      Quality Assurance and Commissioning of an Infrared Marker-Based Patient Positioning System for Frameless Extracranial Stereotactic Radiotherapy

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Rapid advancements in imaging technology have led to remarkable improvements in identification and localization of tumors, ushering the era of high-precision techniques in contemporary radiotherapy practice. However, uncertainties in patient set-up and organ motion during a course of fractionated radiotherapy can compromise precision of radiation therapy. Excellent accuracy has been achieved with invasive and non-invasive fixation systems for stereotactic radiotherapy. This report describes the commissioning procedure and Quality Assurance studies done to evaluate the accuracy of isocenter localization by an infrared marker-based positioning system (ExacTrac). The ExacTrac has two infrared cameras that emit and detect infrared rays from reflective markers and construct three-dimensional coordinates of each marker. It detects the difference of the actual isocenter position from the planned isocenter coordinates in three translational (lateral, longitudinal, vertical, or x,y,z axes) and three rotational axes (six degree of freedom). This study performed on a flat and static phantom shows excellent accuracy achieved by the ExacTrac system. The positioning accuracy of ExacTrac (± 1 mm translational displacement and ± 1° rotational errors) can be a valuable tool in implementing frameless extracranial stereotactic radiotherapy. Nevertheless, it needs to be further evaluated on patients with inherent motion and greater positional uncertainty before being adopted in clinical practice.

          Related collections

          Most cited references22

          • Record: found
          • Abstract: found
          • Article: not found

          Errors and margins in radiotherapy.

          Clinical radiotherapy procedures aim at high accuracy. However, there are many error sources that act during treatment preparation and execution that limit the accuracy. As a consequence, a safety margin is required to ensure that the planned dose is actually delivered to the target for (almost) all patients. Before treatment planning, a planning computed tomography scan is made. In particular, motion of skin with respect to the internal anatomy limits the reproducibility of this step, introducing a systematic setup error. The second important error source is organ motion. The tumor is imaged in an arbitrary position, leading to a systematic organ motion error. The image may also be distorted because of the interference of the scanning process and organ motion. A further systematic error introduced during treatment planning is caused by the delineation process. During treatment, the most important errors are setup error and organ motion leading to day-to-day variations. There are many ways to define the margins required for these errors. In this article, an overview is given of errors in radiotherapy and margin recipes, based on physical and biological considerations. Respiration motion is treated separately.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Target localization and real-time tracking using the Calypso 4D localization system in patients with localized prostate cancer.

            The Calypso 4D Localization System is being developed to provide accurate, precise, objective, and continuous target localization during radiotherapy. This study involves the first human use of the system, to evaluate the localization accuracy of this technique compared with radiographic localization and to assess its ability to obtain real-time prostate-motion information. Three transponders were implanted in each of 20 patients. Eleven eligible patients of the 20 patients participated in a study arm that compared radiographic triangulated transponder locations to electromagnetically recorded transponder locations. Transponders were tracked for 8-min periods. The implantations were all successful, with no major complications. Intertransponder distances were largely stable. Comparison of the patient localization on the basis of transponder locations as per the Calypso system with the radiographic transponder localization showed an average (+/-SD) 3D difference of 1.5 +/- 0.9 mm. Upon tracking during 8 min, 2 of the 11 patients showed significant organ motion (>1 cm), with some motion lasting longer that 1 min. Calypso transponders can be used as magnetic intraprostatic fiducials. Clinical evaluation of this novel 4D nonionizing electromagnetic localization system with transponders indicates a comparable localization accuracy to isocenter, (within 2 mm) compared with X-ray localization.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              The utility of megavoltage computed tomography images from a helical tomotherapy system for setup verification purposes.

              To evaluate the utility of relatively low-dose megavoltage computed tomography (MVCT) images from a clinical helical tomotherapy system for setup verification purposes. Cross-sectional kilovolt computed tomography (kVCT) images were obtained for treatment planning purposes on a diagnostic third-generation CT scanner, followed by MVCT images from a helical tomotherapy system in 8 pet dogs with spontaneously occurring tumors. The kVCT and MVCT images were aligned for setup verification purposes, allowing repositioning before treatment delivery. Tumors are readily visualized on the MVCT images. At a dose of 2-3 cGy, the MVCT images are of sufficient quality for verification of treatment setup, but soft-tissue contrast is inferior to that with conventional kVCT. The MV and kVCT images were successfully aligned. When necessary, patients undergoing helical tomotherapy were repositioned before treatment. Megavoltage CT image quality is sufficient for tumor identification and three-dimensional setup verification in dogs with spontaneous tumors. The MVCT images can be aligned with the planning kVCT to ensure proper patient registration before treatment. Image alignment was successful in these canine patients, despite no skin markings defining patient positioning between the two scans. MVCT images facilitate setup verification, and their tomographic nature offers improvements over conventional portal imaging.
                Bookmark

                Author and article information

                Journal
                Int J Biomed Sci
                Int J Biomed Sci
                IJBS
                International Journal of Biomedical Science : IJBS
                Master Publishing Group
                1550-9702
                1555-2810
                December 2007
                : 3
                : 4
                : 298-301
                Affiliations
                [1 ] Department of Radiation Oncology, Advanced Centre for Treatment Research & Education in Cancer (ACTREC), Tata Memorial Centre, Kharghar, Navi Mumbai, India;
                [2 ] Department of Radiation Oncology, Tata Memorial Hospital, Mumbai, India
                Author notes
                Corresponding author: Tejpal Gupta, Assistant Professor and Consultant Radiation Oncologist, Department of Radiation Oncology, ACTREC, Tata Memorial Centre, Kharghar, Navi Mumbai: 410208, India. Tel: 91-22-27405057; Fax: 91-22-27405061; E-mail: tejpalgupta@ 123456rediffmail.com .

                Note: It attributed to Department of Radiation Oncology, ACTREC, Tata Memorial Centre, and no source of funding was involved in this study.

                Article
                IJBS-3-298
                3614655
                23675057
                bae780fb-f3c8-4def-9d99-f4c09280c465
                © Tejpal Gupta et al. Licensee Master Publishing Group

                This is an open-access article distributed under the terms of the Creative Commons Attribution License ( http://creativecommons.org/licenses/by/2.5/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.

                History
                Categories
                Article

                stereotactic radiotherapy,patient positioning system,extracranial,infrared marker,quality assurance

                Comments

                Comment on this article