      Comparisons of perventricular device closure, conventional surgical repair, and transcatheter device closure in patients with perimembranous ventricular septal defects: a network meta-analysis

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          Treatments for perimembranous ventricular septal defects (pmVSD) mainly include conventional surgical repair (CSR), transcatheter device closure (TDC), and perventricular device closure (PDC). We aimed to perform a network meta-analysis to compare the three approaches in patients with pmVSD.

          Methods

          We searched for comparative studies on device closure and conventional repair for pmVSD to April 2020. A network meta-analysis was performed under the frequentist frame with risk ratio and 95% confidence interval. The main outcome was the procedural success rate. Additional outcomes were postoperative complications, including residual shunt, intra-cardiac conduction block, valvular insufficiency, incision infection, and pericardial effusion.

          Results

          Twenty-four studies of 8113 patients were included in the comparisons. The pooled estimates of success rate favored the CSR compared with the PDC. No significant differences of success rate were found in the TDC versus CSR and the PDC versus TDC. The pooled estimates of incidences of the residual shunt, new tricuspid regurgitation, incision infection, and pericardial effusion favored the PDC compared with the CSR. There were no significant differences between the PDC and TDC approaches in all outcomes except new aortic regurgitation.

          Conclusion

          The PDC technique not only reduces the risk of significant complications compared with the CSR, but also produces not inferior results compared with the TDC in selected pmVSD patients.

          PROSPERO registration number

          CRD42019125257.

          Related collections

          Most cited references44

          • Record: found
          • Abstract: not found
          • Article: not found

          Network Meta-analysis

            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Transcatheter closure of perimembranous ventricular septal defects: early and long-term results.

            We sought to analyze safety, efficacy, and follow-up results of percutaneous closure of perimembranous ventricular septal defects (pmVSD). Results of pmVSD transcatheter closure have been reported in the literature; however, follow-up data are still limited. Between January 1999 and June 2006, 104 patients underwent percutaneous closure of a pmVSD at our institution. An Amplatzer VSD device (muscular or eccentric) (AGA Medical Corp., Golden Valley, Minnesota) was used in all subjects. The mean age at closure was 14 years (range 0.6 to 63 years). The attempt to place a device was successful in 100 patients (96.2%). The median device size used was 8 mm (range 4 to 16 mm). No deaths occurred. Total occlusion rate was 47% at completion of the procedure, rising to 84% at discharge and 99% during the follow-up. A total of 13 early complications occurred (11.5%), but in all but 2 subjects (1.9%) these were transient. The median follow-up was 38.5 months. The most significant complication was complete atrioventricular block (cAVB), which required pacemaker implantation in 6 subjects (5.7%; 2 in the early phase and 4 during the follow-up). Cox proportional hazards regression analysis showed that the only variable significantly associated with the occurrence of this complication was age at the time of the procedure (p = 0.028; relative risk 0.25). All subjects experiencing this problem were <6 years old. In the current era and in experienced hands, pmVSD closure can be performed safely and successfully. The major concern is the occurrence of cAVB; therefore, very careful monitoring of rhythm is mandatory during follow-up.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              Transcatheter closure of perimembranous ventricular septal defects using the amplatzer membranous VSD occluder: immediate and midterm results of an international registry.

              To report the immediate and midterm results of transcatheter closure of perimembranous ventricular septal defect (PmVSD) using the Amplatzer membranous VSD occluder (AMVSD). Between April 2002 and August 2004, 100 patients underwent an attempt of percutaneous device closure of PmVSD using the AMVSD in 24 international centers. The median age was 9.0 years (0.7-58 years) and the median weight was 27.5 kg (7-121 kg). A device was successfully deployed in 93/100 (93%) patients. Reasons for procedural failure were an increased gradient across the left ventricle outflow tract in one patient, aortic regurgitation in 2 patients, and inability to securely position the device in 4 patients. The median VSD size by TEE was 7.0 mm (1.5-13 mm), median device size 10 mm (4-16 mm) and median fluoroscopy time 22.1 min (8.9-96.0 min). Weight below 10 kg (P = 0.0392), inlet extension of the VSD (P = 0.0139) and aortic cusp prolapse into the VSD (P = 0.0084) were significantly associated with a lower procedural success. Patients have been followed up for a median of 182 days (1-763 days). There were no procedure-related deaths. Complications were encountered in 29/100 (29%) patients, including rhythm or conduction anomalies in 13 patients (two with complete heart block requiring permanent pacemaker implantation), new or increased aortic (9 patients) or tricuspid (9 patients) regurgitation, most of which were classified as trivial or mild. Patients with a weight below 10 kg had a significantly higher incidence of adverse events than patients with a weight above 10 kg (58.3% versus 25.0%, P = 0.0285). Immediately after device release complete closure of the defect was present in 54/93 (58.1%) patients, increasing to 46/55 (83.6%) patients at 6-months follow-up (P = 0.0012). Left ventricle end-diastolic diameter decreased from a median of 44 mm prior to device closure to a median of 39 mm at 6-months postprocedure (P = 0.0015). Closure of PmVSDs using the AMVSD occluder is safe and effective. However, longer follow-up period is warranted prior to the wide spread use of this device.
                Bookmark

                Author and article information

                Contributors
                bohreinstein@126.com
                Journal
                BMC Surg
                BMC Surg
                BMC Surgery
                BioMed Central (London )
                1471-2482
                26 May 2020
                26 May 2020
                2020
                : 20
                : 115
                Affiliations
                [1 ]GRID grid.412901.f, ISNI 0000 0004 1770 1022, Department of Cardiovascular Surgery, , West China Hospital, Sichuan University, ; No. 37 Guo Xue Xiang, Chengdu, Sichuan 610041 P.R. China
                [2 ]GRID grid.411868.2, ISNI 0000 0004 1798 0690, Evidence-based Medicine Research Center, School of Basic Medical Sciences, , Jiangxi University of Traditional Chinese Medicine, ; Nanchang, Jiangxi P.R. China
                [3 ]GRID grid.412901.f, ISNI 0000 0004 1770 1022, Department of Anesthesiology, , West China Hospital, Sichuan University, ; Chengdu, Sichuan P.R. China
                Author information
                http://orcid.org/0000-0002-2630-4286
                Article
                777
                10.1186/s12893-020-00777-w
                7249310
                32456641
                b5b6bc52-1bd6-4ed5-9f7e-ce4f1f1e4526
                © The Author(s) 2020

                Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver ( http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

                History
                : 29 September 2019
                : 18 May 2020
                Funding
                Funded by: China Post-Doctoral Science Foundation
                Award ID: 2019TQ0219
                Award Recipient :
                Funded by: Post-Doctor Research Project, West China Hospital, Sichuan University
                Award ID: 2019HXBH019
                Award Recipient :
                Categories
                Research Article
                Custom metadata
                © The Author(s) 2020

                Surgery
                ventricular septal defect,conventional surgical repair,transcatheter device closure,perventricular device closure,network meta-analysis