      Analytical isotachophoresis of lactate in human serum using dry film photoresist microfluidic chips compatible with a commercially available field-deployable instrument platform.

      Read this article at

      ScienceOpenPublisherPubMed
      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          A dry film resist (DFR) chip compatible with the Agilent Bioanalyzer 2100 was designed and fabricated for use in the analysis of lactate in serum by chip isotachophoresis (ITP). The Agilent Bioanalyzer 2100 is a commercially available field deployable analytical instrument originally developed for the electrophoretic analysis of DNA, RNA and proteins. The DFR chip was designed for the ITP separation of lactate in human serum within 1 min and was made compatible with the Bioanalyzer after packaging in the plastic caddies normally used for the DNA chips. A 20-fold improvement in sensitivity was obtained for the DFR chips in comparison with the standard chips used in earlier work. The limit of detection and limit of quantification for lactate were 24 μM and 80 μM, respectively. This new approach enables the use of commercial platforms like the Agilent Bioanalyzer for new applications including the analysis of small molecules.

          Related collections

          Author and article information

          Journal
          Anal. Chim. Acta
          Analytica chimica acta
          Elsevier BV
          1873-4324
          0003-2670
          Nov 25 2013
          : 803
          Affiliations
          [1 ] University of Tasmania, ACROSS and School of Chemistry, Hobart, Australia.
          Article
          S0003-2670(13)00194-3
          10.1016/j.aca.2013.01.046
          24216207
          cf85eb45-6ff2-4646-81cc-52ce91b46e69
          History

          Serum,R6G,PC,Microfluidics,LE,Isotachophoresis,Indirect detection,HPMC,Fluorescence,DFR,terminating electrolyte,rhodamine 6G,polycarbonate,leading electrolyte,hydroxypropylmethyl cellulose,dry film resist,TE

          Comments

          Comment on this article