      Evaluation of the Dade Behring N Latex Cystatin C assay on the Dade Behring Nephelometer II System.

      Scandinavian Journal of Clinical and Laboratory Investigation
      Calibration, Cystatin C, Cystatins, blood, Cysteine Proteinase Inhibitors, Humans, Immunoassay, Nephelometry and Turbidimetry, Regression Analysis

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          The Dade Behring N Latex Cystatin C assay, a particle-enhanced nephelometric immunoassay for measuring serum cystatin C, was evaluated on the Dade Behring Nephelometer II. The assay time was 6 min and the throughput was 75 samples per hour. The sample volume was 40 microL and the measuring range was 0.25-7.90 mg/L. Imprecision studies revealed within-run CVs < 1.8% and between-run CVs < 1.8% in the concentration range 0.87-4.63 mg/L. Recovery was 92.4-101.3%. Linearity studies showed excellent correlation between the theoretical and obtained values. No interferences were detected from haemoglobin < 1.0 mmol/ L, bilirubin <512 micromol/L and Intralipid <20 g/L. Stability of cystatin C in serum was 7 days at temperatures from 20 degrees C to 20 degrees C and 6 months at -80 degrees C. Measurements of cystatin C in heparin-plasma and EDTA-plasma did not differ significantly from cystatin C measured in serum. Fifty patient samples run on the Dade Behring Nephelometer II (y) were compared to the Dako Cystatin C assay (x). The Passing-Bablok regression analysis revealed y = 1.105x - 0.340. In conclusion, the Dade Behring N Latex Cystatin C assay was precise and correlated with the Dako Cystatin C assay.

          Related collections

          Author and article information

          Journal
          10206092
          10.1080/00365519950185940

          Chemistry
          Calibration,Cystatin C,Cystatins,blood,Cysteine Proteinase Inhibitors,Humans,Immunoassay,Nephelometry and Turbidimetry,Regression Analysis

          Comments

          Comment on this article