      Continuous Flash Glucose Monitoring in children with Congenital Hyperinsulinism; first report on accuracy and patient experience

      research-article

      Read this article at

      Bookmark
          There is no author summary for this article yet. Authors can add summaries to their articles on ScienceOpen to make them more accessible to a non-specialist audience.

          Abstract

          Background

          The factory calibrated FreeStyle Libre (FSL) flash glucose monitoring system has been recently introduced for use in patients with diabetes mellitus. There are no reports available regarding its use in patients with congenital hyperinsulinism (CHI). We have assessed the accuracy of FSL compared to the finger prick capillary blood glucose (CBG) over 2 weeks period in patients with CHI and evaluated the parents’ experience of using FSL.

          Methods

          Four hundred sixty-seven episodes of CBG along with corresponding swipe FSL readings were available from 11 children with CHI (0.5–5 years). A detailed questionnaire was completed by the parents.

          Results

          The mean variation between the two methods was 0.29 mmol/l (SD ±1.07), higher readings by FSL compared to CBG. The FSL sensors stayed in-situ for an average period of 11.5 days. There was a positive correlation between the two methods ( r = 0.7). The FSL tended to overestimate compared to CBG (bias = 0.29 mmol/l; 95% CI: 0.19 to 0.38). Only 70% of values were within the reference standard (±0.83 mmol/l) at glucose concentrations less than 5.6 mmol/l. The overall Mean Absolute Relative Difference (MARD) was 17.9%. Forty two episodes of hypoglycaemia (CBG < 3.5 mmol/l) were noted but FSL identified only 52% of these episodes. The Bland Altman analysis showed the 95% limits of agreement between the two methods ranging from − 1.8 (95% CI: -1.97 to − 1.64) to 2.37 (95% CI: 2.21 to 2.54). Majority of the parents found the glucose trend on FSL to be useful to detect and prevent hypoglycaemic episodes. All parents felt that FSL is a very easy and convenient method to measure the glucose especially during sleep. A significant proportion of parents felt that FSL readings were not accurate and 56% of parents expressed interest to continue using FSL after the trial period.

          Conclusion

          Noticeable variability between the two methods of measuring the glucose was noted. Despite the ease of using the FSL system, concerns related to accuracy, especially at low glucose values do remain although parents find the glucose trend to be very useful. Further larger trials are needed in CHI patients before FSL is recommended as a routine alternative method for measuring glucose levels.

          Related collections

          Most cited references9

          • Record: found
          • Abstract: found
          • Article: not found

          Effectiveness of sensor-augmented insulin-pump therapy in type 1 diabetes.

          Recently developed technologies for the treatment of type 1 diabetes mellitus include a variety of pumps and pumps with glucose sensors. In this 1-year, multicenter, randomized, controlled trial, we compared the efficacy of sensor-augmented pump therapy (pump therapy) with that of a regimen of multiple daily insulin injections (injection therapy) in 485 patients (329 adults and 156 children) with inadequately controlled type 1 diabetes. Patients received recombinant insulin analogues and were supervised by expert clinical teams. The primary end point was the change from the baseline glycated hemoglobin level. At 1 year, the baseline mean glycated hemoglobin level (8.3% in the two study groups) had decreased to 7.5% in the pump-therapy group, as compared with 8.1% in the injection-therapy group (P<0.001). The proportion of patients who reached the glycated hemoglobin target (<7%) was greater in the pump-therapy group than in the injection-therapy group. The rate of severe hypoglycemia in the pump-therapy group (13.31 cases per 100 person-years) did not differ significantly from that in the injection-therapy group (13.48 per 100 person-years, P=0.58). There was no significant weight gain in either group. In both adults and children with inadequately controlled type 1 diabetes, sensor-augmented pump therapy resulted in significant improvement in glycated hemoglobin levels, as compared with injection therapy. A significantly greater proportion of both adults and children in the pump-therapy group than in the injection-therapy group reached the target glycated hemoglobin level. (Funded by Medtronic and others; ClinicalTrials.gov number, NCT00417989.) 2010 Massachusetts Medical Society
            Bookmark
            • Record: found
            • Abstract: found
            • Article: not found

            Practical management of hyperinsulinism in infancy.

