      Highly accelerated 3D myocardial late fadolinium enhancement MRI using ESPIRiT compressed sensing: initial feasibility

      research-article
      1 , , 2 , 3 , 2
      Journal of Cardiovascular Magnetic Resonance
      BioMed Central
      17th Annual SCMR Scientific Sessions
      16-19 January 2014

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background Myocardial late gadolinium enhancement (LGE) MRI has become a standard clinical exam for characterizing myocardial viability after infarction and for assessing many nonischemic myocardial diseaseses. Typically, a 3D slab covering the myocardium is scanned within a single breath-hold. However, 3D LGE with sufficient resolution and slice coverage requires a long scan time and is challenging in cardiac patients with limited breath-hold capability. Recently, compressed sensing (CS) [1] has demonstrated the ability to further accelerate MRI on top of parallel imaging (PI) and has demonstrated promising results for 3D LGE [2]. This work intended to evaluate the feasibility of a CS-PI method based on ESPIRiT [3-5] and optimize its data acquisition for improving the robustness of 3D LGE with high acceleration. Methods k-space sampling: As shown in Figure 1, a variable density Poisson disk sampling (VDPDS) scheme was used for CS acceleration, with variable acceleration depending on off-center distance and Poisson disk distribution at each local k-space region. For accurate reconstruction, the acceleration quadratically increases from center to outer k-space. Partial Fourier along ky was used to further reduce scan time. Image reconstruction: As described in ESPIRiT [3-5], first, coil sensitivity was estimated from calibration data-based eigenvector computations. Next, k-space was completed by iteratively enforcing coil sensitivity and signal sparsity until convergence. Full-kspace 3D LGE datasets were collected from 3 patients on a GE 1.5T system using 3D FGRE and an 8 channel cardiac coil. Typical imaging parameters included: 15° flip angle, matrix size of 200 × 200, 14 slices with 10 mm thickness, TR/TE = 2.5/1.1 ms, partial Fourier factor of 0.6 along ky, 28 s scan time. TI was determined on a per-patient basis from a TI scout scan. Full k-space was downsampled offline using the VDPDS scheme and reconstructed using ESPIRiT. For comparison, conventional autocalibrated PI was also simulated using regular 2D acceleration and reconstructed using a GE product reconstruction (ARC). Figure 1 Variable density Poisson disk sampling for ESPIRiT. Results Figure 2 shows 2 representative cases. In case 1 (left), ARC (b, c) produces substantial noise amplification and visible aliasing artifacts at these high accelerations, which interfere with pathology at inferoposterior endocardium. In comparison, ESPIRiT (d, e) generates image quality and delineation of the delayed enhancement comparable to full k-space image (a), though with slight edge blurring. In another case (right), similarly, ESPIRiT (i, j) significantly suppresses artifacts and provides superior SNR vs. ARC (g, h). Figure 2 Comparison of different accelerations and reconstructions in case 1 (left) and case 2 (right). Full k-space image (a, f) and simulated reconstruction using ARC with 2 × 2 (b, g) & 3 × 2 (c, h) and ESPIRiT with net acceleration same as ARC 2 × 2 (d, i) & 3 × 2 (e, j). Conclusions In this initial study, ESPIRiT with VDPDS was shown to enable higher acceleration with better image quality than conventional PI and is a promising approach for improving the robustness of 3D LGE by shortening breathholding time.

          Related collections

          Author and article information

          Conference
          J Cardiovasc Magn Reson
          J Cardiovasc Magn Reson
          Journal of Cardiovascular Magnetic Resonance
          BioMed Central
          1097-6647
          1532-429X
          2014
          16 January 2014
          : 16
          : Suppl 1
          : W25
          Affiliations
          [1 ]MR Applications and Workflow, GE Healthcare, Menlo Park, California, USA
          [2 ]MR Applications and Workflow, GE Healthcare, Garching, Munchen, Germany
          [3 ]Department of Cardiovascular Imaging, Centro Cardiologico Monzino, Milan, Italy
          Article
          1532-429X-16-S1-W25
          10.1186/1532-429X-16-S1-W25
          4043535
          d2a0d748-6c1e-4a2b-92e2-64366314263a
          Copyright © 2014 Lai et al.; licensee BioMed Central Ltd.

          This is an Open Access article distributed under the terms of the Creative Commons Attribution License ( http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver ( http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

          17th Annual SCMR Scientific Sessions
          New Orleans, LA, USA
          16-19 January 2014
          History
          Categories
          Workshop Presentation

          Cardiovascular Medicine
          Cardiovascular Medicine

          Comments

          Comment on this article