing bone. Strains above a 1500 microstrain region (the MESm) tend to turn lamellar bone modeling drifts on. By adding to, reshaping and strengthening bone, those drifts reduce future strains under the same mechanical loads towards that strain region. Strains above a 3000 microstrain region (the MESp) can turn woven bone drifts on to suppress local lamellar drifts but can strengthen bone faster than lamellar drifts can. Such strains also increase bone microdamage and the remodeling that normally repairs it. Those values compare to bone's fracture strain of about 25,000 microstrain.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Mechanical properties of the human cervical spine as shown by three-dimensional load-displacement curves.

            The mechanical properties of multilevel human cervical spines were investigated by applying pure rotational moments to each specimen and measuring multidirectional intervertebral motions. To document intervertebral main and coupled motions of the cervical spine in the form of load-displacement curves. Although a number of in vivo and in vitro studies have attempted to delineate normal movement patterns of the cervical spine, none has explored the complexity of the whole cervical spine as a three-dimensional structure. Sixteen human cadaveric specimens (C0-C7) were used for this study. Pure rotational moments of flexion-extension, bilateral axial torque, and bilateral lateral bending were applied using a specially designed loading fixture. The resulting intervertebral motions were recorded using stereophotogrammetry and depicted as a series of load-displacement curves. The resulting load-displacement curves were found to be nonlinear, and both rotation and translation motions were coupled with main motions. With flexion-extension moment loading, the greatest degree of flexion occurred at C1-C2 (12.3 degrees), whereas the greatest degree of extension was observed at C0-C1 (20.2 degrees). With axial moment loading, rotation at C1-C2 was the largest recorded (56.7 degrees). With lateral bending moments, the average range of motion for all vertebral levels was 7.9 degrees. The findings of the present study are relevant to the clinical practice of examining motions of the cervical spine in three dimensions and to the understanding of spinal trauma and degenerative diseases.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Comparison of Outcomes for Anterior Cervical Discectomy and Fusion With and Without Anterior Plate Fixation: A Systematic Review and Meta-Analysis.

              Systematic review and meta-analysis.
                Bookmark

                Author and article information

                Journal
                Med Sci Monit
                Med. Sci. Monit
                Medical Science Monitor
                Medical Science Monitor : International Medical Journal of Experimental and Clinical Research
                International Scientific Literature, Inc.
                1234-1010
                1643-3750
                2019
                25 February 2019
                : 25
                : 1489-1498
                Affiliations
                [1 ]Department of Orthopedic Surgery, The 2 nd Hospital of Xi’an Jiaotong University, Xi’an, Shaanxi, P.R. China
                [2 ]State Key Laboratory for Manufacturing Systems Engineering, School of Mechanical Engineering, Xi’an Jiaotong University, Xi’an, Shaanxi, P.R. China
                Author notes
                Corresponding Author: Xijing He, e-mail: xijing_h@ 123456vip.tom.com
                [A]

                Study Design

                [B]

                Data Collection

                [C]

                Statistical Analysis

                [D]

                Data Interpretation

                [E]

                Manuscript Preparation

                [F]

                Literature Search

                [G]

                Funds Collection

                Article
                913630
                10.12659/MSM.913630
                6400022
                30799836
                b3706105-dcf4-4f28-bd71-c0fe490cc043
                © Med Sci Monit, 2019

                This work is licensed under Creative Common Attribution-NonCommercial-NoDerivatives 4.0 International ( CC BY-NC-ND 4.0)

                History
                : 15 October 2018
                : 16 November 2018
                Categories
                Lab/In Vitro Research

                biomechanical phenomena,cervical vertebrae,finite element analysis,spinal fusion

                Comments

                Comment on this article