A cementless, elastic press-fit socket with and without screws : A 2-year randomized controlled radiostereometric analysis of 37 hips

Read this article at

Bookmark
      There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

      Abstract

      Background The acetabular component has remained the weakest link in hip arthroplasty regarding achievement of long-term survival. Primary fixation is a prerequisite for long-term performance. For this reason, we investigated the stability of a unique cementless titanium-coated elastic monoblock socket and the influence of supplementary screw fixation.Patient and methods During 2006–2008, we performed a randomized controlled trial on 37 patients (mean age 63 years (SD 7), 22 females) in whom we implanted a cementless press-fit socket. The socket was implanted with additional screw fixation (group A, n = 19) and without additional screw fixation (group B, n = 18). Using radiostereometric analysis with a 2-year follow-up, we determined the stability of the socket. Clinically relevant migration was defined as > 1 mm translation and > 2º rotation. Clinical scores were determined.Results The sockets without screw fixation showed a statistically significantly higher proximal translation compared to the socket with additional screw fixation. However, this higher migration was below the clinically relevant threshold. The numbers of migratory sockets were not significantly different between groups. After the 2-year follow-up, there were no clinically relevant differences between groups A and B regarding the clinical scores. 1 patient dropped out of the study. In the others, no sockets were revised.Interpretation We found that additional screw fixation is not necessary to achieve stability of the cementless press-fit elastic RM socket. We saw no postoperative benefit or clinical effect of additional screw fixation.

      Related collections

      Most cited references 40

      • Record: found
      • Abstract: found
      • Article: not found

      Guidelines for standardization of radiostereometry (RSA) of implants.

      There is a need for standardization of radiostereometric (RSA) investigations to facilitate comparison of outcome reported from different research groups. In this document, 6 research centers have agreed upon standards for terminology, description and use of RSA arrangement including radiographic set-up and techniques. Consensus regarding minimum requirements for marker stability and scatter, choice of coordinate systems, and preferred way of describing prosthetic micromotion is of special interest. Some notes on data interpretation are also presented. Validation of RSA should be standardized by preparation of protocols for assessment of accuracy and precision. Practical issues related to loading of the joint by weight bearing or other conditions, follow-up intervals, length of follow-up, radiation dose, and the exclusion of patients due to technical errors are considered. Finally, we present a checklist of standardized output that should be included in any clinical RSA paper.This document will form the basis of a detailed standardization protocol under supervision of ISO and the European Standards Working Group on Joint Replacement Implants (CEN/TC 285/WG4). This protocol will facilitate inclusion of RSA in a standard protocol for implant testing before it is released for general use. Such a protocol-also including other recognized clinical outcome parameters-will reduce the risk of implanting potentially inferior prostheses on a large scale.
        Bookmark
        • Record: found
        • Abstract: found
        • Article: found
        Is Open Access

        Notes on the use and interpretation of radiostereometric analysis

        ABSTRACT With increasing numbers of research groups carrying out radiostereometric analysis (RSA), it is important to reach a consensus on how the main aspects of the technique should be carried out and how the results should be presented in an appropriate and consistent way. In this collection of guidelines, we identify a number of methodological and reporting issues including: measurement error and precision, migration and migration direction data, and the use of RSA as a screening technique. Alternatives are proposed, and a statistical analysis is presented, from which a sample size of 50 is recommended for screening of newly introduced prostheses.
          Bookmark
          • Record: found
          • Abstract: not found
          • Article: not found

          The role of access of joint fluid to bone in periarticular osteolysis. A report of four cases.

            Bookmark

            Author and article information

            Affiliations
            1Department Orthopaedic and Trauma Surgery, Orthopaedic Centre OCON , Hengelo
            2Departments of Research, Development and Education
            3Orthopaedic Surgery, Sint Maartenskliniek, Nijmegen , the Netherlands
            Author notes
            Journal
            Acta Orthop
            Acta Orthop
            ORT
            Acta Orthopaedica
            Informa Healthcare
            1745-3674
            1745-3682
            October 2012
            22 October 2012
            : 83
            : 5
            : 481-487
            23083434
            3488174
            10.3109/17453674.2012.720116
            ORT_A_720116_O
            Copyright: © Nordic Orthopaedic Federation

            This is an open-access article distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the source is credited.

            Categories
            Hip

            Orthopedics

            Comments

            Comment on this article