      Refractive prediction of four different intraocular lens calculation formulas compared between new swept source optical coherence tomography and partial coherence interferometry

      research-article
      , , *
      PLoS ONE
      Public Library of Science

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Purpose

          To compare the biometry and prediction of postoperative refractive outcomes of four different formulae (Haigis, SRK/T, Holladay1, Barrett Universal II) obtained by swept-source optical coherence tomography (SS-OCT) biometers and partial coherence interferometry (PCI; IOLMaster ver 5.4).

          Methods

          We compared the biometric values of SS-OCT (ANTERION, Heidelberg Engineering Inc., Heidelberg, Germany) and PCI (IOLMaster, Carl Zeiss Meditec, Jena, Germany). Predictive errors calculated using four different formulae (Haigis, SRKT, Holladay1, Barrett Universal II) were compared at 1 month after cataract surgery.

          Results

          The mean preoperative axial length (AL) showed no statistically significant difference between SS-OCT and PCI (SS-OCT: 23.78 ± 0.12 mm and PCI: 23.77 ± 0.12 mm). The mean anterior chamber depth (ACD) was 3.30 ± 0.04 mm for SS-OCT and 3.23 ± 0.04 mm for PCI, which was significantly different between the two techniques. The mean corneal curvature also differed significantly between the two techniques. The difference in mean arithmetic prediction error was significant in the Haigis, SRKT, and Holladay1 formulae. The difference in mean absolute prediction error was significant in all four formulae.

          Conclusions

          SS-OCT and PCI demonstrated good agreement on biometric measurements; however, there were significant differences in some biometric values. These differences in some ocular biometrics can cause a difference in refractive error after cataract surgery. New type SS-OCT was not superior to the IOL power prediction calculated by PCI.

          Related collections

          Most cited references26

          • Record: found
          • Abstract: not found
          • Article: not found

          Comparing methods of measurement: why plotting difference against standard method is misleading

            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis.

            The precision of intraocular lens (IOL) calculation is essentially determined by the accuracy of the measurement of axial length. In addition to classical ultrasound biometry, partial coherence interferometry serves as a new optical method for axial length determination. A functional prototype from Carl Zeiss Jena implementing this principle was compared with immersion ultrasound biometry in our laboratory. In 108 patients attending the biometry laboratory for planning of cataract surgery, axial lengths were additionally measured optically. Whereas surgical decisions were based on ultrasound data, we used postoperative refraction measurements to calculate retrospectively what results would have been obtained if optical axial length data had been used for IOL calculation. For the translation of optical to geometrical lengths, five different conversion formulas were used, among them the relation which is built into the Zeiss IOL-Master. IOL calculation was carried out according to Haigis with and without optimization of constants. On the basis of ultrasound immersion data from our Grieshaber Biometric System (GBS), postoperative refraction after implantation of a Rayner IOL type 755 U was predicted correctly within +/- 1 D in 85.7% and within +/- 2 D in 99% of all cases. An analogous result was achieved with optical axial length data after suitable transformation of optical path lengths into geometrical distances. Partial coherence interferometry is a noncontact, user- and patient-friendly method for axial length determination and IOL planning with an accuracy comparable to that of high-precision immersion ultrasound.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: found
              Is Open Access

              Evaluation and comparison of the new swept source OCT-based IOLMaster 700 with the IOLMaster 500

