Intensity-modulated radiation therapy using static ports of tomotherapy (TomoDirect): comparison with the TomoHelical mode

Read this article at

Bookmark
      There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

      Abstract

      PurposeWith the new mode of Tomotherapy, irradiation can be delivered using static ports of the TomoDirect mode. The purpose of this study was to evaluate the characteristics of TomoDirect plans compared to conventional TomoHelical plans.MethodsTomoDirect and TomoHelical plans were compared in 46 patients with a prostate, thoracic wall or lung tumor. The mean target dose was used as the prescription dose. The minimum coverage dose of 95% of the target (D95%), conformity index (CI), uniformity index (UI), dose distribution in organs at risk and treatment time were evaluated. For TomoDirect, 2 to 5 static ports were used depending on the tumor location.ResultsFor the prostate target volume, TomoDirect plans could not reduce the rectal dose and required a longer treatment time than TomoHelical. For the thoracic wall target volume, the V5Gy of the lung or liver was lower in TomoDirect than in TomoHelical (p = 0.02). For the lung target volume, TomoDirect yielded higher CI (p = 0.009) but smaller V5Gy of the lung (p = 0.005) than TomoHelical. Treatment time did not differ significantly between the thoracic wall and lung plans.ConclusionProstate cancers should be treated with the TomoHelical mode. Considering the risk of low-dose radiation to the lung, the TomoDirect mode could be an option for thoracic wall and lung tumors.

      Related collections

      Most cited references 26

      • Record: found
      • Abstract: found
      • Article: not found

      A simple scoring ratio to index the conformity of radiosurgical treatment plans. Technical note.

       I Paddick (2000)
      A conformity index is a measure of how well the volume of a radiosurgical dose distribution conforms to the size and shape of a target volume. Because the success of radiosurgery is related to the extremely conformal irradiation of the target, an accurate method for describing this parameter is important. Existing conformity ratios and indices used in radiosurgery are reviewed and criticized. It will be demonstrated that previously proposed measurements of conformity can, under certain conditions, give false perfect scores. A new conformity index is derived that gives an objective score of conformity for a treatment plan and gives no false scores. An analysis of five different treatment plans is made using both the existing scoring methods and the new conformity index.
        Bookmark
        • Record: found
        • Abstract: found
        • Article: not found

        Radiation dose-volume effects in the lung.

        The three-dimensional dose, volume, and outcome data for lung are reviewed in detail. The rate of symptomatic pneumonitis is related to many dosimetric parameters, and there are no evident threshold "tolerance dose-volume" levels. There are strong volume and fractionation effects. Copyright 2010 Elsevier Inc. All rights reserved.
          Bookmark
          • Record: found
          • Abstract: found
          • Article: not found

          Tomotherapy: a new concept for the delivery of dynamic conformal radiotherapy.

          Tomotherapy, literally "slice therapy," is a proposal for the delivery of radiation therapy with intensity-modulated strips of radiation. The proposed method employs a linear accelerator, or another radiation-emitting device, which would be mounted on a ring gantry like a CT scanner. The patient would move through the bore of the gantry simultaneously with gantry rotation. The intensity modulation would be performed by temporally modulated multiple independent leaves that open and close across the slit opening. At any given time, any leaf would be (1) closed, covering a portion of the slit, (2) open, allowing radiation through, or (3) changing between these states. This method would result in the delivery of highly conformal radiation. Overall treatment times should be comparable with contemporary treatment delivery times. The ring gantry would make it convenient to mount a narrow multisegmented megavoltage detector system for beam verification and a CT scanner on the treatment unit. Such a treatment unit could become a powerful tool for treatment planning, conformal treatment, and verification using tomographic images. The physical properties of this treatment delivery are evaluated and the fundamental design specifications are justified.
            Bookmark

            Author and article information

            Affiliations
            [1 ]Department of Radiology, Nagoya City University Graduate School of Medical Sciences, 1 Kawasumi, Mizuho-cho, Mizuho-ku, Nagoya 467-8601, Japan
            [2 ]Department of Radiation Oncology, JA Suzuka General Hospital, Suzuka, Japan
            Contributors
            Journal
            Radiat Oncol
            Radiat Oncol
            Radiation Oncology (London, England)
            BioMed Central
            1748-717X
            2013
            21 March 2013
            : 8
            : 68
            23517931
            3643840
            1748-717X-8-68
            10.1186/1748-717X-8-68
            Copyright ©2013 Murai et al.; licensee BioMed Central Ltd.

            This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

            Categories
            Research

            Comments

            Comment on this article