      Retrieval analysis of Harris-Galante I and II acetabular liners in situ for more than 10 years

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background and purpose

          There have been few reports documenting the wear and oxidation performance of the polyethylene bearing surface of HGPI and HGPII THA devices.

          We evaluated retrieved HGPI and HGPII acetabular liners that had been in situ for more than 10 years and determined whether there was a relationship between clinical and radiographic factors, surface damage, wear, and oxidation.

          Materials and methods

          129 HGPI and II acetabular liners with implantation times of > 10 years were retrieved at 4 institutions between 1997 and 2010. The liners were made from a single resin and were gamma radiation-sterilized in air. Surface damage, linear wear, and oxidation index (OI) were assessed. Differences in clinical and radiographic factors, surface damage, linear wear, and OI for the 2 designs were statistically evaluated separately and together.

          Results

          Articular surface damage and backside damage was similar in the 2 designs. The linear penetration rate was 0.14 (SD 0.07) mm/year for the HGPI liners and 0.12 (SD 0.08) mm/year for the HGPII liners. For both cohorts, the rim had a higher OI than the articular surface. 74% of the liners had subsurface cracking and 24% had a complete fracture through the acetabular rim.

          Interpretation

          Despite modification of the HGP locking mechanism in the HGPII design, dissociation of the liner from the acetabular shell can still occur if fracture of the rim of the liner develops due to oxidative degradation.

          Related collections

          Most cited references29

          • Record: found
          • Abstract: found
          • Article: not found

          Radiological demarcation of cemented sockets in total hip replacement.

          The frequency of radiological demarcation of the cement-bone junction in the acetabulum after total hip replacement has been examined in 141 Charnley low-friction arthroplasties followed for an average of 10.1 years. Sixty-nine per cent showed demarcation of various degrees and 9.2 per cent of the series showed evidence of progressive migration of the socket. The vast majority of cases with demarcation were symptomless. In most cases where demarcation was accompanied by migration the operation notes suggested a technical explanation and in three cases low-grade sepsis was responsible. The fact that nearly 30 per cent of cases showed no demarcation even after 10 years supports the idea that there is no fundamental defect in the principle of employing cement in the acetabulum. Better surgical technique may increase the number of cases showing no demarcation.
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            "Modes of failure" of cemented stem-type femoral components: a radiographic analysis of loosening.

            In view of the increasing incidence of stem-type femoral component loosening, a detailed retrospective radiographic zonal analysis of 389 total hip replacements indicated a 19.5% incidence (76 hips) of radiological evidences of mechanical looseness, i.e., fractured acrylic cement and/or a radiolucent gap at the stem-cement or cement-bone interfaces. Detailed serial radiographic examination demonstrated progressive loosening in 56 of the 76 hips and these were categorized into mechanical modes of failure. The 4 modes of failure characterizing stem-type component progressive loosening mechanisms consisted of stem pistoning within the acrylic (3.3%), cement-embedded stem pistoning with the femur (5.1%), medial midstem pivot (2.5%), calcar pivot (0.7%) and bending (fatigue) cantilever (3.3%).
              Bookmark
              • Record: found
              • Abstract: found
              • Article: not found

              A literature review of the association between wear rate and osteolysis in total hip arthroplasty.

              The establishment of a polyethylene wear rate threshold for the development of osteolysis at the hip would allow surgeons to identify patients at risk for osteolysis and to implement selective, more frequent follow-up. We reviewed publications that met certain criteria for wear and osteolysis measurement. Based on this review, the incidence of osteolysis increases as the rate of wear increases. The literature indicates that osteolysis rarely is observed at a wear rate of <0.1 mm/y. We suggest that a practical wear rate threshold of 0.05 mm/y would eliminate osteolysis. This wear threshold suggests that the new cross-linked polyethylenes would reduce osteolysis, provided that in vivo wear rates mirror those observed in vitro. To facilitate future comparison of published data, we suggest that longitudinal wear studies adopt consistent edge detection-based wear measurement techniques and uniform osteolytic lesion classification and measurement schema. Copyright 2002, Elsevier Science (USA). All rights reserved.
                Bookmark

                Author and article information

                Journal
                Acta Orthop
                Acta Orthop
                ORT
                Acta Orthopaedica
                Informa Healthcare
                1745-3674
                1745-3682
                August 2012
                25 August 2012
                : 83
                : 4
                : 366-373
                Affiliations
                1simpleState University of New York Medical Center , New York, NY
                2simpleOrthopaedic Implant Retrieval Analysis Laboratory, Department of Orthopaedics, Case Western Reserve University , Cleveland, OH
                3simpleImplant Research Center, School of Biomedical Engineering, Science, and Health Systems, Drexel University and Exponent Inc. , Philadelphia, PA
                4simpleDepartments of Mechanical and Aerospace Engineering and Orthopaedics, Case Western Reserve University , Cleveland, OH, USA
                Author notes
                Article
                ORT_A_717843_O
                10.3109/17453674.2012.717843
                3427627
                22880709
                d2a4ebbb-58ed-4433-9ba6-f5fdec80e00a
                Copyright: © Nordic Orthopaedic Federation

                This is an open-access article distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the source is credited.

                History
                : 29 June 2011
                : 19 April 2012
                Categories
                Hip

                Orthopedics
                Orthopedics

                Comments

                Comment on this article