              Purpose To compare the measurements and failure rates obtained with a new swept source optical coherence tomography (OCT)-based biometry to IOLMaster 500. Setting Eye Clinic, Baskent University Faculty of Medicine, Ankara, Turkey. Design Observational cross-sectional study and evaluation of a new diagnostic technology. Methods 188 eyes of 101 subjects were included in the study. Measurements of axial length (AL), anterior chamber depth (ACD), corneal power (K1 and K2) and the measurement failure rate with the new Zeiss IOLMaster 700 were compared with those obtained with the IOLMaster 500. The results were evaluated using Bland–Altman analyses. The differences between both methods were assessed using the paired samples t test, and their correlation was evaluated by intraclass correlation coefficient (ICC). Results The mean age was 68.32±12.71 years and the male/female ratio was 29/72. The agreements between two devices were outstanding regarding AL (ICC=1.0), ACD (ICC=0.920), K1 (ICC=0.992) and K2 (ICC=0.989) values. IOLMaster 700 was able to measure ACD AL, K1 and K2 in all eyes within high-quality SD limits of the manufacturer. IOLMaster 500 was able to measure ACD in 175 eyes, whereas measurements were not possible in the remaining 13 eyes. AL measurements were not possible for 17 eyes with IOLMaster 500. Nine of these eyes had posterior subcapsular cataracts and eight had dense nuclear cataracts. Conclusions Although the agreement between the two devices was excellent, the IOLMaster 700 was more effective in obtaining biometric measurements in eyes with posterior subcapsular and dense nuclear cataracts.
                Bookmark

                Author and article information

                Contributors
                Role: MethodologyRole: VisualizationRole: Writing – original draft
                Role: Data curationRole: Formal analysisRole: ResourcesRole: Software
                Role: ConceptualizationRole: Data curationRole: Formal analysisRole: MethodologyRole: SupervisionRole: Writing – original draftRole: Writing – review & editing
                Role: Editor
                Journal
                PLoS One
                PLoS One
                plos
                plosone
                PLoS ONE
                Public Library of Science (San Francisco, CA USA )
                1932-6203
                4 May 2021
                2021
                : 16
                : 5
                : e0251152
                Affiliations
                [001]Kim’s Eye Hospital, Seoul, Korea
                University of Toronto, CANADA
                Author notes

                Competing Interests: The authors have declared that no competing interests exist.

                Author information
                https://orcid.org/0000-0003-4752-3098
                Article
                PONE-D-21-00885
                10.1371/journal.pone.0251152
                8096100
                33945581
                b44b3198-20fa-4174-89b2-ac74954cc84f
                © 2021 Song et al

                This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

                History
                : 10 January 2021
                : 20 April 2021
                Page count
                Figures: 3, Tables: 5, Pages: 13
                Funding
                The authors received no specific funding for this work.
                Categories
                Research Article
                Biology and Life Sciences
                Anatomy
                Head
                Eyes
                Medicine and Health Sciences
                Anatomy
                Head
                Eyes
                Biology and Life Sciences
                Anatomy
                Ocular System
                Eyes
                Medicine and Health Sciences
                Anatomy
                Ocular System
                Eyes
                Biology and Life Sciences
                Anatomy
                Ocular System
                Ocular Anatomy
                Cornea
                Medicine and Health Sciences
                Anatomy
                Ocular System
                Ocular Anatomy
                Cornea
                Physical Sciences
                Mathematics
                Statistics
                Biostatistics
                Research and Analysis Methods
                Computational Techniques
                Biometrics
                Medicine and Health Sciences
                Surgical and Invasive Medical Procedures
                Ophthalmic Procedures
                Cataract Surgery
                Biology and Life Sciences
                Anatomy
                Ocular System
                Ocular Anatomy
                Eye Lens
                Medicine and Health Sciences
                Anatomy
                Ocular System
                Ocular Anatomy
                Eye Lens
                Medicine and Health Sciences
                Diagnostic Medicine
                Diagnostic Radiology
                Tomography
                Research and Analysis Methods
                Imaging Techniques
                Diagnostic Radiology
                Tomography
                Medicine and Health Sciences
                Radiology and Imaging
                Diagnostic Radiology
                Tomography
                Physical Sciences
                Mathematics
                Arithmetic
                Custom metadata
                All relevant data are within the manuscript and its Supporting Information files.

                Uncategorized
                Uncategorized

                Comments

                Comment on this article