            Hyperinsulinism in infancy is one of the most difficult problems to manage in contemporary paediatric endocrinology. Although the diagnosis can usually be achieved without difficulty, it presents the paediatrician with formidable day to day management problems. Despite recent advances in understanding the pathophysiology of hyperinsulinism, the neurological outcome remains poor, and there is often a choice of unsatisfactory treatments, with life long sequelae for the child and his or her family. This paper presents a state of the art overview on management derived from a consensus workshop held by the European network for research into hyperinsulinism (ENRHI). The consensus is presented as an educational aid for paediatricians and children's nurses. It offers a practical guide to management based on the most up to date knowledge. It presents a proposed management cascade and focuses on the clinical recognition of the disease, the immediate steps that should be taken to stabilise the infant during diagnostic investigations, and the principles of definitive treatment.
              Bookmark
              • Record: found
              • Abstract: found
              • Article: found
              Is Open Access

              An alternative sensor-based method for glucose monitoring in children and young people with diabetes

              Objective To determine accuracy, safety and acceptability of the FreeStyle Libre Flash Glucose Monitoring System in the paediatric population. Design, setting and patients Eighty-nine study participants, aged 4–17 years, with type 1 diabetes were enrolled across 9 diabetes centres in the UK. A factory calibrated sensor was inserted on the back of the upper arm and used for up to 14 days. Sensor glucose measurements were compared with capillary blood glucose (BG) measurements. Sensor results were masked to participants. Results Clinical accuracy of sensor results versus BG results was demonstrated, with 83.8% of results in zone A and 99.4% of results in zones A and B of the consensus error grid. Overall mean absolute relative difference (MARD) was 13.9%. Sensor accuracy was unaffected by patient factors such as age, body weight, sex, method of insulin administration or time of use (day vs night). Participants were in the target glucose range (3.9–10.0 mmol/L) ∼50% of the time (mean 12.1 hours/day), with an average of 2.2 hours/day and 9.5 hours/day in hypoglycaemia and hyperglycaemia, respectively. Sensor application, wear/use of the device and comparison to self-monitoring of blood glucose were rated favourably by most participants/caregivers (84.3–100%). Five device related adverse events were reported across a range of participant ages. Conclusions Accuracy, safety and user acceptability of the FreeStyle Libre System were demonstrated for the paediatric population. Accuracy of the system was unaffected by subject characteristics, making it suitable for a broad range of children and young people with diabetes. Trial registration number NCT02388815.
                Bookmark

                Author and article information

                Contributors
                01512525281 , senthilkss@yahoo.co.uk
                Journal
                Int J Pediatr Endocrinol
                Int J Pediatr Endocrinol
                International Journal of Pediatric Endocrinology
                BioMed Central (London )
                1687-9848
                1687-9856
                27 March 2018
                27 March 2018
                2018
                : 2018
                : 3
                Affiliations
                [1 ]ISNI 0000 0004 0398 4295, GRID grid.415892.3, Department of Paediatrics, , Leighton Hospital, ; Crewe, CW1 4QJ UK
                [2 ]ISNI 0000 0001 0503 2798, GRID grid.413582.9, Department of Paediatric Endocrinology, , Alder Hey Children’s Hospital, ; L14 5AB, Liverpool, UK
                Article
                57
                10.1186/s13633-018-0057-2
                5870486
                29599801
                d3e69e81-a25b-4b33-9b40-9672717b5e4a
                © The Author(s). 2018

                Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License ( http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver ( http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

                History
                : 6 December 2017
                : 9 March 2018
                Categories
                Research
                Custom metadata
                © The Author(s) 2018

                Pediatrics
                hyperinsulinism,flash glucose monitoring,freestyle libre
                Pediatrics
                hyperinsulinism, flash glucose monitoring, freestyle libre

                Comments

                Comment on this article