      Performance testing and status analysis of nuclear medicine imaging device in Qinghai Province

      research-article

      Read this article at

      ScienceOpenPublisher
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Objective To analyze the quality control situation of nuclear medicine equipment in Qinghai Province by testing the performance of SPECT and PET/CT devices.

          Methods According to WS 523—2019, NEMA NU 2—2007 and GB 17589—2011 standard, the performanc indexes of four SPECTs and one PET/CT were measured in 2019 and 2020.

          Results For SPECTS, according to the qualified limits requirements of WS 523—2019 Standard, one device with intrinsic spatial linearity is unqualified. For PET/CT, the equipment performance test results in 2020 were better than those in 2019.

          Conclusion Regular calibration and maintenance are very important for the quality control of nuclear medicine equipment. Strengthening daily quality control can improve equipment performance and clinical application.

          Abstract

          摘要: 目的 通过对青海省核医学影像设备SPECT和PET/CT进行性能检测, 分析青海省核医学设备质量控制现 状; 方法 依据 WS523—2019、NEMA NU2—2007 和 GB 17589—2011 标准, 分别在 2019 年和 2020 年测量了 4 台 SPECT设备和 1 台PET/CT的性能指标; 结果 对于SPECT, 按照 WS523 标准对合格限值的要求, 2019 年有 1 台 SPECT的固有空间线性不合格, 对于PET/CT, 2020 年的设备性能检测结果优于 2019 的结果; 结论 定期的校正、保 养和维护对核医学设备的质量控制很重要, 加强日常质量控制, 可以提升设备性能, 为临床应用保驾护航。

          Related collections

          Author and article information

          Journal
          CJRH
          Chinese Journal of Radiological Health
          Chinese Preventive Medical Association (Ji’an, China )
          1004-714X
          01 December 2021
          01 January 2022
          : 30
          : 6
          : 697-700
          Affiliations
          [1] 1Qing Hai Center for Diseases Prevention and Control, Xining 810007 China
          [2] 2Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC. Notional Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088 China
          Author notes
          Corresponding author: LIU Hui, E-mail: liuhui@ 123456nirp.chinacdc.cn
          Article
          j.issn.1004-714X.2021.06.008
          10.13491/j.issn.1004-714X.2021.06.008
          b7e90e6b-cded-4906-ad44-d46b33b224d3
          © 2021 Chinese Journal of Radiological Health

          This is an open-access article distributed under the terms of the Creative Commons Attribution 4.0 Unported License (CC BY-NC 4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. See https://creativecommons.org/licenses/by-nc/4.0/.

          History
          Categories
          Journal Article

          Medicine,Image processing,Radiology & Imaging,Bioinformatics & Computational biology,Health & Social care,Public health
          ECT,Quality Control,Intrinsic Spatial Linearity

          Comments

          Comment on this article